{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2025,10,22]],"date-time":"2025-10-22T17:46:48Z","timestamp":1761155208925,"version":"3.44.0"},"reference-count":15,"publisher":"BMJ","issue":"3","content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":["Arch Dis Child Fetal Neonatal Ed"],"published-print":{"date-parts":[[1995,11]]},"abstract":"<jats:p>A trial of the Nellcor Stat Cap, which detects exhaled carbon dioxide, as an aid to determining whether endotracheal tubes are placed in the oesophagus or the trachea, was carried out in the neonatal unit of this hospital. Twenty five neonates, with a mean (range) gestational age of 33 (24-42) weeks and a mean (range) birthweight of 2.17 (0.54-4.1) kg were studied over two months. These babies underwent a total of 58 intubations and 20 suspected self-extubations. The Nellcor Stat Cap was easy to use. It confirmed clinical findings on 20\/20 occasions when the endotracheal tube was in the oesophagus and on 57\/58 (98%) occasions when it was in the trachea. On 14\/78 (19%) occasions the machine provided helpful additional information in reaching a decision on the adequacy of tube placement. Failure to detect rapidly accidental oesophageal intubation or unintentional tracheal extubation can result in rapid deterioration of the condition of ventilated newborns. The machine is a valuable aid to intubation and rapid diagnosis of self-extubation.<\/jats:p>","DOI":"10.1136\/fn.73.3.f184","type":"journal-article","created":{"date-parts":[[2008,10,31]],"date-time":"2008-10-31T17:55:51Z","timestamp":1225475751000},"page":"F184-F186","source":"Crossref","is-referenced-by-count":5,"title":["Nellcor Stat Cap differentiates oesophageal from tracheal intubation."],"prefix":"10.1136","volume":"73","author":[{"given":"P. D.","family":"Sutherland","sequence":"first","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"M.","family":"Quinn","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]}],"member":"239","published-online":{"date-parts":[[1995,11,1]]},"reference":[{"key":"ref_1","first-page":"886","article-title":"Esophageal intubation. A review of detection techniques","volume":"65","author":"Birmingham, P.K.; Cheney, F.W.; Ward, R.J.","year":"1986","unstructured":"Birmingham PK, Cheney FW, Ward RJ. Esophageal intubation. A review of detection techniques. Anesth Analg 1986; 65: 886-91.","journal-title":"Anesth Analg"},{"key":"ref_2","doi-asserted-by":"crossref","first-page":"232","DOI":"10.1007\/BF01619265","article-title":"Auscultation cannot distinguish esophageal from tracheal passage of tube","volume":"7","author":"Mizutani, A.R.; Ozaki, G.; Benumof, J.L.; Scheller, M.S.","year":"1991","unstructured":"Mizutani AR, Ozaki G, Benumof JL, Scheller MS. Auscultation cannot distinguish esophageal from tracheal passage of tube. J Clin Monit 1991; 7: 232-6.","journal-title":"J Clin Monit"},{"key":"ref_3","doi-asserted-by":"crossref","first-page":"608","DOI":"10.1177\/0310057X9302100519","article-title":"Oesophageal intubation: An analysis of 2000 patients","volume":"21","author":"Holland, R.; Webb, R.K.; Runciman, W.B.","year":"1993","unstructured":"Holland R, Webb RK, Runciman WB. Oesophageal intubation: An analysis of 2000 patients. Anaesth Intens Care 1993; 21: 608-10.","journal-title":"Anaesth Intens Care"},{"key":"ref_4","doi-asserted-by":"crossref","first-page":"183","DOI":"10.1177\/0310057X8000800215","article-title":"Accidental intubation of the oesophagus","volume":"8","author":"Pollard, B.J.; Junius, F.","year":"1980","unstructured":"Pollard BJ, Junius F. Accidental intubation of the oesophagus. Anasthes Intens Care 1980; 8: 183-6.","journal-title":"Anasthes Intens Care"},{"key":"ref_5","doi-asserted-by":"crossref","first-page":"984","DOI":"10.1111\/j.1365-2044.1980.tb04997.x","article-title":"Signs of endotracheal intubation","volume":"35","author":"Howells, T.H.; Riethmuller, R.J.","year":"1980","unstructured":"Howells TH, Riethmuller RJ. Signs of endotracheal intubation. Anaesthesia 1980; 35: 984-6.","journal-title":"Anaesthesia"},{"key":"ref_6","doi-asserted-by":"crossref","first-page":"344","DOI":"10.1097\/00000542-198310000-00013","article-title":"Early detection of endotracheal tube accidents by monitoring CO2 in respiratory gas","volume":"59","author":"Murray, I.P.; Modell, J.H.","year":"1983","unstructured":"Murray IP, Modell JH. Early detection of endotracheal tube accidents by monitoring CO2 in respiratory gas. Anaesthesiol 1983; 59: 344-6.","journal-title":"Anaesthesiol"},{"key":"ref_7","doi-asserted-by":"crossref","first-page":"199","DOI":"10.1111\/j.1399-6576.1983.tb01934.x","article-title":"Capnography for detection of accidental