{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2026,2,10]],"date-time":"2026-02-10T06:44:34Z","timestamp":1770705874667,"version":"3.49.0"},"reference-count":38,"publisher":"Ovid Technologies (Wolters Kluwer Health)","issue":"6","content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":[],"published-print":{"date-parts":[[2018,11]]},"abstract":"<jats:p>Relative blood volume (RBV) monitoring during hemodialysis has been used to help guide fluid management for decades, although with little supporting evidence. The technique relies on the assumption that variation in RBV during fluid removal reflects the capacity for vascular refilling and that efficient refilling is related to fluid overload. This study investigated the relationship between RBV variation and bioimpedance-based fluid overload in 47 patients on stable hemodialysis. Mean treatment ultrafiltration volume (UFV) was 1.7\u2009L and RBV reduction was 3.2%\/hour. Relative blood volume slopes were grouped based on trajectory: flatline (no decrease), linear decrease, or linear decrease followed by flatline. Fluid overload was similar (<jats:italic toggle=\"yes\">p<\/jats:italic> &gt; 0.05) across groups pre-dialysis (1.0, 2.2, and 1.6\u2009L, respectively) and post-dialysis (\u22120.8, \u22120.1, and \u22120.1\u2009L), whereas UFV was higher in patients with a linear decrease (1.8, 2.5, and 1.6\u2009L; <jats:italic toggle=\"yes\">p<\/jats:italic> = 0.02). Specific ultrafiltration rate, but not fluid overload, was associated with RBV change over dialysis. At least half the patients in each group finished dialysis fluid depleted based on bioimpedance, suggesting that the link between refilling and fluid overload is not as straightforward as previously assumed. These results question the assumptions that the absence of an appreciable decrease in RBV indicates fluid overload, and a rapid fall suggests fluid depletion.<\/jats:p>","DOI":"10.1097\/mat.0000000000000795","type":"journal-article","created":{"date-parts":[[2018,4,20]],"date-time":"2018-04-20T22:32:55Z","timestamp":1524263575000},"page":"812-818","source":"Crossref","is-referenced-by-count":16,"title":["Time to Reconsider the Role of Relative Blood Volume Monitoring for Fluid Management in Hemodialysis"],"prefix":"10.1097","volume":"64","author":[{"given":"David F.","family":"Keane","sequence":"first","affiliation":[{"name":"Department of Renal Medicine, Leeds Teaching Hospitals NHS Trust, Leeds, United Kingdom"},{"name":"NIHR Devices for Dignity Healthcare Technology Co-Operative, Sheffield, United Kingdom"},{"name":"Leeds Institute for Cardiovascular and Metabolic Medicine, University of Leeds, Leeds, United Kingdom"}],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Paul","family":"Baxter","sequence":"additional","affiliation":[{"name":"Leeds Institute for Cardiovascular and Metabolic Medicine, University of Leeds, Leeds, United Kingdom"}],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Elizabeth","family":"Lindley","sequence":"additional","affiliation":[{"name":"Department of Renal Medicine, Leeds Teaching Hospitals NHS Trust, Leeds, United Kingdom"},{"name":"NIHR Devices for Dignity Healthcare Technology Co-Operative, Sheffield, United Kingdom"}],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Laura","family":"Rhodes","sequence":"additional","affiliation":[{"name":"Leeds Institute for Cardiovascular and Metabolic Medicine, University of Leeds, Leeds, United Kingdom"}],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Sue","family":"Pavitt","sequence":"additional","affiliation":[{"name":"School of Dentistry, University of Leeds, Leeds, United Kingdom."