{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2026,11,5]],"date-time":"2026-11-05T20:53:55Z","timestamp":1793912035239,"version":"4.8.0"},"reference-count":25,"publisher":"American Speech Language Hearing Association","issue":"1","content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":["Am J Audiol"],"published-print":{"date-parts":[[2022,3,3]]},"abstract":"<jats:sec>\n                    <jats:title>Purpose:<\/jats:title>\n                    <jats:p>This investigation aims to provide outcomes from a clinical perspective on the validity and efficacy of a wireless automated audiometer system that could be used in multiple settings when a sound booth is not accessible. Testing was conducted in a clinical setting under modified protocols meeting safety precautions during the COVID-19 pandemic.<\/jats:p>\n                  <\/jats:sec>\n                  <jats:sec>\n                    <jats:title>Method:<\/jats:title>\n                    <jats:p>\n                      Four doctoral students in audiology served as examiners. Participants were 69 adults between the ages of 20 and 69 years, with normal hearing (\u2264 25 dB HL;\n                      <jats:italic>n<\/jats:italic>\n                      = 110 ears) or hearing loss (&gt; 25 dB HL;\n                      <jats:italic>n<\/jats:italic>\n                      = 25 ears). Two versions of a pure-tone air-conduction threshold test following a modified Hughson-Westlake approach were performed and compared at 500, 1000, 2000, 3000, 4000, 6000, and 8000 Hz (a) in a sound-treated test booth using standard manual audiometry and (b) in a quiet, nonsound-treated clinical room (sound booth free) using automated KUDUwave audiometry. Participants were asked to complete a five-item feedback questionnaire, and examiners were interviewed to report on their experience.\n                    <\/jats:p>\n                  <\/jats:sec>\n                  <jats:sec>\n                    <jats:title>Results:<\/jats:title>\n                    <jats:p>\n                      Clinical validity to within \u00b110 dB of standard audiometry was demonstrated for 94.5% of the total thresholds (\n                      <jats:italic>n<\/jats:italic>\n                      = 937) measured with the sound booth\u2013free approach. Less accuracy (73.3%) was observed using a \u00b15 dB comparison. When comparing the mean thresholds, there were significant differences (\n                      <jats:italic>p<\/jats:italic>\n                      &lt; .01) between the mean thresholds at most frequencies, with mean sound booth thresholds being higher than the sound booth\u2013free mean thresholds. A strong threshold correlation (.91\u2013.98) was found between the methods across frequencies. Participant and examiner feedback supported the efficacy of the sound booth\u2013free technology.\n                    <\/jats:p>\n                  <\/jats:sec>\n                  <jats:sec>\n                    <jats:title>Conclusions:<\/jats:title>\n                    <jats:p>Findings support sound booth\u2013free, automated software-controlled audiometry with active noise monitoring as a valid and efficient procedure for pure-tone hearing threshold assessment. This method offers an effective alternative when circumstances require more transportable hearing assessment technology or do not allow for standard manual audiometry in a sound booth.