{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2026,3,19]],"date-time":"2026-03-19T00:47:30Z","timestamp":1773881250057,"version":"3.50.1"},"reference-count":26,"publisher":"Ovid Technologies (Wolters Kluwer Health)","issue":"6","content-domain":{"domain":["lww.com","ovid.com"],"crossmark-restriction":true},"short-container-title":[],"published-print":{"date-parts":[[2023,6]]},"abstract":"<jats:sec>\n            <jats:title>Objectives:<\/jats:title>\n            <jats:p>To explore the performance of headless screws with FiberWire Suture as a tension band and headless screws with a mini-cable tension band in patella fixation.<\/jats:p>\n          <\/jats:sec>\n          <jats:sec>\n            <jats:title>Methods:<\/jats:title>\n            <jats:p>A transverse osteotomy was created in 6 matched pairs of fresh-frozen cadaver knee joints. One knee was randomly assigned to receive fixation with headless screws plus a cable tension band while the other was fixed with headless screws plus a suture tension band. Using a servo-hydraulic material testing system, the specimens were first tested nondestructively under 20% of the reported mean failure load with a standard technique of cannulated screws with tension band wiring. The specimen was then loaded to 1000 N to test the construct's failure strength. All tests were run under displacement-control with loading threshold. A motion analysis system was used to track the interfragmentary motion to assess fixation stability.<\/jats:p>\n          <\/jats:sec>\n          <jats:sec>\n            <jats:title>Results:<\/jats:title>\n            <jats:p>In the nondestructive loading test, gap displacement under 150 N was 0.10 mm or less for 11 of 12 specimens, and the difference between the 2 groups was not statistically significant. In the destructive test, 3 of 12 specimens maintained reduction (gap &lt;2 mm) at the maximum load of 1000 N. Of the failed specimens, the mean strength was 648 \u00b1 185 N for suture and 784 \u00b1 228 N for cable.<\/jats:p>\n          <\/jats:sec>\n          <jats:sec>\n            <jats:title>Conclusions:<\/jats:title>\n            <jats:p>There was no significant difference in fixation strength or subfailure fragment displacement between the suture and cable tension band techniques when using headless screws.<\/jats:p>\n          <\/jats:sec>","DOI":"10.1097\/bot.0000000000002556","type":"journal-article","created":{"date-parts":[[2023,2,2]],"date-time":"2023-02-02T17:16:05Z","timestamp":1675358165000},"page":"e253-e257","update-policy":"https:\/\/doi.org\/10.1097\/lww.0000000000001000","source":"Crossref","is-referenced-by-count":5,"title":["Biomechanical Evaluation of Transverse Patella Fracture Fixation Using Headless Screws: A Comparison of Suture Versus Cable as a Tension Band"],"prefix":"10.1097","volume":"37","author":[{"given":"Justin D.","family":"Bric","sequence":"first","affiliation":[{"name":"Department of Orthopaedic Surgery, Medical College of Wisconsin, Milwaukee, WI; and"}],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Jill M.","family":"Martin","sequence":"additional","affiliation":[{"name":"Department of Orthopaedic Surgery, Medical College of Wisconsin, Milwaukee, WI; and"}],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Linda M.","family":"McGrady","sequence":"additional","affiliation":[{"name":"Department of Orthopaedic Surgery, Medical College of Wisconsin, Milwaukee, WI; and"},{"name":"Marquette University, Orthopaedic and Rehabilitation Engineering Center, Milwaukee, WI."}],"role":[{"role":"author","vocabulary":"crossref"}]},{"ORCID":"https:\/\/orcid.org\/0000-0001-8352-2477","authenticated-orcid":false,"given":"Mei","family":"Wang","sequence":"additional","affiliation":[{"name":"Department of Orthopaedic Surgery, Medical College of Wisconsin, Milwaukee, WI; and"},{"name":"Marquette University, Orthopaedic and Rehabilitation Engineering Center, Milwaukee, WI."}],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Gregory J.","family":"Schmeling","sequence":"additional","affiliation":[{"name":"Department of Orthopaedic Surgery, Medical College of Wisconsin, Milwaukee, WI; and"},{"name":"Marquette University, Orthopaedic and Rehabilitation Engineering Center, Milwaukee, WI."