{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2026,2,16]],"date-time":"2026-02-16T17:38:59Z","timestamp":1771263539868,"version":"3.50.1"},"reference-count":13,"publisher":"Ovid Technologies (Wolters Kluwer Health)","issue":"9","content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":[],"published-print":{"date-parts":[[2018,9]]},"abstract":"<jats:sec>\n            <jats:title>Objectives:<\/jats:title>\n            <jats:p>To investigate how compensating for endotracheal tube leaks by targeting the leak-compensated tidal volume affects measured physiologic and ventilator variables during neonatal mechanical ventilation.<\/jats:p>\n          <\/jats:sec>\n          <jats:sec>\n            <jats:title>Design:<\/jats:title>\n            <jats:p>Retrospective observational study.<\/jats:p>\n          <\/jats:sec>\n          <jats:sec>\n            <jats:title>Setting:<\/jats:title>\n            <jats:p>A level III Neonatal ICU.<\/jats:p>\n          <\/jats:sec>\n          <jats:sec>\n            <jats:title>Patients:<\/jats:title>\n            <jats:p>We enrolled 30 neonates who were ventilated using synchronized intermittent positive pressure mode with volume guarantee and had at least 12 hours of continuous detailed recording of ventilation variables.<\/jats:p>\n          <\/jats:sec>\n          <jats:sec>\n            <jats:title>Interventions:<\/jats:title>\n            <jats:p>Infants were treated using the Dr\u00e4ger VN500 ventilator (Dr\u00e4ger, L\u00fcbeck, Germany), which uses a proprietary algorithm to measure and compensate for endotracheal tube leaks. Eleven were ventilated without leak compensation and 19 with leak compensation.<\/jats:p>\n          <\/jats:sec>\n          <jats:sec>\n            <jats:title>Measurements and Main Results:<\/jats:title>\n            <jats:p>Detailed ventilation data were collected and analyzed at 1 Hz, with intermittent blood gas values. The percentage of leak was less than 20% in 73% of leak-compensated inflations, and the volume of the leak compensation was less than 1\u2009mL\/kg in 97.3% of inflations. Between the two groups, ventilation variables were comparable, except the percentage of leak that was significantly (<jats:italic toggle=\"yes\">p<\/jats:italic> = 0.005) higher in the recordings with leak compensation. Without leak compensation, the mean expired tidal volume was maintained very close to the set level up to 50% leak, but with leaks greater than 50%, it declined progressively. With leak compensation, the mean leak-compensated expired tidal volume was well maintained even with leak greater than 90% although with large variability. Without leak compensation, the difference between the maximum allowed inflating pressure and the peak inflating pressure decreased progressively as the leak increased. This did not occur with leak compensation. The median P<jats:sc>co<\/jats:sc>\n                                 <jats:sub>2<\/jats:sub> was slightly higher with leak compensation.<\/jats:p>\n          <\/jats:sec>\n          <jats:sec>\n            <jats:title>Conclusions:<\/jats:title>\n            <jats:p>During volume guarantee ventilation with a Dr\u00e4ger VN500 ventilator, without leak compensation the expired tidal volume declined after 50% leak. With leak compensation, the tidal volume was maintained even with a large leak. With leak compensation, there was a more stable peak inflating pressure, although the P<jats:sc>co<\/jats:sc>\n                                 <jats:sub>2<\/jats:sub> was slightly higher.<\/jats:p>\n          <\/jats:sec>","DOI":"10.1097\/pcc.0000000000001638","type":"journal-article","created":{"date-parts":[[2018,6,22]],"date-time":"2018-06-22T22:42:16Z","timestamp":1529707336000},"page":"861-868","source":"Crossref","is-referenced-by-count":18,"title":["Leak Compensation During Volume Guarantee With the Dr\u00e4ger Babylog VN500 Neonatal Ventilator*"],"prefix":"10.1097","volume":"19","author":[{"given":"Eniko","family":"Szakmar","sequence":"first","affiliation":[{"name":"Neonatal Intensive Care Unit, The Rosie Hospital, Cambridge University Hospitals NHS Foundation Trust, Cambridge, United Kingdom."},{"name":"1st Department of Paediatrics, Semmelweis University, Budapest, Hungary."}],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Colin J.","family":"Morley","sequence":"additional","affiliation":[{"name":"Neonatal Intensive Care Unit, The Rosie Hospital, Cambridge University Hospitals NHS Foundation Trust, Cambridge, United Kingdom."}],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Gusztav","family":"Belteki","sequence":"additional","affiliation":[{"name":"Neonatal Intensive Care Unit, The Rosie Hospital, Cambridge University Hospitals NHS Foundation Trust, Cambridge, United Kingdom."