<\/jats:p>\n                  <\/jats:sec>","DOI":"10.1044\/2021_aja-21-00108","type":"journal-article","created":{"date-parts":[[2021,12,29]],"date-time":"2021-12-29T19:31:46Z","timestamp":1640806306000},"page":"91-100","source":"Crossref","is-referenced-by-count":12,"title":["Adapting Audiology Procedures During the Pandemic: Validity and Efficacy of Testing Outside a Sound Booth"],"prefix":"10.1044","volume":"31","author":[{"ORCID":"https:\/\/orcid.org\/0000-0002-6079-9139","authenticated-orcid":false,"given":"Yula C.","family":"Serpanos","sequence":"first","affiliation":[{"name":"Department of Communication Sciences and Disorders, Adelphi University, Garden City, NY"},{"name":"Long Island Doctor of Audiology (AuD) Consortium, Adelphi University, Garden City, NY"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"ORCID":"https:\/\/orcid.org\/0000-0001-7548-2006","authenticated-orcid":false,"given":"Melissa","family":"Hobbs","sequence":"additional","affiliation":[{"name":"RTI International, Research Triangle Park, NC"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Karina","family":"Nunez","sequence":"additional","affiliation":[{"name":"Department of Communication Sciences and Disorders, Adelphi University, Garden City, NY"},{"name":"Long Island Doctor of Audiology (AuD) Consortium, Adelphi University, Garden City, NY"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Lucia","family":"Gambino","sequence":"additional","affiliation":[{"name":"Department of Communication Sciences and Disorders, Adelphi University, Garden City, NY"},{"name":"Long Island Doctor of Audiology (AuD) Consortium, Adelphi University, Garden City, NY"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Jasmin","family":"Butler","sequence":"additional","affiliation":[{"name":"Department of Communication Sciences and Disorders, Adelphi University, Garden City, NY"},{"name":"Long Island Doctor of Audiology (AuD) Consortium, Adelphi University, Garden City, NY"}],"role":[{"vocabulary":"crossref","role":"author"}]}],"member":"1310","reference":[{"key":"e_1_3_2_2_1","unstructured":"American Academy of Audiology. (2020). Booth safety: COVID-19 and beyond. \nhttps:\/\/www.audiology.org\/news\/booth-safety-covid-19-and-beyond"},{"key":"e_1_3_2_3_1","unstructured":"American National Standards Institute. (2018). Maximum permissible ambient noise levels for audiometric test rooms (ANSI\/ASA S3.1-1999 (R2018)). \nhttps:\/\/webstore.ansi.org\/standards\/asa\/ansiasas31999r2018"},{"key":"e_1_3_2_4_1","unstructured":"American Speech-Language-Hearing Association. (n.d.). Audiology service delivery considerations in health care during coronavirus\/COVID-19. \nhttps:\/\/www.asha.org\/aud\/Audiology-Service-Delivery-Considerations-in-Health-Care-During-Coronavirus-COVID-19\/"},{"key":"e_1_3_2_5_1","unstructured":"American Speech-Language-Hearing Association. (2005). Guidelines for manual pure-tone threshold audiometry. \nhttps:\/\/www.asha.org\/policy\/gl2005-00014\/"},{"key":"e_1_3_2_6_1","doi-asserted-by":"publisher","DOI":"10.1044\/2018_AJA-17-0070"},{"key":"e_1_3_2_7_1","unstructured":"Centers for Disease Control and Prevention. (2020). Healthcare facilities: Managing operations during the COVID-19 pandemic. \nhttps:\/\/www.cdc.gov\/coronavirus\/2019-ncov\/hcp\/guidance-hcf.html?CDC_AA_refVal=https%3A%2F%2Fwww.cdc.gov%2Fcoronavirus%2F2019-ncov%2Fhealthcare-facilities%2Fguidance-hcf.html#:~:text=be- placed -on"},{"key":"e_1_3_2_8_1","doi-asserted-by":"publisher","DOI":"10.1044\/2016_AJA-16-0004"},{"key":"e_1_3_2_9_1","unstructured":"eMoyo