}],"role":[{"role":"author","vocabulary":"crossref"}]}],"member":"276","reference":[{"issue":"suppl 143","key":"R1-20250614","doi-asserted-by":"crossref","first-page":"1","DOI":"10.3109\/ort.1972.43.suppl-143.01","article-title":"Fracture of the patella. A study of 422 patellar fractures","volume":"43","author":"Bostrom","year":"1972","journal-title":"Acta Orthop Scand"},{"key":"R2-20250614","first-page":"421","article-title":"Basic kinematics and biomechanics of the patello-femoral joint. Part 1: the native patella","volume":"77","author":"Schindler","year":"2011","journal-title":"Acta Orthop Belg."},{"key":"R3-20250614","doi-asserted-by":"crossref","first-page":"3297","DOI":"10.1007\/s11999-009-1052-y","article-title":"Clinical anatomy of the quadriceps femoris and extensor apparatus of the knee","volume":"467","author":"Waligora","year":"2009","journal-title":"Clin Orthop Relat Res."},{"key":"R4-20250614","first-page":"51","article-title":"Patellar biomechanics","volume":"144","author":"Kaufer","year":"1979","journal-title":"Clin Orthop Relat Res"},{"key":"R6-20250614","doi-asserted-by":"crossref","first-page":"345","DOI":"10.1097\/BOT.0b013e31826623eb","article-title":"An alternative patellar fracture fixation: a biomechanical study","volume":"27","author":"Banks","year":"2013","journal-title":"J Orthop Trauma"},{"key":"R7-20250614","first-page":"9","article-title":"Biomechanics of the patellofemoral joint","author":"Hungerford","year":"1979","journal-title":"Clin Orthop Relat Res"},{"key":"R8-20250614","doi-asserted-by":"crossref","first-page":"17","DOI":"10.1016\/0021-9290(88)90187-X","article-title":"Experimental determination of forces transmitted through the patello-femoral joint","volume":"21","author":"Buff","year":"1988","journal-title":"J Biomechanics"},{"key":"R9-20250614","doi-asserted-by":"crossref","first-page":"126","DOI":"10.3109\/17453677208991251","article-title":"Experimental analysis of the quadriceps muscle force and patello-femoral joint reaction force for various activities","volume":"43","author":"Reilly","year":"1972","journal-title":"Acta Orthop Scand"},{"key":"R10-20250614","doi-asserted-by":"crossref","first-page":"198","DOI":"10.5435\/00124635-201104000-00004","article-title":"Patellar fractures in adults","volume":"19","author":"Melvin","year":"2011","journal-title":"Am Acad Orthop Surg"},{"key":"R11-20250614","doi-asserted-by":"crossref","first-page":"e4992","DOI":"10.1097\/MD.0000000000004992","article-title":"Clinical results of treatment using a modified K-wire tension band versus a cannulated screw tension band in transverse patella fractures: a strobe-compliant retrospective observational study","volume":"95","author":"Tan","year":"2016","journal-title":"Med Baltim"},{"key":"R12-20250614","doi-asserted-by":"crossref","first-page":"1051","DOI":"10.1097\/TA.0b013e3182405296","article-title":"Factors predicting failure of patella fixation","volume":"72","author":"Miller","year":"2012","journal-title":"J Trauma Acute Care Surg"},{"key":"R13-20250614","doi-asserted-by":"crossref","first-page":"351","DOI":"10.1097\/00005131-199707000-00009","article-title":"Biomechanical evaluation of current patella fracture fixation techniques","volume":"11","author":"Carpenter","year":"1997","journal-title":"J Orthop Trauma"},{"key":"R14-20250614","doi-asserted-by":"crossref","first-page":"573","DOI":"10.1097\/00005131-199711000-00005","article-title":"Open reduction internal fixation of displaced transverse patella fractures with figure-eight wiring through parallel cannulated compression screws","volume":"11","author":"Berg","year":"1997","journal-title":"J Orthop Trauma"},{"key":"R15-20250614","doi-asserted-by":"crossref","first-page":"301","DOI":"10.3928\/01477447-20100329-14","article-title":"Implant removal following surgical stabilization of patella fracture","volume":"33","author":"Kumar","year":"2010","journal-title":"Orthopedics"},{"key":"R16-20250614","doi-asserted-by":"crossref","first-page":"928","DOI":"10.1097\/TA.0b013e31825168b6","article-title":"Meta-analysis of re-operation, nonunion, and infection after open reduction and internal fixation of patella fractures","volume":"73","author":"Dy","year":"2012","journal-title":"J