{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2023,11,2]],"date-time":"2023-11-02T00:43:44Z","timestamp":1698885824856},"reference-count":10,"publisher":"Wiley","issue":"6","license":[{"start":{"date-parts":[[2006,6,30]],"date-time":"2006-06-30T00:00:00Z","timestamp":1151625600000},"content-version":"vor","delay-in-days":9007,"URL":"http:\/\/onlinelibrary.wiley.com\/termsAndConditions#vor"}],"content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":["Pacing Clinical Electrophis"],"published-print":{"date-parts":[[1981,11]]},"abstract":"<jats:p>A variety of commercial pacemaker system analyzers have been used to evaluate the electrical performance of pulse generators prior to implant or at the time of reoperation for pacemaker malfunction. We evaluated the ability of seven commercially available pacemaker system analyzers to provide the data required to determine if a given pulse generator was operating according to its manufacturer's specifications. While all pacemaker system analyzers provided accurate data for rate and pulse duration, values displayed for amplitude and sensitivity were inconsistent due to the methods each pacemaker system analyzer employed to measure these parameters. Overall, 22% of representative test sequences yielded data which would have resulted in the inappropriate rejection of the subject pulse generators for implantation. These observations suggest that pacemaker system analyzers are not suitable for evaluating all pulse generators. Moreover, physicians must be aware of these deficiencies in order to avoid the pitfalls of relying exclusively on pacemaker system analyzer data when troubleshooting pacemaker malfunction. (PACE, Vol. 4, November\u2010December, 1981)<\/jats:p>","DOI":"10.1111\/j.1540-8159.1981.tb06248.x","type":"journal-article","created":{"date-parts":[[2006,7,2]],"date-time":"2006-07-02T08:34:00Z","timestamp":1151829240000},"page":"650-656","source":"Crossref","is-referenced-by-count":7,"title":["Limitations of Pacemaker System Analyzers for the Evaluation of Implantable Pulse Generators"],"prefix":"10.1111","volume":"4","author":[{"given":"ROBERT G.","family":"HAUSER","sequence":"first","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"LYNN M.","family":"EDWARDS","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"VERLIN W.","family":"GIUFFRE","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]}],"member":"311","published-online":{"date-parts":[[2006,6,30]]},"reference":[{"key":"e_1_2_1_2_2","volume-title":"Medtronic Pacing System Analyzer, Model 5300","author":"Manual","year":"1974"},{"key":"e_1_2_1_3_2","volume-title":"The Cordis Pacer System Analyzer Model 209A: Instructions for Use, Cordis Corp.","author":"Manual","year":"1976"},{"key":"e_1_2_1_4_2","volume-title":"Cardiotest Pacemaker System Tester, Model 2200","author":"Manual","year":"1979"},{"key":"e_1_2_1_5_2","volume-title":"Electrode\/Pacer System Analyzer, Model PMA200","author":"Manual","year":"1976"},{"key":"e_1_2_1_6_2","volume-title":"PSA\u20105 Pacer System Analyzer","author":"Manual","year":"1978"},{"key":"e_1_2_1_7_2","volume-title":"Pacer System Analyzer Model PM200A","author":"Manual","year":"1978"},{"key":"e_1_2_1_8_2","volume-title":"Compupace Pacing System Computer, Model 524\u201301","author":"Manual","year":"1980"},{"key":"e_1_2_1_9_2","doi-asserted-by":"publisher","DOI":"10.1111\/j.1540-8159.1980.tb05278.x"},{"key":"e_1_2_1_10_2","unstructured":"Pacemaker standard labelling requirements performance requirements and terminology for implantable artificial cardiac pacemakers. FDA Contract No. 223\u201374\u20135083 Association for the Advancement of Medical Instrumentation Arlington Virginia 1975."},{"key":"e_1_2_1_11_2","first-page":"19","article-title":"Clinical assessment of cardiac pacemaker performance","volume":"14","author":"Hauser R.G.","year":"1979","journal-title":"Contin. Educ. Cardiol"}],"container-title":["Pacing and Clinical Electrophysiology"],"original-title":[],"language":"en","link":[{"URL":"https:\/\/api.wiley.com\/onlinelibrary\/tdm\/v1\/articles\/10.1111%2Fj.1540-8159.1981.tb06248.x","content-type":"unspecified","content-version":"vor","intended-application":"text-mining"},{"URL":"https:\/\/onlinelibrary.wiley.com\/doi\/pdf\/10.1111\/j.1540-8159.1981.tb06248.x","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2023,11,2]],"date-time":"2023-11-02T00:10:40Z","timestamp":1698883840000},"score":1,"resource":{"primary":{"URL":"https:\/\/onlinelibrary.wiley.com\/doi\/10.1111\/j.1540-8159.1981.tb06248.x"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[1981,11]]},"references-count":10,"journal-issue":{"issue":"6","published-print":{"date-parts":[[1981,11]]}},"alternative-id":["10.1111\/j.1540-8159.1981.tb06248.x"],"URL":"https:\/\/doi.org\/10.1111\/j.1540-8159.1981.tb06248.x","archive":["Portico"],"relation":{},"ISSN":["0147-8389","1540-8159"],"issn-type":[{"value":"0147-8389","type":"print"},{"value":"1540-8159","type":"electronic"}],"subject":[],"published":{"date-parts":[[1981,11]]}}}