{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2026,9,11]],"date-time":"2026-09-11T15:26:04Z","timestamp":1789140364403,"version":"build-2803163510"},"reference-count":14,"publisher":"Ovid Technologies (Wolters Kluwer Health)","issue":"6","content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":[],"abstract":"<jats:sec>\n            <jats:title>Objectives:<\/jats:title>\n            <jats:p>A new bone conduction transducer, the Radioear B-81, has been designed to be an improvement over the commonly used transducer, the Radioear B-71. Reference Equivalent Threshold Force Levels (RETFLs) were obtained with the new Radioear B-81.<\/jats:p>\n          <\/jats:sec>\n          <jats:sec>\n            <jats:title>Design:<\/jats:title>\n            <jats:p>Thresholds were obtained in accordance with ANSI-S3.6-2018 (Annex D) and participants were selected as prescribed in ISO 389.9-2009. Thresholds were obtained with automatic audiometry using circumaural earphones (Radioear DD450) and forehead placement of the bone vibrators.<\/jats:p>\n          <\/jats:sec>\n          <jats:sec>\n            <jats:title>Results:<\/jats:title>\n            <jats:p>Mean bone conduction thresholds obtained using the B-81 and B-71 bone oscillators for frequencies from 250 to 4000 Hz were not statistically different. RETFLs for the B-81 are identical to the values in ANSI S3.6-2018 for the B-71 bone vibrator. Air-bone gaps were observed for both transducers at low frequencies (250 and 500 Hz) due to occlusion effects produced by the circumaural earphone and at high frequencies (3000 and 4000 Hz), previously reported in several studies that used standard RETFLs. Test-retest differences for air conduction thresholds were analyzed and the results are presented in the Appendix A (Supplemental Digital Content 1, http:\/\/links.lww.com\/EANDH\/A639).<\/jats:p>\n          <\/jats:sec>\n          <jats:sec>\n            <jats:title>Conclusions:<\/jats:title>\n            <jats:p>RETFLs in ANSI S3.6-2018 and ISO 389.3-2016 are appropriate for use with the B-81 bone vibrator.<\/jats:p>\n          <\/jats:sec>","DOI":"10.1097\/aud.0000000000000873","type":"journal-article","created":{"date-parts":[[2020,3,24]],"date-time":"2020-03-24T04:34:23Z","timestamp":1585024463000},"page":"1775-1778","source":"Crossref","is-referenced-by-count":10,"title":["Comparison of Thresholds Obtained With the Radioear B-71 and B-81 Bone-Conduction Transducers"],"prefix":"10.1097","volume":"41","author":[{"given":"Aparna","family":"Rao","sequence":"first","affiliation":[{"name":"Department of Speech and Hearing Science, Arizona State University, Tempe, Arizona, USA"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Bryce","family":"Altus","sequence":"additional","affiliation":[{"name":"Department of Speech and Hearing Science, Arizona State University, Tempe, Arizona, USA"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Roselynn","family":"Schroeder","sequence":"additional","affiliation":[{"name":"Department of Speech and Hearing Science, Arizona State University, Tempe, Arizona, USA"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Robert H.","family":"Margolis","sequence":"additional","affiliation":[{"name":"Department of Speech and Hearing Science, Arizona State University, Tempe, Arizona, USA"},{"name":"Audiology Incorporated, Arden Hills, Minnesota, USA."}],"role":[{"vocabulary":"crossref","role":"author"}]}],"member":"276","published-online":{"date-parts":[[2020,11]]},"reference":[{"key":"R3-20230928","doi-asserted-by":"crossref","first-page":"EL96","DOI":"10.1121\/1.4885771","article-title":"Effect of bone-conduction harmonic distortions on hearing thresholds.","volume":"136","author":"Eichenauer","year":"2014","journal-title":"J Acoust Soc Am"},{"key":"R4-20230928","doi-asserted-by":"crossref","first-page":"e0195233","DOI":"10.1371\/journal.pone.0195233","article-title":"Influence of transducer types on bone conduction hearing thresholds.","volume":"13","author":"Fr\u00f6hlich","year":"2018","journal-title":"PLoS