{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2025,10,30]],"date-time":"2025-10-30T06:54:01Z","timestamp":1761807241788},"reference-count":17,"publisher":"Wiley","issue":"3","license":[{"start":{"date-parts":[[2006,6,30]],"date-time":"2006-06-30T00:00:00Z","timestamp":1151625600000},"content-version":"vor","delay-in-days":3773,"URL":"http:\/\/onlinelibrary.wiley.com\/termsAndConditions#vor"}],"content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":["Pacing Clinical Electrophis"],"published-print":{"date-parts":[[1996,3]]},"abstract":"<jats:p>A series of 78 consecutive implants of the transvene PCD (Medtronic, Inc.) defibrillator system is presented and the occurrence of right ventricular perforation in 4 patients reported (5.2%). Diagnosis of perforation is made using four signs: (1) decrease in arterial blood pressure without any other explanation; (2) decrease in pulsatility of the cardiac silhouette as monitored by fluoroscopy; (3) increased size of the cardiac silhouette; and (4) abnormal position of the transvenous lead too far out toward the left ventricle along the pericardial outline. Perforation causes rapid and dramatic cardiac tamponade due to the large diameter and stiffness of the coil carrier lead. Immediate drainage of the hemopericardium must be carried out using the transxiphoid approach. The use of a thin blue\u2010coded lead stylet (0.014\u2010inch gauge) is recommended over the stiffer maroon\u2010coded stylet. Since treatment must be carried out immediately, it is advised that a surgeon either perform, assist, or be immediately available whenever one of these systems is implanted.<\/jats:p>","DOI":"10.1111\/j.1540-8159.1996.tb03329.x","type":"journal-article","created":{"date-parts":[[2006,7,2]],"date-time":"2006-07-02T17:00:11Z","timestamp":1151859611000},"page":"288-292","source":"Crossref","is-referenced-by-count":35,"title":["Perforation of the Right Ventricle by Transvenous Defibrillator Leads: Prevention and Treatment"],"prefix":"10.1111","volume":"19","author":[{"given":"J. ERNESTO","family":"MOLINA","sequence":"first","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]}],"member":"311","published-online":{"date-parts":[[2006,6,30]]},"reference":[{"key":"e_1_2_1_2_2","first-page":"119","volume-title":"A history of the heart and the circulation","author":"Willius FA","year":"1948"},{"key":"e_1_2_1_3_2","doi-asserted-by":"publisher","DOI":"10.1002\/bjs.1800540712"},{"key":"e_1_2_1_4_2","doi-asserted-by":"publisher","DOI":"10.1016\/S0003-4975(10)65393-3"},{"key":"e_1_2_1_5_2","doi-asserted-by":"publisher","DOI":"10.1016\/S0003-4975(10)65732-3"},{"key":"e_1_2_1_6_2","doi-asserted-by":"publisher","DOI":"10.1016\/S0003-4975(10)64168-9"},{"key":"e_1_2_1_7_2","doi-asserted-by":"publisher","DOI":"10.1378\/chest.88.1.30"},{"key":"e_1_2_1_8_2","doi-asserted-by":"publisher","DOI":"10.1016\/S0003-4975(10)62862-7"},{"key":"e_1_2_1_9_2","first-page":"95","article-title":"Subxiphoid pericardial window. A simple and safe method for diagnosing and treating acute and chronic pericardial effusions","volume":"62","author":"Fontenelle LJ","year":"1971","journal-title":"J Thorac Cardiovasc Surg"},{"key":"e_1_2_1_10_2","doi-asserted-by":"publisher","DOI":"10.1111\/j.1540-8159.1989.tb01931.x"},{"key":"e_1_2_1_11_2","doi-asserted-by":"publisher","DOI":"10.1378\/chest.63.2.185"},{"key":"e_1_2_1_12_2","doi-asserted-by":"publisher","DOI":"10.1111\/j.1540-8159.1983.tb04419.x"},{"key":"e_1_2_1_13_2","doi-asserted-by":"publisher","DOI":"10.1111\/j.1540-8159.1987.tb04521.x"},{"key":"e_1_2_1_14_2","doi-asserted-by":"publisher","DOI":"10.7326\/0003-4819-74-3-361"},{"key":"e_1_2_1_15_2","doi-asserted-by":"publisher","DOI":"10.1001\/archsurg.1972.04180110030009"},{"key":"e_1_2_1_16_2","first-page":"51","article-title":"Pericardial lamponade due lo perforalion wilh a permanenl endocardial pacing calheler","volume":"74","author":"Bassan MM","year":"1977","journal-title":"J Cardiovasc Surg"},{"key":"e_1_2_1_17_2","doi-asserted-by":"publisher","DOI":"10.1016\/0002-9149(69)90416-0"},{"key":"e_1_2_1_18_2","doi-asserted-by":"publisher","DOI":"10.1111\/j.1540-8159.1987.tb06049.x"}],"container-title":["Pacing and Clinical Electrophysiology"],"original-title":[],"language":"en","link":[{"URL":"https:\/\/api.wiley.com\/onlinelibrary\/tdm\/v1\/articles\/10.1111%2Fj.1540-8159.1996.tb03329.x","content-type":"unspecified","content-version":"vor","intended-application":"text-mining"},{"URL":"https:\/\/onlinelibrary.wiley.com\/doi\/pdf\/10.1111\/j.1540-8159.1996.tb03329.x","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2023,10,27]],"date-time":"2023-10-27T07:41:28Z","timestamp":1698392488000},"score":1,"resource":{"primary":{"URL":"https:\/\/onlinelibrary.wiley.com\/doi\/10.1111\/j.1540-8159.1996.tb03329.x"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[1996,3]]},"references-count":17,"journal-issue":{"issue":"3","published-print":{"date-parts":[[1996,3]]}},"alternative-id":["10.1111\/j.1540-8159.1996.tb03329.x"],"URL":"https:\/\/doi.org\/10.1111\/j.1540-8159.1996.tb03329.x","archive":["Portico"],"relation":{},"ISSN":["0147-8389","1540-8159"],"issn-type":[{"value":"0147-8389","type":"print"},{"value":"1540-8159","type":"electronic"}],"subject":[],"published":{"date-parts":[[1996,3]]}}}