{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2025,10,30]],"date-time":"2025-10-30T07:05:26Z","timestamp":1761807926897,"version":"build-2065373602"},"reference-count":8,"publisher":"BMJ","issue":"1","content-domain":{"domain":["bmj.com"],"crossmark-restriction":true},"short-container-title":["Heart"],"accepted":{"date-parts":[[1998,6,24]]},"published-print":{"date-parts":[[1999,1,1]]},"abstract":"<jats:p>Two cases are reported (both men, one 72 and one 54 years old) of inappropriate shocks delivered by an implantable cardiac defibrillator (ICD) device, which oversensed the myopotentials induced by deep breathing and Valsalva manoeuvre. No damage to leads was associated with the oversensing of myopotentials. The mechanism of the inappropriate shocks was determined using real time electrograms. Modification of the duration of ventricular detection and decrease in sensitivity made it possible to avoid the oversensing of myopotentials and to deliver ICD treatment.<\/jats:p>","DOI":"10.1136\/hrt.81.1.94","type":"journal-article","created":{"date-parts":[[2014,10,22]],"date-time":"2014-10-22T15:31:45Z","timestamp":1413991905000},"page":"94-96","update-policy":"https:\/\/doi.org\/10.1136\/crossmarkpolicy","source":"Crossref","is-referenced-by-count":18,"title":["Inappropriate shocks delivered by implantable cardiac defibrillators during oversensing of activity of diaphagmatic muscle"],"prefix":"10.1136","volume":"81","author":[{"given":"D","family":"Babuty","sequence":"first","affiliation":[{"name":"Cardiology B Department, Hospital Trousseau, 37044 Tours, France"}],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"L","family":"Fauchier","sequence":"additional","affiliation":[{"name":"Cardiology B Department, Hospital Trousseau, 37044 Tours, France"}],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"P","family":"Cosnay","sequence":"additional","affiliation":[{"name":"Cardiology B Department, Hospital Trousseau, 37044 Tours, France"}],"role":[{"role":"author","vocabulary":"crossref"}]}],"member":"239","published-online":{"date-parts":[[1999,1,1]]},"reference":[{"key":"2025101800250505000_81.1.94.1","doi-asserted-by":"publisher","DOI":"10.1161\/01.CIR.92.6.1651"},{"key":"2025101800250505000_81.1.94.2","doi-asserted-by":"publisher","DOI":"10.1161\/01.CIR.91.8.2204"},{"key":"2025101800250505000_81.1.94.3","doi-asserted-by":"crossref","first-page":"1756","DOI":"10.1111\/j.1540-8159.1993.tb01049.x","article-title":"Interference from a hand radiofrequency remote control causing discharge of an implantable defibrillator.","volume":"16","author":"Man","year":"1993","journal-title":"PACE"},{"key":"2025101800250505000_81.1.94.4","doi-asserted-by":"crossref","first-page":"665","DOI":"10.1111\/j.1540-8159.1994.tb02401.x","article-title":"Inappropriate discharge by an implantable cardioverter defibrillator: recognition of myopotential sensing using telemetered intracardiac electrograms.","volume":"17","author":"Sandler","year":"1994","journal-title":"PACE"},{"key":"2025101800250505000_81.1.94.5","doi-asserted-by":"crossref","first-page":"1592","DOI":"10.1111\/j.1540-8159.1995.tb06748.x","article-title":"Erroneous discharge of an implantable cardioverter defibrillator caused by an electric razor.","volume":"19","author":"Seifert","year":"1995","journal-title":"PACE"},{"key":"2025101800250505000_81.1.94.6","doi-asserted-by":"crossref","first-page":"2857","DOI":"10.1111\/j.1540-8159.1997.tb05445.x","article-title":"Interaction between electronic article surveillance systems and implantable defibrillators: insights from a fourth generation ICD.","volume":"20","author":"Mathew","year":"1997","journal-title":"PACE"},{"key":"2025101800250505000_81.1.94.7","doi-asserted-by":"crossref","first-page":"1589","DOI":"10.1111\/j.1540-8159.1995.tb06747.x","article-title":"Cardioverter discharges following sensing of electrical artifact due to fluid penetration in the connector port.","volume":"18","author":"Hief","year":"1995","journal-title":"PACE"},{"key":"2025101800250505000_81.1.94.8","doi-asserted-by":"crossref","first-page":"124","DOI":"10.1111\/j.1540-8159.1996.tb04802.x","article-title":"Inhibition of bradycardia pacing caused by far-field atrial sensing in a third-generation cardioverter defibrillator with an automatic gain feature.","volume":"19","author":"Curwin","year":"1996","journal-title":"PACE"}],"container-title":["Heart"],"original-title":[],"language":"en","link":[{"URL":"https:\/\/syndication.highwire.org\/content\/doi\/10.1136\/hrt.81.1.94","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2025,10,18]],"date-time":"2025-10-18T07:25:14Z","timestamp":1760772314000},"score":1,"resource":{"primary":{"URL":"https:\/\/heart.bmj.com\/lookup\/doi\/10.1136\/hrt.81.1.94"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[1999,1,1]]},"references-count":8,"journal-issue":{"issue":"1","published-online":{"date-parts":[[1999,1,1]]},"published-print":{"date-parts":[[1999,1,1]]}},"alternative-id":["10.1136\/hrt.81.1.94"],"URL":"https:\/\/doi.org\/10.1136\/hrt.81.1.94","relation":{},"ISSN":["1355-6037","1468-201X"],"issn-type":[{"type":"print","value":"1355-6037"},{"type":"electronic","value":"1468-201X"}],"subject":[],"published":{"date-parts":[[1999,1,1]]}}}