{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2025,9,23]],"date-time":"2025-09-23T14:51:13Z","timestamp":1758639073180},"reference-count":18,"publisher":"Wiley","issue":"12","license":[{"start":{"date-parts":[[2009,11,24]],"date-time":"2009-11-24T00:00:00Z","timestamp":1259020800000},"content-version":"vor","delay-in-days":0,"URL":"http:\/\/onlinelibrary.wiley.com\/termsAndConditions#vor"}],"content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":["Medical Physics"],"published-print":{"date-parts":[[2009,12]]},"abstract":"<jats:sec><jats:title>Purpose:<\/jats:title><jats:p>Cardiac CT achieves its high temporal resolution by lowering the scan range from<jats:inline-graphic xmlns:xlink=\"http:\/\/www.w3.org\/1999\/xlink\" xlink:href=\"graphic\/mp9734-math-0001.png\" xlink:title=\"urn:x-wiley:00942405:media:mp9734:mp9734-math-0001\" \/> to <jats:inline-graphic xmlns:xlink=\"http:\/\/www.w3.org\/1999\/xlink\" xlink:href=\"graphic\/mp9734-math-0002.png\" xlink:title=\"urn:x-wiley:00942405:media:mp9734:mp9734-math-0002\" \/> plus fan angle (partial scan). This, however, introduces CT\u2010value variations, depending on the angular position of the <jats:inline-graphic xmlns:xlink=\"http:\/\/www.w3.org\/1999\/xlink\" xlink:href=\"graphic\/mp9734-math-0003.png\" xlink:title=\"urn:x-wiley:00942405:media:mp9734:mp9734-math-0003\" \/> range. These partial scan artifacts are of the order of a few HU and prevent the quantitative evaluation of perfusion measurements. The authors present the new algorithm partial scan artifact reduction (PSAR) that corrects a dynamic phase\u2010correlated scan without <jats:italic>a priori<\/jats:italic> information.<\/jats:p><\/jats:sec><jats:sec><jats:title>Methods:<\/jats:title><jats:p>In general, a full scan does not suffer from partial scan artifacts since all projections in [0,<jats:inline-graphic xmlns:xlink=\"http:\/\/www.w3.org\/1999\/xlink\" xlink:href=\"graphic\/mp9734-math-0004.png\" xlink:title=\"urn:x-wiley:00942405:media:mp9734:mp9734-math-0004\" \/>] contribute to the data. To maintain the optimum temporal resolution and the phase correlation, PSAR creates an artificial full scan <jats:inline-graphic xmlns:xlink=\"http:\/\/www.w3.org\/1999\/xlink\" xlink:href=\"graphic\/mp9734-math-0005.png\" xlink:title=\"urn:x-wiley:00942405:media:mp9734:mp9734-math-0005\" \/> by projectionwise averaging a set of neighboring partial scans <jats:inline-graphic xmlns:xlink=\"http:\/\/www.w3.org\/1999\/xlink\" xlink:href=\"graphic\/mp9734-math-0006.png\" xlink:title=\"urn:x-wiley:00942405:media:mp9734:mp9734-math-0006\" \/> from the same perfusion examination (typically <jats:inline-graphic xmlns:xlink=\"http:\/\/www.w3.org\/1999\/xlink\" xlink:href=\"graphic\/mp9734-math-0007.png\" xlink:title=\"urn:x-wiley:00942405:media:mp9734:mp9734-math-0007\" \/> phase\u2010correlated partial scans distributed over <jats:inline-graphic xmlns:xlink=\"http:\/\/www.w3.org\/1999\/xlink\" xlink:href=\"graphic\/mp9734-math-0008.png\" xlink:title=\"urn:x-wiley:00942405:media:mp9734:mp9734-math-0008\" \/> and <jats:inline-graphic xmlns:xlink=\"http:\/\/www.w3.org\/1999\/xlink\" xlink:href=\"graphic\/mp9734-math-0009.png\" xlink:title=\"urn:x-wiley:00942405:media:mp9734:mp9734-math-0009\" \/>). Corresponding to the angular range of each partial scan, the authors extract virtual partial scans <jats:inline-graphic xmlns:xlink=\"http:\/\/www.w3.org\/1999\/xlink\" xlink:href=\"graphic\/mp9734-math-0010.png\" xlink:title=\"urn:x-wiley:00942405:media:mp9734:mp9734-math-0010\" \/> from