}],"role":[{"role":"author","vocabulary":"crossref"}]}],"member":"276","reference":[{"key":"R1-20230823","doi-asserted-by":"crossref","first-page":"575","DOI":"10.1038\/jp.2011.98","article-title":"A practical guide to neonatal volume guarantee ventilation.","volume":"31","author":"Klingenberg","year":"2011","journal-title":"J Perinatol"},{"key":"R2-20230823","doi-asserted-by":"crossref","first-page":"294","DOI":"10.1164\/ajrccm.157.1.9604014","article-title":"Ventilator-induced lung injury: Lessons from experimental studies.","volume":"157","author":"Dreyfuss","year":"1998","journal-title":"Am J Respir Crit Care Med"},{"key":"R3-20230823","first-page":"CD003666","article-title":"Volume-targeted versus pressure-limited ventilation in neonates.","volume":"10","author":"Klingenberg","year":"2017","journal-title":"Cochrane Database Syst Rev"},{"key":"R4-20230823","doi-asserted-by":"crossref","first-page":"107","DOI":"10.1016\/j.clp.2006.12.011","article-title":"Volume guarantee ventilation.","volume":"34","author":"Keszler","year":"2007","journal-title":"Clin Perinatol"},{"key":"R5-20230823","article-title":"Frequency, duration and cause of ventilator alarms on a neonatal intensive care unit.","author":"Belteki","year":"2017","journal-title":"Arch Dis Child Fetal Neonatal Ed"},{"key":"R6-20230823","doi-asserted-by":"crossref","first-page":"10","DOI":"10.4187\/respcare.05012","article-title":"Performance of leak compensation in all-age ICU ventilators during volume-targeted neonatal ventilation: A lung model study.","volume":"62","author":"Itagaki","year":"2017","journal-title":"Respir Care"},{"key":"R7-20230823","doi-asserted-by":"crossref","first-page":"F120","DOI":"10.1136\/archdischild-2017-312640","article-title":"Volume guaranteed? Accuracy of a volume-targeted ventilation mode in infants.","volume":"103","author":"Farrell","year":"2018","journal-title":"Arch Dis Child Fetal Neonatal Ed"},{"key":"R8-20230823","doi-asserted-by":"crossref","first-page":"348","DOI":"10.4187\/respcare.01839","article-title":"A rational framework for selecting modes of ventilation.","volume":"58","author":"Mireles-Cabodevila","year":"2013","journal-title":"Respir Care"},{"key":"R9-20230823","doi-asserted-by":"crossref","first-page":"F279","DOI":"10.1136\/adc.2008.147157","article-title":"Evolution of tidal volume requirement during the first 3 weeks of life in infants <800\u2009g ventilated with volume guarantee.","volume":"94","author":"Keszler","year":"2009","journal-title":"Arch Dis Child Fetal Neonatal Ed"},{"key":"R10-20230823","doi-asserted-by":"crossref","first-page":"357","DOI":"10.1002\/ppul.20363","article-title":"Lung inflammation in preterm infants with respiratory distress syndrome: Effects of ventilation with different tidal volumes.","volume":"41","author":"Lista","year":"2006","journal-title":"Pediatr Pulmonol"},{"key":"R11-20230823","doi-asserted-by":"crossref","first-page":"326","DOI":"10.1007\/s00134-006-0450-9","article-title":"Volume-targeted modes of modern neonatal ventilators: How stable is the delivered tidal volume?","volume":"33","author":"Jaecklin","year":"2007","journal-title":"Intensive Care Med"},{"key":"R12-20230823","doi-asserted-by":"crossref","first-page":"22","DOI":"10.4187\/respcare.04825","article-title":"Effects of leak compensation on patient-ventilator synchrony during premature\/neonatal invasive and noninvasive ventilation: A lung model study.","volume":"62","author":"Itagaki","year":"2017","journal-title":"Respir Care"},{"key":"R13-20230823","doi-asserted-by":"crossref","first-page":"2027","DOI":"10.4187\/respcare.02466","article-title":"A comparison of leak compensation in acute care ventilators during noninvasive and invasive ventilation: A lung model study.","volume":"58","author":"Oto","year":"2013","journal-title":"Respir Care"}],"container-title":["Pediatric Critical Care Medicine"],"original-title":[],"language":"en","link":[{"URL":"https:\/\/journals.lww.com\/00130478-201809000-00010","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2023,8,24]],"date-time":"2023-08-24T00:36:04Z","timestamp":1692837364000},"score":1,"resource":{"primary":{"URL":"https:\/\/journals.lww.com\/00130478-201809000-00010"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[2018,9]]},"references-count":13,"journal-issue":{"issue":"9","published-print":{"date-parts":[[2018]]}},"URL":"https:\/\/doi.org\/10.1097\/pcc.0000000000001638","relation":{},"ISSN":["1529-7535"],"issn-type":[{"value":"1529-7535","type":"print"}],"subject":[],"published":{"date-parts":[[2018,9]]}}}