{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2026,10,7]],"date-time":"2026-10-07T05:39:17Z","timestamp":1791351557790,"version":"4.3.1"},"reference-count":18,"publisher":"Wiley","issue":"2","license":[{"start":{"date-parts":[[2007,2,12]],"date-time":"2007-02-12T00:00:00Z","timestamp":1171238400000},"content-version":"vor","delay-in-days":11,"URL":"http:\/\/onlinelibrary.wiley.com\/termsAndConditions#vor"}],"content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":["Artificial Organs"],"published-print":{"date-parts":[[2007,2]]},"abstract":"<jats:p><jats:bold>Abstract:\u2002<\/jats:bold> During hemodialysis microembolic findings have been noted after the venous chamber (subclavian vein). The aim of this study was to evaluate if air could pass the venous chamber and, if so, if it passes the safety\u2010system detector for air\u2010infusion without triggering an alarm.Various in vitro dialysis settings were performed using regular dialysis devices. A dextran fluid was used instead of blood to avoid the risk of development of emboli. Optical visualization as well as recirculation and collection of eventual air into an intermediate bag were investigated. In addition, a specifically designed ultrasound monitor was placed after the venous air trap to measure the presence of eventual microbubbles. Speed of dialysis fluid was changed, as was the level of the fluid in the air trap. Thereby a fluid level was considered \u201chigh\u201d if it was close to the top of the air trap and \u201clow\u201d if it was around the mid part of the air trap. By optical vision microbubbles were seen at the bottom of the air trap and could pass the air trap towards the venous line without alarming. During recirculation several mL of air were collected in an intermediate bag after the venous line. Ultrasound monitoring exhibited the presence of microbubbles of the size of approximately 5\u2003\u00b5m upwards passing to the venous line in all runs performed. Amount of bubbles differed between devices and in general an increased fluid speed correlated significantly with the increased counts of microbubbles\/min. No alarming of the detector occurred. A more concentrated fluid allowed higher counts\/min when flow was increased to 600\u2003mL\/min. Data revealed that air passes the safety\u2010sensor in the air trap without alarming. The presence of air increased in general with fluid speed and a lower fluid level in the air trap. Differences were present between devices. If this affects the patients has to be elucidated.<\/jats:p>","DOI":"10.1111\/j.1525-1594.2007.00352.x","type":"journal-article","created":{"date-parts":[[2007,2,12]],"date-time":"2007-02-12T20:09:47Z","timestamp":1171310987000},"page":"132-139","source":"Crossref","is-referenced-by-count":41,"title":["Air Bubbles Pass the Security System of the Dialysis Device Without Alarming"],"prefix":"10.1111","volume":"31","author":[{"given":"Per","family":"Jonsson","sequence":"first","affiliation":[],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Lars","family":"Karlsson","sequence":"additional","affiliation":[],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Ulf","family":"Forsberg","sequence":"additional","affiliation":[],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Margareta","family":"Gref","sequence":"additional","affiliation":[],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Christofer","family":"Stegmayr","sequence":"additional","affiliation":[],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Bernd","family":"Stegmayr","sequence":"additional","affiliation":[],"role":[{"vocabulary":"crossref","role":"author"}]}],"member":"311","published-online":{"date-parts":[[2007,2,12]]},"reference":[{"key":"e_1_2_7_2_2","volume-title":"Dialysis and Transplantation: A Companion to Brenner & Rektor's The Kidney","author":"Owen WF","year":"2000"},{"key":"e_1_2_7_3_2","first-page":"8","volume-title":"Particular Requirements for the Safety of Haemodialysis, Haemodiafiltration and Haemofiltration Equipment","author":"International Electrotechnical Commission (IEC)","year":"1998"},{"key":"e_1_2_7_4_2","unstructured":"Socialstyrelsen.Infusion of air contaminated blood through a central dialysis catheter by mistake. Report\u2014SOSFS1991:34 19930622 dnr:5287\/1993. Socialstyrelsen Sweden."},{"key":"e_1_2_7_5_2","volume-title":"CMD\u201010 Cardiovascular Micro Bubble Detector USER MANUAL","author":"Hatteland K.","year":"1988"},{"key":"e_1_2_7_6_2","first-page":"1","volume-title":"The Scandinavian Society of Extracorporeal Technology. Post Graduate Course (Aug 21, 1991)","author":"Pedersen TH.","year":"1991"},{"key":"e_1_2_7_7_2","doi-asserted-by":"publisher","DOI":"10.3109\/14017438509102706"},{"key":"e_1_2_7_8_2","doi-asserted-by":"publisher","DOI":"10.1007\/978-1-4020-2275-3_14"},{"key":"e_1_2_7_9_2","unstructured":"International Electrotechnical Commission (IEC).Medical electrical equipment\u2014part 2\u201024: particular requirements for the safety of infusion pumps and controllers. IEC 60601\u20102\u201024. Rationale to 51.104."},{"key":"e_1_2_7_10_2","doi-asserted-by":"publisher","DOI":"10.1093\/ndt\/gfg352"},{"key":"e_1_2_7_11_2","doi-asserted-by":"publisher","DOI":"10.1093\/ndt\/17.3.462"},{"key":"e_1_2_7_12_2","doi-asserted-by":"publisher","DOI":"10.1093\/ndt\/15.9.1420"},{"key":"e_1_2_7_13_2","doi-asserted-by":"publisher","DOI":"10.1111\/j.1525-1594.1984.tb04275.x"},{"key":"e_1_2_7_14_2","doi-asserted-by":"publisher","DOI":"10.1111\/j.1525-1594.2007.00358.x"},{"key":"e_1_2_7_15_2","first-page":"97","article-title":"Lung pathology in chronic hemodialysis patients","volume":"5","author":"Fairshter RD","year":"1982","journal-title":"Int J Artif Organs"},{"key":"e_1_2_7_16_2","doi-asserted-by":"publisher","DOI":"10.1177\/039139889201500103"},{"key":"e_1_2_7_17_2","doi-asserted-by":"publisher","DOI":"10.3109\/00365519609090600"},{"key":"e_1_2_7_18_2","doi-asserted-by":"publisher","DOI":"10.1378\/chest.128.4.2918"},{"key":"e_1_2_7_19_2","doi-asserted-by":"publisher","DOI":"10.1111\/j.1492-7535.2005.01116.x"}],"container-title":["Artificial Organs"],"original-title":[],"language":"en","link":[{"URL":"https:\/\/api.wiley.com\/onlinelibrary\/tdm\/v1\/articles\/10.1111%2Fj.1525-1594.2007.00352.x","content-type":"unspecified","content-version":"vor","intended-application":"text-mining"},{"URL":"https:\/\/onlinelibrary.wiley.com\/doi\/pdf\/10.1111\/j.1525-1594.2007.00352.x","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2023,10,16]],"date-time":"2023-10-16T11:21:12Z","timestamp":1697455272000},"score":1,"resource":{"primary":{"URL":"https:\/\/onlinelibrary.wiley.com\/doi\/10.1111\/j.1525-1594.2007.00352.x"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[2007,2]]},"references-count":18,"journal-issue":{"issue":"2","published-print":{"date-parts":[[2007,2]]}},"alternative-id":["10.1111\/j.1525-1594.2007.00352.x"],"URL":"https:\/\/doi.org\/10.1111\/j.1525-1594.2007.00352.x","archive":["Portico"],"relation":{},"ISSN":["0160-564X","1525-1594"],"issn-type":[{"value":"0160-564X","type":"print"},{"value":"1525-1594","type":"electronic"}],"subject":[],"published":{"date-parts":[[2007,2]]}}}