oesophageal intubation","volume":"27","author":"Linko, K.; Paloheimo, M.; Tammisto, T.","year":"1983","unstructured":"Linko K, Paloheimo M, Tammisto T. Capnography for detection of accidental oesophageal intubation. Acta Anaesthesiol Scand 1983; 27: 199-202.","journal-title":"Acta Anaesthesiol Scand"},{"key":"ref_8","first-page":"47","article-title":"Validity of a disposable end-tidal carbon dioxide detector in verifying endotracheal tube placement in infants and children","volume":"21","author":"Bhende, M.S.; Thompson, A.E.; Cook, D.R.; Saville, A.I.","year":"1992","unstructured":"Bhende MS, Thompson AE, Cook DR, Saville AI. Validity of a disposable end-tidal carbon dioxide detector in verifying endotracheal tube placement in infants and children. Ann Emerg Med 1992; 21: 47-50.","journal-title":"Ann Emerg Med"},{"key":"ref_9","doi-asserted-by":"crossref","first-page":"1042","DOI":"10.1542\/peds.89.6.1042","article-title":"Utility of an end-tidal carbon dioxide detector during stabilization and transport of critically ill children","volume":"89","author":"Bhende, M.S.; Thompson, A.E.; Orr, R.A.","year":"1992","unstructured":"Bhende MS, Thompson AE, Orr RA. Utility of an end-tidal carbon dioxide detector during stabilization and transport of critically ill children. Pediatrics 1992; 89: 1042-4.","journal-title":"Pediatrics"},{"key":"ref_10","doi-asserted-by":"crossref","first-page":"887","DOI":"10.1093\/bja\/59.7.877","article-title":"Pitfalls in anaesthetic practice","volume":"59","author":"Utting, J.E.","year":"1987","unstructured":"Utting JE. Pitfalls in anaesthetic practice. Br J Anaesth 1987; 59: 887-90.","journal-title":"Br J Anaesth"},{"key":"ref_11","first-page":"43","article-title":"The spectrum of medical liability cases","volume-title":"Anaesthesia Clinics","author":"Solazzi, R.W.; Ward, R.J.","year":"1984","unstructured":"Solazzi RW, Ward RJ. The spectrum of medical liability cases. In: Pierce EC, Cooper JB, eds. International Anaesthesia Clinics. Boston: Little, Brown and Company, 1984: 43-59."},{"key":"ref_12","doi-asserted-by":"crossref","first-page":"259","DOI":"10.1016\/0300-9572(78)90006-0","article-title":"The capnogram as a guide to the efficiency of cardiac massage","volume":"6","author":"Kalenda, Z.","year":"1978","unstructured":"Kalenda Z. The capnogram as a guide to the efficiency of cardiac massage. Resuscitation 1978; 6: 259-63.","journal-title":"Resuscitation"},{"key":"ref_13","doi-asserted-by":"crossref","first-page":"607","DOI":"10.1056\/NEJM198803103181005","article-title":"End-tidal carbon dioxide concentration during cardio-pulmonary resuscitation","volume":"318","author":"Falk, J.L.; Rackow, E.C.; Weil, M.H.","year":"1988","unstructured":"Falk JL, Rackow EC, Weil MH. End-tidal carbon dioxide concentration during cardio-pulmonary resuscitation. N Engl JMed 1988; 318: 607-11.","journal-title":"N Engl JMed"},{"key":"ref_14","doi-asserted-by":"crossref","first-page":"1285","DOI":"10.1097\/00000542-199012000-00037","article-title":"Tracheal intubation is not invariably confirmed by capnography","volume":"73","author":"Dunn, S.M.; Mushlin, P.S.; Lind, L.J.; Raemer, D.","year":"1990","unstructured":"Dunn SM, Mushlin PS, Lind LJ, Raemer D. Tracheal intubation is not invariably confirmed by capnography. Anaesthesiol 1990; 73: 1285-7.","journal-title":"Anaesthesiol"},{"key":"ref_15","doi-asserted-by":"crossref","first-page":"992","DOI":"10.1097\/00000542-198912000-00029","article-title":"Unusual cause for an absent capnogram","volume":"71","author":"Markovitz, B.P.; Silverburg, M.; Godinez, R.I.","year":"1989","unstructured":"Markovitz BP, Silverburg M, Godinez RI. Unusual cause for an absent capnogram. Anaesthesiol 1989; 71: 992-3.","journal-title":"Anaesthesiol"}],"container-title":["Archives of Disease in Childhood - Fetal and Neonatal Edition"],"original-title":[],"language":"en","link":[{"URL":"https:\/\/syndication.highwire.org\/content\/doi\/10.1136\/fn.73.3.F184","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2025,9,5]],"date-time":"2025-09-05T22:51:25Z","timestamp":1757112685000},"score":1,"resource":{"primary":{"URL":"https:\/\/fn.bmj.com\/lookup\/doi\/10.1136\/fn.73.3.F184"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[1995,11]]},"references-count":15,"journal-issue":{"issue":"3","published-online":{"date-parts":[[1995,11,1]]},"published-print":{"date-parts":[[1995,11]]}},"alternative-id":["10.1136\/fn.73.3.F184"],"URL":"https:\/\/doi.org\/10.1136\/fn.73.3.f184","relation":{},"ISSN":["1359-2998","1468-2052"],"issn-type":[{"type":"print","value":"1359-2998"},{"type":"electronic","value":"1468-2052"}],"subject":[],"published":{"date-parts":[[1995,11]]}}}