}],"role":[{"role":"author","vocabulary":"crossref"}]}],"member":"276","reference":[{"key":"R1-20240126","doi-asserted-by":"crossref","first-page":"1474","DOI":"10.2215\/CJN.06930613","article-title":"Time to improve fluid management in hemodialysis: Should we abandon clinical assessment and routinely use bioimpedance?","volume":"8","author":"Covic","year":"2013","journal-title":"Clin J Am Soc Nephrol"},{"key":"R2-20240126","doi-asserted-by":"crossref","first-page":"342","DOI":"10.1038\/ki.1990.207","article-title":"A blood protein monitor for the continuous measurement of blood volume changes during hemodialysis.","volume":"38","author":"Schneditz","year":"1990","journal-title":"Kidney Int"},{"key":"R3-20240126","doi-asserted-by":"crossref","first-page":"2043","DOI":"10.1093\/ndt\/gfi056","article-title":"Measurement of relative blood volume changes during haemodialysis: Merits and limitations.","volume":"20","author":"Dasselaar","year":"2005","journal-title":"Nephrol Dial Transplant"},{"key":"R4-20240126","doi-asserted-by":"crossref","first-page":"120","DOI":"10.1111\/hdi.12338","article-title":"Concordance of absolute and relative plasma volume changes and stability of cells in routine hemodialysis.","volume":"20","author":"Schneditz","year":"2016","journal-title":"Hemodial Int"},{"key":"R5-20240126","doi-asserted-by":"crossref","first-page":"1034","DOI":"10.1046\/j.1523-1755.2002.00511.x","article-title":"Blood volume controlled hemodialysis in hypotension-prone patients: A randomised, multicentre controlled trial.","volume":"62","author":"Santoto","year":"2002","journal-title":"Kidney Int"},{"key":"R6-20240126","doi-asserted-by":"crossref","first-page":"52","DOI":"10.1111\/j.1755-6686.2007.tb00040.x","article-title":"Continuous blood volume monitoring and \u201cdry weight\u201d assessment.","volume":"33","author":"Lopot","year":"2007","journal-title":"J Ren Care"},{"key":"R7-20240126","doi-asserted-by":"crossref","first-page":"2162","DOI":"10.1681\/ASN.2004121053","article-title":"Intradialytic blood volume monitoring in ambulatory hemodialysis patients: A randomized trial.","volume":"16","author":"Reddan","year":"2005","journal-title":"J Am Soc Nephrol"},{"key":"R8-20240126","doi-asserted-by":"crossref","first-page":"80","DOI":"10.1093\/ajcn\/85.1.80","article-title":"A whole-body model to distinguish excess fluid from the hydration of major body tissues.","volume":"85","author":"Chamney","year":"2007","journal-title":"Am J Clin Nutr"},{"key":"R9-20240126","doi-asserted-by":"crossref","first-page":"75","DOI":"10.1159\/000167013","article-title":"Importance of whole-body bioimpedance spectroscopy for the management of fluid balance.","volume":"27","author":"Wabel","year":"2009","journal-title":"Blood Purif"},{"key":"R10-20240126","doi-asserted-by":"crossref","first-page":"1574","DOI":"10.1093\/ndt\/gfn707","article-title":"The mortality risk of overhydration in haemodialysis patients.","volume":"24","author":"Wizemann","year":"2009","journal-title":"Nephrol Dial Transplant"},{"key":"R11-20240126","first-page":"3","article-title":"The body composition monitor: A flexible tool for routine fluid management across the haemodialysis population.","volume":"25","author":"Keane","year":"2017","journal-title":"Biomed Phys Eng Express"},{"key":"R12-20240126","doi-asserted-by":"crossref","first-page":"305","DOI":"10.1161\/HYPERTENSIONAHA.109.143974","article-title":"Relative plasma volume monitoring during hemodialysis AIDS the assessment of dry weight.","volume":"55","author":"Sinha","year":"2010","journal-title":"Hypertension"},{"issue":"suppl 6","key":"R14-20240126","first-page":"137","article-title":"Reproducibility of bioimpedance spectroscopy (BIS) in health and disease (abstract).","volume":"22","author":"Wabel","year":"2007","journal-title":"Nephrol Dial Transplant"},{"key":"R15-20240126","doi-asserted-by":"crossref","first-page":"1214","DOI":"10.1016\/j.kint.2016.12.008","article-title":"Impact of fluid status and inflammation and their interaction on survival: A study in an international hemodialysis patient cohort.","volume":"91","author":"Dekker","year":"2017","journal-title":"Kidney