Trauma Acute Care Surg"},{"key":"R17-20250614","doi-asserted-by":"crossref","first-page":"183","DOI":"10.1097\/00005131-199704000-00008","article-title":"Early complications in the operative treatment of patella fractures","volume":"11","author":"Smith","year":"1997","journal-title":"J Orthop Trauma"},{"key":"R18-20250614","doi-asserted-by":"crossref","first-page":"439","DOI":"10.1097\/00005373-199509000-00007","article-title":"Open fractures of the patella: long-term functional outcome","volume":"39","author":"Catalano","year":"1995","journal-title":"J Trauma"},{"key":"R19-20250614","doi-asserted-by":"crossref","first-page":"e240","DOI":"10.1097\/BOT.0000000000001447","article-title":"Biomechanical comparison of tension band fixation of patella transverse fracture: headless screws versus headed screws","volume":"33","author":"Martin","year":"2019","journal-title":"J Orthop Trauma"},{"key":"R20-20250614","doi-asserted-by":"crossref","first-page":"1200","DOI":"10.1016\/j.injury.2009.04.011","article-title":"FiberWire is superior in strength to stainless steel wire for tension band fixation of transverse patellar fractures","volume":"40","author":"Wright","year":"2009","journal-title":"Injury"},{"key":"R21-20250614","doi-asserted-by":"crossref","first-page":"372","DOI":"10.1002\/jor.1100040314","article-title":"Patellar surface strain","volume":"4","author":"Goldstein","year":"1986","journal-title":"J Orthop Res."},{"key":"R22-20250614","doi-asserted-by":"crossref","first-page":"1290","DOI":"10.1016\/j.injury.2012.04.020","article-title":"Biomechanical cadaver testing of a fixed-angle plate in comparison to tension wiring and screw fixation in transverse patella fractures","volume":"43","author":"Thelen","year":"2012","journal-title":"Injury"},{"key":"R23-20250614","doi-asserted-by":"crossref","first-page":"1749","DOI":"10.1016\/j.injury.2017.05.037","article-title":"Biomechanical evaluation of the tension band wiring principle. A comparison between two different techniques for transverse patella fracture fixation","volume":"48","author":"Zderic","year":"2017","journal-title":"Injury"},{"key":"R24-20250614","doi-asserted-by":"crossref","first-page":"99","DOI":"10.1186\/s12891-019-2474-7","article-title":"Roles of the screw types, proximity and anterior band wiring in the surgical fixation of transverse patellar fractures: a finite element investigation","volume":"20","author":"Chen","year":"2019","journal-title":"BMC Musculoskelet Disord."},{"key":"R25-20250614","doi-asserted-by":"crossref","first-page":"156","DOI":"10.1016\/j.injury.2009.07.007","article-title":"Biomechanical comparison of tension band- and interfragmentary screw fixation with a new implant in transverse patella fractures","volume":"41","author":"Dargel","year":"2010","journal-title":"Injury"},{"key":"R26-20250614","doi-asserted-by":"crossref","first-page":"1249","DOI":"10.1097\/CORR.0000000000000683","article-title":"Cannulated screw prominence in tension band wiring of patella fractures increases fracture gapping: a cadaver study","volume":"477","author":"Avery","year":"2019","journal-title":"Clin Orthop Relat Res."},{"key":"R27-20250614","first-page":"84","article-title":"Material properties of common suture materials in orthopaedic surgery","volume":"30","author":"Najibi","year":"2010","journal-title":"Iowa Orthop J."}],"container-title":["Journal of Orthopaedic Trauma"],"original-title":[],"language":"en","link":[{"URL":"https:\/\/journals.lww.com\/10.1097\/BOT.0000000000002556","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2025,6,14]],"date-time":"2025-06-14T18:03:43Z","timestamp":1749924223000},"score":1,"resource":{"primary":{"URL":"https:\/\/journals.lww.com\/10.1097\/BOT.0000000000002556"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[2023,6]]},"references-count":26,"journal-issue":{"issue":"6","published-print":{"date-parts":[[2023]]}},"URL":"https:\/\/doi.org\/10.1097\/bot.0000000000002556","relation":{},"ISSN":["0890-5339"],"issn-type":[{"value":"0890-5339","type":"print"}],"subject":[],"published":{"date-parts":[[2023,6]]},"assertion":[{"value":"2022-12-19","name":"accepted","label":"Accepted","group":{"name":"publication_history","label":"Publication History"}}]}}