the artificial full scan <jats:inline-graphic xmlns:xlink=\"http:\/\/www.w3.org\/1999\/xlink\" xlink:href=\"graphic\/mp9734-math-0011.png\" xlink:title=\"urn:x-wiley:00942405:media:mp9734:mp9734-math-0011\" \/>. A standard reconstruction yields the corresponding images <jats:inline-graphic xmlns:xlink=\"http:\/\/www.w3.org\/1999\/xlink\" xlink:href=\"graphic\/mp9734-math-0012.png\" xlink:title=\"urn:x-wiley:00942405:media:mp9734:mp9734-math-0012\" \/>, <jats:inline-graphic xmlns:xlink=\"http:\/\/www.w3.org\/1999\/xlink\" xlink:href=\"graphic\/mp9734-math-0013.png\" xlink:title=\"urn:x-wiley:00942405:media:mp9734:mp9734-math-0013\" \/>, and <jats:inline-graphic xmlns:xlink=\"http:\/\/www.w3.org\/1999\/xlink\" xlink:href=\"graphic\/mp9734-math-0014.png\" xlink:title=\"urn:x-wiley:00942405:media:mp9734:mp9734-math-0014\" \/>. Subtracting the virtual partial scan image <jats:inline-graphic xmlns:xlink=\"http:\/\/www.w3.org\/1999\/xlink\" xlink:href=\"graphic\/mp9734-math-0015.png\" xlink:title=\"urn:x-wiley:00942405:media:mp9734:mp9734-math-0015\" \/> from the artificial full scan image <jats:inline-graphic xmlns:xlink=\"http:\/\/www.w3.org\/1999\/xlink\" xlink:href=\"graphic\/mp9734-math-0016.png\" xlink:title=\"urn:x-wiley:00942405:media:mp9734:mp9734-math-0016\" \/> yields an artifact image that can be used to correct the original partial scan image: <jats:inline-graphic xmlns:xlink=\"http:\/\/www.w3.org\/1999\/xlink\" xlink:href=\"graphic\/mp9734-math-0017.png\" xlink:title=\"urn:x-wiley:00942405:media:mp9734:mp9734-math-0017\" \/>, where <jats:inline-graphic xmlns:xlink=\"http:\/\/www.w3.org\/1999\/xlink\" xlink:href=\"graphic\/mp9734-math-0018.png\" xlink:title=\"urn:x-wiley:00942405:media:mp9734:mp9734-math-0018\" \/> is the corrected image.<\/jats:p><\/jats:sec><jats:sec><jats:title>Results:<\/jats:title><jats:p>The authors evaluated the effects of scattered radiation on the partial scan artifacts using simulated and measured water phantoms and found a strong correlation. The PSAR algorithm has been validated with a simulated semianthropomorphic heart phantom and with measurements of a dynamic biological perfusion phantom. For the stationary phantoms, real full scans have been performed to provide theoretical reference values. The improvement in the root mean square errors between the full and the partial scans with respect to the errors between the full and the corrected scans is up to 54% for the simulations and 90% for the measurements.<\/jats:p><\/jats:sec><jats:sec><jats:title>Conclusions:<\/jats:title><jats:p>The phase\u2010correlated data now appear accurate enough for a quantitative analysis of cardiac perfusion.<\/jats:p><\/jats:sec>","DOI":"10.1118\/1.3259734","type":"journal-article","created":{"date-parts":[[2009,11,24]],"date-time":"2009-11-24T23:37:16Z","timestamp":1259105836000},"page":"5683-5694","source":"Crossref","is-referenced-by-count":21,"title":["Partial scan artifact reduction (PSAR) for the assessment of cardiac perfusion in dynamic phase\u2010correlated