Int"},{"key":"R16-20240126","doi-asserted-by":"crossref","first-page":"202","DOI":"10.1159\/000346630","article-title":"Sensitivity of blood volume monitoring for fluid status assessment in hemodialysis patients.","volume":"35","author":"Maduell","year":"2013","journal-title":"Blood Purif"},{"key":"R17-20240126","doi-asserted-by":"crossref","first-page":"627","DOI":"10.1046\/j.1525-1594.1998.06036.x","article-title":"Enhanced fluid removal guided by blood volume monitoring during chronic hemodialysis.","volume":"22","author":"Steuer","year":"1998","journal-title":"Artif Organs"},{"key":"R18-20240126","doi-asserted-by":"crossref","first-page":"854","DOI":"10.1111\/j.1523-1755.2005.00467.x","article-title":"Assessment of dry weight by monitoring changes in blood volume during hemodialysis using Crit-Line.","volume":"68","author":"Rodriguez","year":"2005","journal-title":"Kidney Int"},{"key":"R19-20240126","doi-asserted-by":"crossref","first-page":"463","DOI":"10.1097\/01.ASN.0000108970.48370.33","article-title":"The relationship between systemic and whole-body hematocrit is not constant during ultrafiltration on hemodialysis.","volume":"15","author":"Mitra","year":"2004","journal-title":"J Am Soc Nephrol"},{"key":"R20-20240126","doi-asserted-by":"crossref","first-page":"525","DOI":"10.1016\/S0272-6386(96)90163-8","article-title":"Reducing symptoms during hemodialysis by continuously monitoring the hematocrit.","volume":"27","author":"Steuer","year":"1996","journal-title":"Am J Kidney Dis"},{"key":"R21-20240126","doi-asserted-by":"crossref","first-page":"556","DOI":"10.1053\/ajkd.2002.34914","article-title":"Linear decay of relative blood volume during ultrafiltration predicts hemodynamic instability.","volume":"40","author":"Mitra","year":"2002","journal-title":"Am J Kidney Dis"},{"issue":"suppl 6","key":"R22-20240126","doi-asserted-by":"crossref","first-page":"55","DOI":"10.1093\/ndt\/13.suppl_6.55","article-title":"Ultrafiltration behaviour with different dialysis schedules.","volume":"13","author":"Santoro","year":"1998","journal-title":"Nephrol Dial Transplant"},{"key":"R23-20240126","doi-asserted-by":"crossref","first-page":"1098","DOI":"10.1177\/039139880703001210","article-title":"Integration of blood volume, blood pressure, heart rate and bioimpedance monitoring for the achievement of optimal dry body weight during chronic hemodialysis.","volume":"30","author":"Bonello","year":"2007","journal-title":"Artif Organs"},{"key":"R24-20240126","doi-asserted-by":"crossref","first-page":"813","DOI":"10.1007\/s00467-007-0724-7","article-title":"Blood-volume monitoring in paediatric haemodialysis.","volume":"23","author":"Hothi","year":"2008","journal-title":"Pediatr Nephrol"},{"key":"R25-20240126","doi-asserted-by":"crossref","first-page":"8","DOI":"10.1111\/hdi.2000.4.1.8","article-title":"Continuous blood volume monitoring and ultrafiltration control.","volume":"4","author":"Lopot","year":"2000","journal-title":"Hemodial Int"},{"key":"R26-20240126","doi-asserted-by":"crossref","first-page":"1419","DOI":"10.1007\/s00467-009-1150-9","article-title":"Importance of the curve shape for interpretation of blood volume monitor changes during haemodiafiltration.","volume":"24","author":"Dheu","year":"2009","journal-title":"Pediatr Nephrol"},{"key":"R27-20240126","first-page":"224","article-title":"Prediction of blood volume in normal human adults.","volume":"51","author":"Nadler","year":"1962","journal-title":"Surgery"},{"key":"R28-20240126","doi-asserted-by":"crossref","first-page":"19","DOI":"10.1159\/000189276","article-title":"A study on regulating factors of plasma refilling during hemodialysis.","volume":"74","author":"Iimura","year":"1996","journal-title":"Nephron"},{"key":"R29-20240126","doi-asserted-by":"crossref","first-page":"848","DOI":"10.1038\/ki.1984.227","article-title":"Plasma volume recovery after