{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2026,3,31]],"date-time":"2026-03-31T10:16:05Z","timestamp":1774952165057,"version":"3.50.1"},"reference-count":5,"publisher":"BMJ","issue":"4","content-domain":{"domain":["bmj.com"],"crossmark-restriction":true},"short-container-title":["Arch Dis Child Fetal Neonatal Ed"],"accepted":{"date-parts":[[2020,12,14]]},"published-print":{"date-parts":[[2021,7]]},"abstract":"<jats:p>We wished to determine the accuracy of thermometers used to measure temperature in newborn infants. We measured the temperature of a water bath with three types of thermometer set at 0.5\u00b0C increments between 32.5\u00b0C and 38.5\u00b0C and compared the values to a control. We recorded the time to display steady-state temperature. The Microlife thermometer most closely approximated control temperature (mean difference &lt;0.1\u00b0C (SD&lt;0.1\u00b0C)) and displayed a reading in a mean time of 29 s (SD 2 s). Used in \u2018predictive\u2019 (default) mode, the Welch Allyn SureTemp Plus 692 thermometer differed from the control by a mean of 0.6\u00b0C (SD 0.3\u00b0C), displaying a temperature at 15 s (SD 3 s). This device consistently overestimated temperature. In \u2018continuous\u2019 mode, the mean difference was &lt;0.1\u00b0C (SD&lt;0.1\u00b0C) at 5\u2009min. The Phillips probe differed from the control by a mean of 0.4\u00b0C (SD 0.2\u00b0C). Thermometers used to measure temperature in newborn infants may underestimate hypothermia. A prospective study in newborn infants is needed.<\/jats:p>","DOI":"10.1136\/archdischild-2020-320123","type":"journal-article","created":{"date-parts":[[2020,12,28]],"date-time":"2020-12-28T13:20:55Z","timestamp":1609161655000},"page":"435-437","update-policy":"https:\/\/doi.org\/10.1136\/crossmarkpolicy","source":"Crossref","is-referenced-by-count":5,"title":["Bench-top comparison of thermometers used in newborn infants"],"prefix":"10.1136","volume":"106","author":[{"ORCID":"https:\/\/orcid.org\/0000-0001-6618-4402","authenticated-orcid":false,"given":"Emma A","family":"Dunne","sequence":"first","affiliation":[{"name":"Department of Neonatology, National Maternity Hospital, Dublin, Ireland"},{"name":"School of Medicine, University College Dublin, Dublin, Ireland"}],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Katie","family":"Cunningham","sequence":"additional","affiliation":[{"name":"Department of Neonatology, National Maternity Hospital, Dublin, Ireland"},{"name":"School of Medicine, University College Dublin, Dublin, Ireland"}],"role":[{"role":"author","vocabulary":"crossref"}]},{"ORCID":"https:\/\/orcid.org\/0000-0001-7453-5452","authenticated-orcid":false,"given":"Colm Patrick Finbarr","family":"O'Donnell","sequence":"additional","affiliation":[{"name":"Department of Neonatology, National Maternity Hospital, Dublin, Ireland"},{"name":"School of Medicine, University College Dublin, Dublin, Ireland"}],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Lisa K","family":"McCarthy","sequence":"additional","affiliation":[{"name":"Department of Neonatology, National Maternity Hospital, Dublin, Ireland"},{"name":"School of Medicine, University College Dublin, Dublin, Ireland"}],"role":[{"role":"author","vocabulary":"crossref"}]}],"member":"239","published-online":{"date-parts":[[2020,12,28]]},"reference":[{"key":"2025090510384310000_106.4.435.1","unstructured":"World Health Organization . Maternal and newborn Health\/Safe motherhood. thermal protection of the newborn: a practical guide. Geneva: WHO, 1997."},{"key":"2025090510384310000_106.4.435.2","doi-asserted-by":"crossref","first-page":"e643","DOI":"10.1542\/peds.2006-0943","article-title":"Admission temperature of low birth weight infants: predictors and associated morbidities","volume":"119","author":"Laptook","year":"2007","journal-title":"Pediatrics"},{"key":"2025090510384310000_106.4.435.3","doi-asserted-by":"crossref","first-page":"199","DOI":"10.1080\/03091900600711464","article-title":"Which thermometer? factors influencing best choice for intermittent clinical temperature assessment","volume":"30","author":"Crawford","year":"2006","journal-title":"J Med Eng Technol"},{"key":"2025090510384310000_106.4.435.4","unstructured":"ASTM . Standard specification for electronic thermometer for intermittent determination of patient temperature. ASTM International, West Conshohocken, PA, 2018."},{"key":"2025090510384310000_106.4.435.5","unstructured":"Baura G . Thermometers. In: Medical device technologies: a systems based overview using engineering standards. Waltham, MA, USA: Academic Press, 2011."}],"container-title":["Archives of Disease in Childhood - Fetal and Neonatal Edition"],"original-title":[],"language":"en","link":[{"URL":"https:\/\/syndication.highwire.org\/content\/doi\/10.1136\/archdischild-2020-320123","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2025,9,5]],"date-time":"2025-09-05T17:59:28Z","timestamp":1757095168000},"score":1,"resource":{"primary":{"URL":"https:\/\/fn.bmj.com\/lookup\/doi\/10.1136\/archdischild-2020-320123"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[2020,12,28]]},"references-count":5,"journal-issue":{"issue":"4","published-online":{"date-parts":[[2021,6,18]]},"published-print":{"date-parts":[[2021,7]]}},"alternative-id":["10.1136\/archdischild-2020-320123"],"URL":"https:\/\/doi.org\/10.1136\/archdischild-2020-320123","relation":{},"ISSN":["1359-2998","1468-2052"],"issn-type":[{"value":"1359-2998","type":"print"},{"value":"1468-2052","type":"electronic"}],"subject":[],"published":{"date-parts":[[2020,12,28]]}}}