CT"],"prefix":"10.1002","volume":"36","author":[{"given":"Philip","family":"Stenner","sequence":"first","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Bernhard","family":"Schmidt","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Herbert","family":"Bruder","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Thomas","family":"Allmendinger","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Ulrike","family":"Haberland","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Thomas","family":"Flohr","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Marc","family":"Kachelrie\u00df","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]}],"member":"311","published-online":{"date-parts":[[2009,11,24]]},"reference":[{"key":"e_1_2_7_2_1","doi-asserted-by":"publisher","DOI":"10.1007\/s00330\u2010005\u20102919\u20102"},{"key":"e_1_2_7_3_1","unstructured":"2009 SPIE Bellingham WA H. Bruder M. Petersilka H. Mehldau W. Heidinger B. T. Schmidt C. Thierfeld K. Stierstorfer T. Flohr E. Samei J. Hsieh Flash imaging in dual source CT (DSCT) 72580D"},{"key":"e_1_2_7_4_1","doi-asserted-by":"publisher","DOI":"10.1118\/1.2207236"},{"key":"e_1_2_7_5_1","doi-asserted-by":"publisher","DOI":"10.3109\/9780203089743-33"},{"key":"e_1_2_7_6_1","doi-asserted-by":"publisher","DOI":"10.1259\/bjr\/13564625"},{"key":"e_1_2_7_7_1","doi-asserted-by":"publisher","DOI":"10.1109\/TVCG.2007.70569"},{"key":"e_1_2_7_8_1","doi-asserted-by":"publisher","DOI":"10.3109\/9780203089743-34"},{"key":"e_1_2_7_9_1","doi-asserted-by":"publisher","DOI":"10.1118\/1.2805476"},{"key":"e_1_2_7_10_1","doi-asserted-by":"publisher","DOI":"10.1016\/j.acra.2005.08.007"},{"key":"e_1_2_7_11_1","volume-title":"European Congress of Radiology","author":"Raupach R.","year":"2009"},{"key":"e_1_2_7_12_1","doi-asserted-by":"publisher","DOI":"10.1118\/1.3013700"},{"key":"e_1_2_7_13_1","doi-asserted-by":"publisher","DOI":"10.1118\/1.1561286"},{"key":"e_1_2_7_14_1","volume-title":"SPIE Medical Imaging Proceedings","author":"Bruder H.","year":"2003"},{"key":"e_1_2_7_15_1","doi-asserted-by":"publisher","DOI":"10.1088\/0031\u20109155\/37\/7\/010"},{"key":"e_1_2_7_16_1","doi-asserted-by":"publisher","DOI":"10.1118\/1.2820901"},{"key":"e_1_2_7_17_1","doi-asserted-by":"publisher","DOI":"10.1088\/0031\u20109155\/52\/23\/014"},{"key":"e_1_2_7_18_1","doi-asserted-by":"publisher","DOI":"10.1097\/RLI.0b013e3181afbb03"},{"key":"e_1_2_7_19_1","unstructured":"2008 SPIE Bellingham WA H. Bruder K. Stierstorfer M. Petersilka C. Wiegand C. Suess T. Flohr J. Hsieh E. Samei Correction of cross\u2010scatter in next generation dual source CT (DSCT) scanners 69131W"}],"container-title":["Medical Physics"],"original-title":[],"language":"en","link":[{"URL":"https:\/\/api.wiley.com\/onlinelibrary\/tdm\/v1\/articles\/10.1118%2F1.3259734","content-type":"application\/pdf","content-version":"vor","intended-application":"text-mining"},{"URL":"https:\/\/api.wiley.com\/onlinelibrary\/tdm\/v1\/articles\/10.1118%2F1.3259734","content-type":"unspecified","content-version":"vor","intended-application":"text-mining"},{"URL":"https:\/\/aapm.onlinelibrary.wiley.com\/doi\/pdf\/10.1118\/1.3259734","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2023,10,3]],"date-time":"2023-10-03T11:39:53Z","timestamp":1696333193000},"score":1,"resource":{"primary":{"URL":"https:\/\/aapm.onlinelibrary.wiley.com\/doi\/10.1118\/1.3259734"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[2009,11,24]]},"references-count":18,"journal-issue":{"issue":"12","published-print":{"date-parts":[[2009,12]]}},"alternative-id":["10.1118\/1.3259734"],"URL":"https:\/\/doi.org\/10.1118\/1.3259734","archive":["Portico"],"relation":{},"ISSN":["0094-2405","2473-4209"],"issn-type":[{"value":"0094-2405","type":"print"},{"value":"2473-4209","type":"electronic"}],"subject":[],"published":{"date-parts":[[2009,11,24]]}}}