ultrafiltration in patients with chronic renal failure.","volume":"26","author":"Koomans","year":"1984","journal-title":"Kidney Int"},{"key":"R30-20240126","doi-asserted-by":"crossref","first-page":"1244","DOI":"10.1053\/ajkd.2002.36894","article-title":"The role of blood volume reduction in the genesis of intradialytic hypotension.","volume":"40","author":"Andrulli","year":"2002","journal-title":"Am J Kidney Dis"},{"key":"R31-20240126","doi-asserted-by":"crossref","first-page":"1353","DOI":"10.1093\/ndt\/gfg171","article-title":"Characteristics of hypotension-prone haemodialysis patients: Is there a critical relative blood volume?","volume":"18","author":"Barth","year":"2003","journal-title":"Nephrol Dial Transplant"},{"key":"R32-20240126","doi-asserted-by":"crossref","first-page":"673","DOI":"10.1093\/ndt\/15.5.673","article-title":"Variability of relative blood volume during haemodialysis.","volume":"15","author":"Krepel","year":"2000","journal-title":"Nephrol Dial Transplant"},{"key":"R33-20240126","first-page":"232","article-title":"Improved intradialytic stability during haemodialysis with blood volume-controlled ultrafiltration.","volume":"22","author":"Gabrielli","year":"2009","journal-title":"J Nephrol"},{"key":"R34-20240126","doi-asserted-by":"crossref","first-page":"1831","DOI":"10.2215\/CJN.01030117","article-title":"Randomized crossover trial of blood volume monitoring-guided ultrafiltration biofeedback to reduce intradialytic hypotensive episodes with hemodialysis.","volume":"12","author":"Leung","year":"2017","journal-title":"Clin J Am Soc Nephrol"},{"key":"R35-20240126","doi-asserted-by":"crossref","first-page":"238","DOI":"10.1186\/s12882-017-0639-x","article-title":"Blood volume-monitored regulation of ultrafiltration to decrease the dry weight in fluid-overloaded hemodialysis patients: A randomized controlled trial.","volume":"18","author":"Antlanger","year":"2017","journal-title":"BMC Nephrol"},{"key":"R36-20240126","doi-asserted-by":"crossref","first-page":"436","DOI":"10.1097\/MAT.0000000000000086","article-title":"On-line dialysate infusion to estimate absolute blood volume in dialysis patients.","volume":"60","author":"Schneditz","year":"2014","journal-title":"ASAIO J"},{"key":"R37-20240126","doi-asserted-by":"crossref","first-page":"484","DOI":"10.1111\/hdi.12417","article-title":"Vascular refilling is independent of volume overload in hemodialysis with moderate ultrafiltration requirements.","volume":"20","author":"Kron","year":"2016","journal-title":"Hemodial Int"},{"key":"R38-20240126","doi-asserted-by":"crossref","first-page":"254","DOI":"10.1111\/hdi.12582","article-title":"Adjustment of target weight based on absolute blood volume reduces the frequency of intradialytic morbid events.","volume":"22","author":"Kron","year":"2018","journal-title":"Hemodial Int"},{"key":"R39-20240126","doi-asserted-by":"crossref","first-page":"957","DOI":"10.1053\/j.ajkd.2012.12.017","article-title":"Effect of fluid management guided by bioimpedance spectroscopy on cardiovascular parameters in hemodialysis patients: A randomized controlled trial.","volume":"61","author":"Hur","year":"2013","journal-title":"Am J Kidney Dis"}],"container-title":["ASAIO Journal"],"original-title":[],"language":"en","link":[{"URL":"https:\/\/journals.lww.com\/00002480-201811000-00020","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2024,1,26]],"date-time":"2024-01-26T17:01:18Z","timestamp":1706288478000},"score":1,"resource":{"primary":{"URL":"https:\/\/journals.lww.com\/00002480-201811000-00020"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[2018,11]]},"references-count":38,"journal-issue":{"issue":"6","published-print":{"date-parts":[[2018]]}},"URL":"https:\/\/doi.org\/10.1097\/mat.0000000000000795","relation":{},"ISSN":["1058-2916"],"issn-type":[{"value":"1058-2916","type":"print"}],"subject":[],"published":{"date-parts":[[2018,11]]}}}