{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2024,8,4]],"date-time":"2024-08-04T06:00:04Z","timestamp":1722751204563},"reference-count":5,"publisher":"Wiley","issue":"10","license":[{"start":{"date-parts":[[2004,9,23]],"date-time":"2004-09-23T00:00:00Z","timestamp":1095897600000},"content-version":"vor","delay-in-days":0,"URL":"http:\/\/onlinelibrary.wiley.com\/termsAndConditions#vor"}],"content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":["Artificial Organs"],"published-print":{"date-parts":[[2004,10]]},"abstract":"<jats:p><jats:bold>Abstract:<\/jats:bold> A physiological controller is a key component to optimize the performance of left ventricular assist devices (LVADs) from short term to long term (10\u201320\u2003years). The difficulties of its design lie in the inaccurate modeling of any particular human cardiovascular system because of limited knowledge, the large fluctuation of systemic parameters in normal daily life, nonlinearity introduced by one\u2010way heart valves, and restriction on available sensors. The desired physiological controller for a permanent LVAD is expected to handle different activity levels such as sleeping, resting and exercising, and different pathological levels of left ventricular failure. In this study, a physiological controller designed for a long\u2010term continuous flow (CF) LVAD was tested in a mock circulatory loop. The pressure head across the LVAD and motor signals were used as feedback signals, which were obtained from long\u2010term reliable sensors. Three activities and five different pathological levels of left ventricular failure were simulated to test the performance of the designed controller. The experimental results showed that the designed controller restored the total peripheral flow and arterial pressure, and decreased the left ventricular end\u2010diastolic pressure to a safe range. The suction of the left ventricle was successfully reversed by the designed controller. The controller thus was proven to be able to control the LVAD properly in many of the possible physiological conditions experienced by LVAD recipients in long term.<\/jats:p>","DOI":"10.1111\/j.1525-1594.2004.07381.x","type":"journal-article","created":{"date-parts":[[2004,9,23]],"date-time":"2004-09-23T09:01:52Z","timestamp":1095930112000},"page":"927-932","source":"Crossref","is-referenced-by-count":21,"title":["A Bridge from Short\u2010term to Long\u2010term Left Ventricular Assist Device\u2014Experimental Verification of a Physiological Controller"],"prefix":"10.1111","volume":"28","author":[{"given":"Yi","family":"Wu","sequence":"first","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Paul E.","family":"Allaire","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Gang","family":"Tao","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Milton","family":"Adams","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Yingjie","family":"Liu","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Houston","family":"Wood","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Don B.","family":"Olsen","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]}],"member":"311","published-online":{"date-parts":[[2004,9,23]]},"reference":[{"key":"e_1_2_7_2_2","volume-title":"Textbook of Medical Physiology","author":"Guyton AC","year":"1996"},{"key":"e_1_2_7_3_2","volume-title":"A Guide to Interpretation of Hemodynamic Data in the Coronary Care Unit.","author":"Sharkey SW","year":"1997"},{"key":"e_1_2_7_4_2","doi-asserted-by":"publisher","DOI":"10.1097\/00002480-200009000-00020"},{"key":"e_1_2_7_5_2","doi-asserted-by":"crossref","unstructured":"WuY AllaireP TaoG WoodH OlsenD TribbleC.An advanced physiological controller design for an LVAD to prevent left ventricular collapse.Artif Organs2003;27:926\u201330.","DOI":"10.1046\/j.1525-1594.2003.00032.x"},{"key":"e_1_2_7_6_2","unstructured":"LiuY.Design of a Performance Test Loop for an Artificial Heart Pump.Master Thesis. University of Virginia August 2003."}],"container-title":["Artificial Organs"],"original-title":[],"language":"en","link":[{"URL":"https:\/\/api.wiley.com\/onlinelibrary\/tdm\/v1\/articles\/10.1111%2Fj.1525-1594.2004.07381.x","content-type":"unspecified","content-version":"vor","intended-application":"text-mining"},{"URL":"https:\/\/onlinelibrary.wiley.com\/doi\/pdf\/10.1111\/j.1525-1594.2004.07381.x","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2023,10,16]],"date-time":"2023-10-16T07:12:27Z","timestamp":1697440347000},"score":1,"resource":{"primary":{"URL":"https:\/\/onlinelibrary.wiley.com\/doi\/10.1111\/j.1525-1594.2004.07381.x"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[2004,9,23]]},"references-count":5,"journal-issue":{"issue":"10","published-print":{"date-parts":[[2004,10]]}},"alternative-id":["10.1111\/j.1525-1594.2004.07381.x"],"URL":"https:\/\/doi.org\/10.1111\/j.1525-1594.2004.07381.x","archive":["Portico"],"relation":{},"ISSN":["0160-564X","1525-1594"],"issn-type":[{"value":"0160-564X","type":"print"},{"value":"1525-1594","type":"electronic"}],"subject":[],"published":{"date-parts":[[2004,9,23]]}}}