{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2026,10,6]],"date-time":"2026-10-06T21:51:14Z","timestamp":1791323474389,"version":"4.3.1"},"reference-count":19,"publisher":"Wiley","issue":"2","license":[{"start":{"date-parts":[[2001,12,25]],"date-time":"2001-12-25T00:00:00Z","timestamp":1009238400000},"content-version":"vor","delay-in-days":633,"URL":"http:\/\/onlinelibrary.wiley.com\/termsAndConditions#vor"}],"content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":["Paediatric Perinatal Epid"],"published-print":{"date-parts":[[2000,4]]},"abstract":"<jats:p>The aim of this study was to evaluate the Dinamap 8100 and the Omron M1 (the test devices) against standard criteria for use in children in a fieldwork setting. Device calibration and validation were carried out in accordance with the British Hypertension Society protocol for special groups such as children. A total of 102 children, recruited from infant and junior schools in Southampton, had six sequential measurements made of their blood pressure\u2013four measurements with a mercury sphygmomanometer and two with one of two test devices, 55 children with the Dinamap 8100 and 47 with the Omron M1. Systolic and diastolic readings with the Dinamap 8100 were on average 11 mmHg higher (95% confidence interval [CI]+9, +12 mmHg) and 3 mmHg lower (95% CI \u22125, \u22121 mmHg), respectively, than measurements with the mercury sphygmomanometer, overestimating systolic pressures and underestimating diastolic pressures across the whole range observed. The Omron M1 gave readings lower by 1 mmHg on average for systolic pressures and 2 mmHg for diastolic pressures compared with the sphygmomanometer (95% CIs \u22124, +1 mmHg and \u22125, +1 mmHg respectively), specifically overestimating higher pressures and underestimating lower pressures. According to the criteria of the British Hypertension Society, neither the Dinamap 8100 nor the Omron M1 can be recommended for use in children in clinical situations in which accuracy of the absolute measurement is required. In epidemiological surveys, in which differences in blood pressure between groups of people are more important than absolute levels, it may be more appropriate to use these devices. Of the two, its more consistent performance supports the Dinamap 8100 as the instrument of choice in such studies of children.<\/jats:p>","DOI":"10.1046\/j.1365-3016.2000.00250.x","type":"journal-article","created":{"date-parts":[[2003,3,12]],"date-time":"2003-03-12T17:43:28Z","timestamp":1047491008000},"page":"179-186","source":"Crossref","is-referenced-by-count":36,"title":["Evaluation of the Dinamap 8100 and Omron M1 blood pressure monitors for use in children"],"prefix":"10.1111","volume":"14","author":[{"given":"Mary E.","family":"Barker","sequence":"first","affiliation":[],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Alistair W.","family":"Shiell","sequence":"additional","affiliation":[],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Catherine M.","family":"Law","sequence":"additional","affiliation":[],"role":[{"vocabulary":"crossref","role":"author"}]}],"member":"311","published-online":{"date-parts":[[2001,12,25]]},"reference":[{"key":"e_1_2_6_2_2","doi-asserted-by":"publisher","DOI":"10.1016\/0140-6736(90)90878-9"},{"key":"e_1_2_6_3_2","doi-asserted-by":"publisher","DOI":"10.1146\/annurev.pu.12.050191.002511"},{"key":"e_1_2_6_4_2","doi-asserted-by":"publisher","DOI":"10.1097\/00004872-199306000-00013"},{"key":"e_1_2_6_5_2","doi-asserted-by":"publisher","DOI":"10.1097\/00004872-199007000-00005"},{"key":"e_1_2_6_6_2","doi-asserted-by":"publisher","DOI":"10.1136\/jech.46.2.164"},{"key":"e_1_2_6_7_2","first-page":"907","article-title":"Accuracy