Technologies. (n.d.). [Homepage]. https:\/\/emoyo.net\/"},{"key":"e_1_3_2_10_1","doi-asserted-by":"publisher","DOI":"10.1044\/2020_AJA-20-00037"},{"key":"e_1_3_2_11_1","doi-asserted-by":"publisher","DOI":"10.1044\/teles4.1.4"},{"key":"e_1_3_2_12_1","doi-asserted-by":"publisher","DOI":"10.3109\/14992027.2012.736692"},{"key":"e_1_3_2_13_1","doi-asserted-by":"publisher","DOI":"10.1080\/14992027.2020.1817994"},{"key":"e_1_3_2_14_1","volume-title":"Foundations of audiology: A practical approach","author":"Peterson M. E.","year":"2008","unstructured":"Peterson, M. E. , & Bell, T. S. (2008). Foundations of audiology: A practical approach. Pearson Merrill Prentice Hall."},{"key":"e_1_3_2_15_1","doi-asserted-by":"publisher","DOI":"10.4258\/hir.2018.24.4.263"},{"key":"e_1_3_2_16_1","doi-asserted-by":"publisher","DOI":"10.3109\/14992027.2014.920110"},{"key":"e_1_3_2_17_1","volume-title":"How the COVID-19 pandemic has affected provision of elective services: The challenges ahead","author":"Stuart B.","year":"2020","unstructured":"Stuart, B. (2020). How the COVID-19 pandemic has affected provision of elective services: The challenges ahead. HealthAffairs. https:\/\/www.healthaffairs.org\/do\/10.1377\/hblog20201006.263687\/full\/"},{"key":"e_1_3_2_18_1","doi-asserted-by":"publisher","DOI":"10.3109\/00206099109072873"},{"key":"e_1_3_2_19_1","doi-asserted-by":"publisher","DOI":"10.1080\/15459624.2011.559417"},{"key":"e_1_3_2_20_1","doi-asserted-by":"publisher","DOI":"10.1097\/01.HJ.0000669852.90548.75"},{"key":"e_1_3_2_21_1","doi-asserted-by":"publisher","DOI":"10.1258\/jtt.2010.090906"},{"key":"e_1_3_2_22_1","doi-asserted-by":"publisher","DOI":"10.3766\/jaaa.24.10.10"},{"issue":"11","key":"e_1_3_2_23_1","first-page":"777","article-title":"Pure-tone audiometry outside a sound booth using earphone attenuation, integrated noise monitoring, and automation","volume":"54","author":"Swanepoel D. W.","year":"2015","unstructured":"Swanepoel, D. W. , Matthysen, C. , Eikelboom, R. H. , Clark, J. L. , & Hall, J. W., III. (2015). Pure-tone audiometry outside a sound booth using earphone attenuation, integrated noise monitoring, and automation. International Journal of Audiology, 54(11), 777\u2013785.","journal-title":"International Journal of Audiology"},{"key":"e_1_3_2_24_1","doi-asserted-by":"publisher","DOI":"10.1089\/tmj.2009.0143"},{"key":"e_1_3_2_25_1","doi-asserted-by":"publisher","DOI":"10.1089\/tmj.2014.0243"},{"key":"e_1_3_2_26_1","unstructured":"World Health Organization. (2020). Pulse survey on continuity of essential health services during the COVID-19 pandemic: Interim report 27 August 2020 (No. WHO\/2019-nCoV\/EHS continuity\/survey\/2020.1). \nhttps:\/\/www.who.int\/publications\/i\/item\/WHO-2019-nCoV-EHS_continuity-survey-2020.1"}],"container-title":["American Journal of Audiology"],"original-title":[],"language":"en","link":[{"URL":"http:\/\/pubs.asha.org\/doi\/pdf\/10.1044\/2021_AJA-21-00108","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2022,3,3]],"date-time":"2022-03-03T13:48:38Z","timestamp":1646315318000},"score":1,"resource":{"primary":{"URL":"http:\/\/pubs.asha.org\/doi\/10.1044\/2021_AJA-21-00108"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[2022,3,3]]},"references-count":25,"journal-issue":{"issue":"1","published-print":{"date-parts":[[2022,3,3]]}},"alternative-id":["10.1044\/2021_AJA-21-00108"],"URL":"https:\/\/doi.org\/10.1044\/2021_aja-21-00108","relation":{},"ISSN":["1059-0889","1558-9137"],"issn-type":[{"value":"1059-0889","type":"print"},{"value":"1558-9137","type":"electronic"}],"subject":[],"published":{"date-parts":[[2022,3,3]]}}}