One"},{"key":"R5-20230928","doi-asserted-by":"crossref","first-page":"EL33","DOI":"10.1121\/1.4804944","article-title":"Acoustic method for calibration of audiometric bone vibrators. II. Harmonic distortion.","volume":"134","author":"Ginter","year":"2013","journal-title":"J Acoust Soc Am"},{"key":"R9-20230928","doi-asserted-by":"crossref","first-page":"334","DOI":"10.3109\/14992027.2014.980521","article-title":"Electro-acoustic performance of the new bone vibrator Radioear B81: a comparison with the conventional Radioear B71.","volume":"54","author":"Jansson","year":"2015","journal-title":"Int J Audiol"},{"key":"R10-20230928","doi-asserted-by":"crossref","first-page":"78","DOI":"10.3766\/jaaa.18.1.7","article-title":"Qualind: A method for assessing the accuracy of automated tests.","volume":"18","author":"Margolis","year":"2007","journal-title":"J Am Acad Audiol"},{"key":"R11-20230928","doi-asserted-by":"crossref","first-page":"185","DOI":"10.3109\/14992020903092608","article-title":"AMTAS: automated method for testing auditory sensitivity: validation studies.","volume":"49","author":"Margolis","year":"2010","journal-title":"Int J Audiol"},{"key":"R12-20230928","doi-asserted-by":"crossref","first-page":"440","DOI":"10.3109\/14992027.2011.575085","article-title":"AMTAS(\u00ae): automated method for testing auditory sensitivity: III. sensorineural hearing loss and air-bone gaps.","volume":"50","author":"Margolis","year":"2011","journal-title":"Int J Audiol"},{"key":"R13-20230928","doi-asserted-by":"crossref","first-page":"434","DOI":"10.3109\/14992027.2011.553206","article-title":"AMTAS(\u00ae): automated method for testing auditory sensitivity: II. air conduction audiograms in children and adults.","volume":"50","author":"Margolis","year":"2011","journal-title":"Int J Audiol"},{"key":"R14-20230928","doi-asserted-by":"crossref","first-page":"526","DOI":"10.3109\/14992027.2013.792437","article-title":"False air-bone gaps at 4\u2009kHz in listeners with normal hearing and sensorineural hearing loss.","volume":"52","author":"Margolis","year":"2013","journal-title":"Int J Audiol"},{"key":"R15-20230928","doi-asserted-by":"crossref","first-page":"796","DOI":"10.3109\/14992027.2015.1033656","article-title":"Distribution characteristics of normal pure-tone thresholds.","volume":"54","author":"Margolis","year":"2015","journal-title":"Int J Audiol"},{"key":"R16-20230928","doi-asserted-by":"crossref","first-page":"177","DOI":"10.1097\/AUD.0000000000000246","article-title":"Distribution Characteristics of Air-Bone Gaps: Evidence of Bias in Manual Audiometry.","volume":"37","author":"Margolis","year":"2016","journal-title":"Ear Hear"},{"issue":"04","key":"R17-20230928","doi-asserted-by":"crossref","first-page":"329","DOI":"10.1055\/s-0034-1390162","article-title":"Bone-Conduction Calibration.","volume":"35","author":"Margolis","year":"2014","journal-title":"Sem Hear"},{"key":"R18-20230928","first-page":"14","article-title":"False 4-k Hz air-bone gaps \u2013 A failure of standards.","volume":"31","author":"Margolis","year":"2019","journal-title":"Audiol Today"},{"key":"R20-20230928","doi-asserted-by":"crossref","first-page":"177","DOI":"10.1097\/AUD.0000000000000585","article-title":"Evaluation of Two Circumaural Earphones for Audiometry.","volume":"40","author":"Smull","year":"2019","journal-title":"Ear Hear"}],"container-title":["Ear &amp; Hearing"],"original-title":[],"language":"en","link":[{"URL":"https:\/\/journals.lww.com\/10.1097\/AUD.0000000000000873","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2023,9,28]],"date-time":"2023-09-28T01:02:49Z","timestamp":1695862969000},"score":1,"resource":{"primary":{"URL":"https:\/\/journals.lww.com\/10.1097\/AUD.0000000000000873"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[2020,3,18]]},"references-count":14,"journal-issue":{"issue":"6","published-online":{"date-parts":[[2020]]}},"URL":"https:\/\/doi.org\/10.1097\/aud.0000000000000873","relation":{},"ISSN":["1538-4667","0196-0202"],"issn-type":[{"value":"1538-4667","type":"electronic"},{"value":"0196-0202","type":"print"}],"subject":[],"published":{"date-parts":[[2020,3,18]]}}}