{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2024,8,5]],"date-time":"2024-08-05T21:33:22Z","timestamp":1722893602502},"reference-count":0,"publisher":"Wiley","issue":"8","license":[{"start":{"date-parts":[[2003,7,22]],"date-time":"2003-07-22T00:00:00Z","timestamp":1058832000000},"content-version":"vor","delay-in-days":720,"URL":"http:\/\/onlinelibrary.wiley.com\/termsAndConditions#vor"}],"content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":["Pacing Clinical Electrophis"],"published-print":{"date-parts":[[2001,8]]},"abstract":"<jats:p> <jats:bold>SCHUCHERT, A., <jats:sc>et al.<\/jats:sc>: A Randomized Study on the Effects of Pacemaker Programming to a Lower Output on Projected Pulse Generator Longevity.<\/jats:bold> The programmability of cardiac pacemakers enables the physician at follow\u2010up to adjust the pacing pulse under consideration of the 100% safety margin with respect to the individual pacing threshold. The purpose of reducing the output is to prolong pacemaker longevity. The aims of this prospective, randomized trial were to compare the effects of nominal output versus a lower output on projected pacemaker longevity in single and dual chamber pacemakers. The secondary aim was to assess how many patients can be programmed to <jats:styled-content>2.5 V\/0.4 ms<\/jats:styled-content> instead of the nominal 3.5\u2010V setting with \u2265 100% safety margin. The patients received the same types of VVI or DDD pacemakers that were connected in the ventricle to the steroid\u2010eluting, high impedance pacing lead. At the 3\u2010month follow\u2010up, patients with ventricular pacing thresholds <jats:styled-content>\u2264 0.15 ms at 2.5\u2010V<\/jats:styled-content> pulse amplitude were randomized to 3.5 V or 2.5 V amplitude at 0.4\u2010ms pulse duration. Lead function and projected device longevity were assessed with the pacemaker's telemetry 6 and 12 months after implantation. Of patients implanted with a VVI pacemaker, at the 3\u2010month follow\u2010up, 3 patients had pacing thresholds <jats:styled-content>&gt; 0.15 ms at 2.5 V and 139<\/jats:styled-content> patients could be randomized. A reprogramming to a higher output was necessary in one patient. The mean percentage of ventricular pacing was about 40% throughout the study time. The programming to 2.5\u2010V output resulted in an insignificant increase of device longevity from <jats:styled-content>117.9 \u00b1 18.7<\/jats:styled-content> months in the nominal group to <jats:styled-content>123.7 \u00b1 11.9<\/jats:styled-content> months at the 12\u2010month follow\u2010up (<jats:styled-content>P = 0.16<\/jats:styled-content>). Of patients implanted with a DDD pacemaker, 166 patients underwent randomization. The mean percentage of ventricular pacing was 85% in the ventricle and 35% in the atrium. The 2.5\u2010V setting significantly prolonged pacemaker longevity from <jats:styled-content>98.1 \u00b1 21.3 to 112.0 \u00b1 13.6 months<\/jats:styled-content> (<jats:styled-content>P &lt; 0.0001<\/jats:styled-content>). In three (1%) patients a late increase of the pacing threshold was observed. Due to the low ventricular pacing thresholds, the <jats:styled-content>2.5\u2010V\/0.4\u2010ms<\/jats:styled-content> setting provided, 3 months after implantation, a \u2265 100% safety margin in 99% of the patients. Programming to a lower output slightly increased projected pacemaker longevity compared to the nominal 3.5\u2010V setting. Longevity increased for 5% in patients with single and for 14% in dual chamber pulse generators.<\/jats:p>","DOI":"10.1046\/j.1460-9592.2001.01234.x","type":"journal-article","created":{"date-parts":[[2003,7,22]],"date-time":"2003-07-22T13:50:59Z","timestamp":1058881859000},"page":"1234-1239","source":"Crossref","is-referenced-by-count":15,"title":["A Randomized Study on the Effects of Pacemaker Programming to a Lower Output on Projected Pulse Generator Longevity"],"prefix":"10.1111","volume":"24","author":[{"given":"ANDREAS","family":"SCHUCHERT","sequence":"first","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"THOMAS","family":"MEINERTZ","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"name":"LOW OUTPUT PROGRAMMING (LOP) INVESTIGATORS","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]}],"member":"311","published-online":{"date-parts":[[2003,7,22]]},"container-title":["Pacing and Clinical Electrophysiology"],"original-title":[],"language":"en","link":[{"URL":"https:\/\/api.wiley.com\/onlinelibrary\/tdm\/v1\/articles\/10.1046%2Fj.1460-9592.2001.01234.x","content-type":"unspecified","content-version":"vor","intended-application":"text-mining"},{"URL":"https:\/\/onlinelibrary.wiley.com\/doi\/pdf\/10.1046\/j.1460-9592.2001.01234.x","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2023,11,19]],"date-time":"2023-11-19T18:23:45Z","timestamp":1700418225000},"score":1,"resource":{"primary":{"URL":"https:\/\/onlinelibrary.wiley.com\/doi\/10.1046\/j.1460-9592.2001.01234.x"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[2001,8]]},"references-count":0,"journal-issue":{"issue":"8","published-print":{"date-parts":[[2001,8]]}},"alternative-id":["10.1046\/j.1460-9592.2001.01234.x"],"URL":"https:\/\/doi.org\/10.1046\/j.1460-9592.2001.01234.x","archive":["Portico"],"relation":{},"ISSN":["0147-8389","1540-8159"],"issn-type":[{"type":"print","value":"0147-8389"},{"type":"electronic","value":"1540-8159"}],"subject":[],"published":{"date-parts":[[2001,8]]}}}