{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2023,11,2]],"date-time":"2023-11-02T09:43:24Z","timestamp":1698918204889},"reference-count":1,"publisher":"Wiley","issue":"5","license":[{"start":{"date-parts":[[2006,6,30]],"date-time":"2006-06-30T00:00:00Z","timestamp":1151625600000},"content-version":"vor","delay-in-days":9068,"URL":"http:\/\/onlinelibrary.wiley.com\/termsAndConditions#vor"}],"content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":["Pacing Clinical Electrophis"],"published-print":{"date-parts":[[1981,9]]},"abstract":"<jats:p>Malfunction of a lithium\u2010powered R\u2010wave inhibited unipolar pulse generator (CPI* Model 0505) necessitated removal 15 months after implantation. Examination of the pulse generator in a water bath revealed abnormal function at temperatures above 31.5\u00b0C and normal function at temperatures below 31.5\u00b0C. CPI suggested that the failure was caused by the 10 kHz timing crystal in the generator. (PACE, Vol. 4, September\u2010October, 1981)<\/jats:p>","DOI":"10.1111\/j.1540-8159.1981.tb06222.x","type":"journal-article","created":{"date-parts":[[2006,7,2]],"date-time":"2006-07-02T04:32:12Z","timestamp":1151814732000},"page":"523-524","source":"Crossref","is-referenced-by-count":2,"title":["Temperature Dependent Pulse Generator Dysfunction"],"prefix":"10.1111","volume":"4","author":[{"given":"ARNE M.","family":"ABRAHAMSEN","sequence":"first","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"TORBJ\u00d8RN","family":"AARSLAND","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]}],"member":"311","published-online":{"date-parts":[[2006,6,30]]},"reference":[{"key":"e_1_2_1_2_2","doi-asserted-by":"publisher","DOI":"10.1111\/j.1540-8159.1980.tb05281.x"}],"container-title":["Pacing and Clinical Electrophysiology"],"original-title":[],"language":"en","link":[{"URL":"https:\/\/api.wiley.com\/onlinelibrary\/tdm\/v1\/articles\/10.1111%2Fj.1540-8159.1981.tb06222.x","content-type":"unspecified","content-version":"vor","intended-application":"text-mining"},{"URL":"https:\/\/onlinelibrary.wiley.com\/doi\/pdf\/10.1111\/j.1540-8159.1981.tb06222.x","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2023,11,1]],"date-time":"2023-11-01T19:50:46Z","timestamp":1698868246000},"score":1,"resource":{"primary":{"URL":"https:\/\/onlinelibrary.wiley.com\/doi\/10.1111\/j.1540-8159.1981.tb06222.x"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[1981,9]]},"references-count":1,"journal-issue":{"issue":"5","published-print":{"date-parts":[[1981,9]]}},"alternative-id":["10.1111\/j.1540-8159.1981.tb06222.x"],"URL":"https:\/\/doi.org\/10.1111\/j.1540-8159.1981.tb06222.x","archive":["Portico"],"relation":{},"ISSN":["0147-8389","1540-8159"],"issn-type":[{"value":"0147-8389","type":"print"},{"value":"1540-8159","type":"electronic"}],"subject":[],"published":{"date-parts":[[1981,9]]}}}