of blood pressure measurement by the Dinamap monitor in infants and children","volume":"79","author":"Park MK","year":"1987","journal-title":"Pediatrics"},{"key":"e_1_2_6_8_2","doi-asserted-by":"publisher","DOI":"10.1097\/00004872-199307000-00012"},{"key":"e_1_2_6_9_2","doi-asserted-by":"publisher","DOI":"10.1016\/S0140-6736(05)62926-1"},{"key":"e_1_2_6_10_2","first-page":"517","article-title":"Evaluation of the DINAMAP blood pressure monitor in an ambulatory primary care setting","volume":"26","author":"Ornstein S","year":"1988","journal-title":"Journal of Family Practice"},{"key":"e_1_2_6_11_2","unstructured":"BerensonGS VoorsAW WebberLS. Epidemiology in a bi\u2010racial community in the rural south of the United States. In:Hypertension in the Young and the Old.Editors: Onesti G Kim K. New York: Grune and Stratton 1981."},{"key":"e_1_2_6_12_2","first-page":"169","article-title":"Accuracy of an oscillometric automatic blood pressure device: the Omron HEM403C","volume":"9","author":"Walma EP","year":"1995","journal-title":"Journal of Human Hypertension"},{"key":"e_1_2_6_13_2","doi-asserted-by":"crossref","unstructured":"O\u2019BrienE MeeF AtkinsN.An accurate automated device for home blood pressure measurement at last! The Omron HEM\u2010705CP.Proceedings of the British Hypertension Society 1994.Royal College of Surgeons Dublin 19\u201321 September. Abs. no. 19. London: BHS.","DOI":"10.1097\/00004872-199411000-00032"},{"key":"e_1_2_6_14_2","unstructured":"British Standard.Non\u2010invasive sphygmomanometers.BSEN 10601997 Parts 1\u20133. London: British Standards Institute."},{"key":"e_1_2_6_15_2","unstructured":"Critikon. DINAMAP portable adult\/pediatric and neonatal vital signs monitor. Model 8100 operations manual. Florida USA: Critikon1988."},{"key":"e_1_2_6_16_2","doi-asserted-by":"publisher","DOI":"10.1136\/bmj.304.6818.23"},{"key":"e_1_2_6_17_2","doi-asserted-by":"publisher","DOI":"10.1097\/00004872-199608000-00002"},{"key":"e_1_2_6_18_2","first-page":"S11","article-title":"Differences between observers in blood pressure measurement with an automatic oscillometric recorder","volume":"8","author":"Bruce NG","year":"1990","journal-title":"Journal of Hypertension Supplement"},{"key":"e_1_2_6_19_2","doi-asserted-by":"publisher","DOI":"10.1097\/00004872-198805000-00006"},{"key":"e_1_2_6_20_2","unstructured":"BollingK.The Dinamap 8100 Calibration Study. London: HMSO 1994."}],"container-title":["Paediatric and Perinatal Epidemiology"],"original-title":[],"language":"en","link":[{"URL":"https:\/\/api.wiley.com\/onlinelibrary\/tdm\/v1\/articles\/10.1046%2Fj.1365-3016.2000.00250.x","content-type":"unspecified","content-version":"vor","intended-application":"text-mining"},{"URL":"https:\/\/onlinelibrary.wiley.com\/doi\/pdf\/10.1046\/j.1365-3016.2000.00250.x","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2023,10,29]],"date-time":"2023-10-29T11:01:42Z","timestamp":1698577302000},"score":1,"resource":{"primary":{"URL":"https:\/\/onlinelibrary.wiley.com\/doi\/10.1046\/j.1365-3016.2000.00250.x"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[2000,4]]},"references-count":19,"journal-issue":{"issue":"2","published-print":{"date-parts":[[2000,4]]}},"alternative-id":["10.1046\/j.1365-3016.2000.00250.x"],"URL":"https:\/\/doi.org\/10.1046\/j.1365-3016.2000.00250.x","archive":["Portico"],"relation":{},"ISSN":["0269-5022","1365-3016"],"issn-type":[{"value":"0269-5022","type":"print"},{"value":"1365-3016","type":"electronic"}],"subject":[],"published":{"date-parts":[[2000,4]]}}}