{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2025,9,27]],"date-time":"2025-09-27T04:05:34Z","timestamp":1758945934504},"reference-count":9,"publisher":"Wiley","issue":"3","license":[{"start":{"date-parts":[[2006,6,30]],"date-time":"2006-06-30T00:00:00Z","timestamp":1151625600000},"content-version":"vor","delay-in-days":3773,"URL":"http:\/\/onlinelibrary.wiley.com\/termsAndConditions#vor"}],"content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":["Pacing Clinical Electrophis"],"published-print":{"date-parts":[[1996,3]]},"abstract":"<jats:p>The influence of body posture on atrial sensing was examined in a single\u2010lead VDD pacemaker in 16 patients. At implantation, a position was searched for the floating atrial bipolar sensor to obtain an adequate atrial signal amplitude. After 1\u201312 months the atrial signals were measured by programming the sensitivity in five postures: supine, sitting, and standing and lying on the right and left sides. The group mean amplitudes of the atrial signal were equal in all postures, and comparable to the supine value (1.01 \u00b1 0.51 mV), However, the values within each individual varied considerably according to position, by a range of 0.46 \u00b1 0.41 mV on average. Testing only supine did not always predict decent sensing when upright, e.g., in 3 patients the atrial signal decreased more than 0.5 mV. In 24\u2010hour ambulatory electrocardiography, with the sensitivity (0.15\u20130.30 mV) set to cover all postures, atrial beats were undersensed not at all in 4 patients, &lt; 0.01% in 6 patients, and &lt; 0.1% in the remaining 6 patients. Thus, with a floating atrial electrode, sensitivity testing in various postures is advisable to ascertain proper sensing function.<\/jats:p>","DOI":"10.1111\/j.1540-8159.1996.tb03332.x","type":"journal-article","created":{"date-parts":[[2006,7,2]],"date-time":"2006-07-02T17:00:11Z","timestamp":1151859611000},"page":"309-313","source":"Crossref","is-referenced-by-count":28,"title":["Dependence of Atrial Sensing Function on Posture in a Single\u2010Lead Atrial Triggered Ventricular (VDD) Pacemaker"],"prefix":"10.1111","volume":"19","author":[{"given":"LAURI","family":"TOIVONEN","sequence":"first","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"JYRI","family":"LOMMI","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]}],"member":"311","published-online":{"date-parts":[[2006,6,30]]},"reference":[{"key":"e_1_2_1_2_2","doi-asserted-by":"publisher","DOI":"10.1111\/j.1540-8159.1991.tb04092.x"},{"key":"e_1_2_1_3_2","doi-asserted-by":"publisher","DOI":"10.1111\/j.1540-8159.1991.tb02757.x"},{"key":"e_1_2_1_4_2","first-page":"865","volume-title":"Cardiac Pacing: Electrophysiology, Tachyarrhythmias","author":"Curzio G","year":"1985"},{"key":"e_1_2_1_5_2","doi-asserted-by":"publisher","DOI":"10.1111\/j.1540-8159.1992.tb02988.x"},{"key":"e_1_2_1_6_2","doi-asserted-by":"publisher","DOI":"10.1111\/j.1540-8159.1990.tb02130.x"},{"key":"e_1_2_1_7_2","doi-asserted-by":"publisher","DOI":"10.1111\/j.1540-8159.1992.tb02990.x"},{"key":"e_1_2_1_8_2","doi-asserted-by":"publisher","DOI":"10.1111\/j.1540-8159.1992.tb02989.x"},{"key":"e_1_2_1_9_2","doi-asserted-by":"publisher","DOI":"10.1111\/j.1540-8159.1992.tb02991.x"},{"key":"e_1_2_1_10_2","doi-asserted-by":"publisher","DOI":"10.1016\/0735-1097(93)90769-W"}],"container-title":["Pacing and Clinical Electrophysiology"],"original-title":[],"language":"en","link":[{"URL":"https:\/\/api.wiley.com\/onlinelibrary\/tdm\/v1\/articles\/10.1111%2Fj.1540-8159.1996.tb03332.x","content-type":"unspecified","content-version":"vor","intended-application":"text-mining"},{"URL":"https:\/\/onlinelibrary.wiley.com\/doi\/pdf\/10.1111\/j.1540-8159.1996.tb03332.x","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2023,10,27]],"date-time":"2023-10-27T07:41:30Z","timestamp":1698392490000},"score":1,"resource":{"primary":{"URL":"https:\/\/onlinelibrary.wiley.com\/doi\/10.1111\/j.1540-8159.1996.tb03332.x"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[1996,3]]},"references-count":9,"journal-issue":{"issue":"3","published-print":{"date-parts":[[1996,3]]}},"alternative-id":["10.1111\/j.1540-8159.1996.tb03332.x"],"URL":"https:\/\/doi.org\/10.1111\/j.1540-8159.1996.tb03332.x","archive":["Portico"],"relation":{},"ISSN":["0147-8389","1540-8159"],"issn-type":[{"value":"0147-8389","type":"print"},{"value":"1540-8159","type":"electronic"}],"subject":[],"published":{"date-parts":[[1996,3]]}}}