{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2026,9,11]],"date-time":"2026-09-11T13:15:43Z","timestamp":1789132543584,"version":"build-2803163510"},"reference-count":10,"publisher":"BMJ","funder":[{"name":"Epsilon Medical, Inc","award":["N\/A"],"award-info":[{"award-number":["N\/A"]}]}],"content-domain":{"domain":["bmj.com"],"crossmark-restriction":true},"short-container-title":["J NeuroIntervent Surg"],"accepted":{"date-parts":[[2026,5,11]]},"abstract":"<jats:sec>\n                    <jats:title>Background<\/jats:title>\n                    <jats:p>Endovascular electroencephalography (eEEG) using the EP-01 electrode requires reliable venous access and secure fixation to maintain stable positioning. This technical note aims to describe a standardized workflow for EP-01 placement, focusing on preprocedural planning, step-by-step deployment, and fixation strategies.<\/jats:p>\n                  <\/jats:sec>\n                  <jats:sec>\n                    <jats:title>Methods<\/jats:title>\n                    <jats:p>We report 12 consecutive EP-01 placements performed between May 2024 and October 2025. Preprocedural planning utilized venous-phase angiography with three-dimensional reconstructions, as well as contrast-enhanced three-dimensional MRI when available, to assess venous anatomy and select appropriate access routes. Procedures were performed under general anesthesia using jugular venous sheaths and arterial angiographic guidance. EP-01 electrodes were deployed through microcatheters, and signal recordability was confirmed. Microcatheters were withdrawn using an extension wire to maintain electrode position. Electrodes were secured using Tuohy-Borst fixation and additional skin-level fixation with film dressing.<\/jats:p>\n                  <\/jats:sec>\n                  <jats:sec>\n                    <jats:title>Results<\/jats:title>\n                    <jats:p>All 12 procedures were completed successfully (mean age 40.6\u00b111.0 years; 4\u2009women). The mean procedure time was 165.6\u00b135.5\u2009min and the mean fluoroscopy time was 51.8\u00b19.4\u2009min. Six electrodes were placed in 11 of 12 cases, while three electrodes were placed in one case due to venous anatomical limitations. Alternative venous routes were used when standard approaches were not feasible. No procedure-related complications occurred. The median electrode dwell time was 2\u2009days.<\/jats:p>\n                  <\/jats:sec>\n                  <jats:sec>\n                    <jats:title>Conclusions<\/jats:title>\n                    <jats:p>A standardized planning and placement workflow enabled feasible EP-01 electrode placement without procedure-related complications in this consecutive series. Careful preprocedural assessment of patient-specific venous anatomy is essential and may necessitate alternative access strategies.<\/jats:p>\n                  <\/jats:sec>","DOI":"10.1136\/jnis-2026-025499","type":"journal-article","created":{"date-parts":[[2026,5,22]],"date-time":"2026-05-22T17:20:31Z","timestamp":1779470431000},"page":"jnis-2026-025499","update-policy":"https:\/\/doi.org\/10.1136\/crossmarkpolicy","source":"Crossref","is-referenced-by-count":1,"title":["Step-by-step placement of endovascular electroencephalography using the EP-01 system"],"prefix":"10.1136","author":[{"ORCID":"https:\/\/orcid.org\/0000-0002-0392-6808","authenticated-orcid":false,"given":"Hisayuki","family":"Hosoo","sequence":"first","affiliation":[{"name":"Division of Stroke Prevention and Treatment, Department of Neurosurgery, Institute of Medicine, University of Tsukuba, Tsukuba, Ibaraki, Japan"},{"name":"Department of Stroke and Cerebrovascular Diseases, University of Tsukuba Hospital, Tsukuba, Ibaraki, Japan"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Shun","family":"Tanaka","sequence":"additional","affiliation":[{"name":"Department of Stroke and Cerebrovascular Diseases, University of Tsukuba Hospital, Tsukuba, Ibaraki, Japan"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Kota","family":"Araki","sequence":"additional","affiliation":[{"name":"Department of Neurosurgery, Institute of Medicine, University of Tsukuba, Tsukuba, Ibaraki, Japan"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Yosuke","family":"Masuda","sequence":"additional","affiliation":[{"name":"Department of Neurosurgery, Institute of Medicine, University of Tsukuba, Tsukuba, Ibaraki, Japan"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Eiichi","family":"Ishikawa","sequence":"additional","affiliation":[{"name":"Department of Neurosurgery, Institute of Medicine, University of Tsukuba, Tsukuba, Ibaraki, Japan"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Yuji","family":"Matsumaru","sequence":"additional","affiliation":[{"name":"Department of Neurosurgery, Institute of Medicine, University of Tsukuba, Tsukuba, Ibaraki, Japan"},{"name":"Epsilon Medical Inc, Chuo-ku, Tokyo, Japan"}],"role":[{"vocabulary":"crossref","role":"author"}]}],"member":"239","published-online":{"date-parts":[[2026,5,22]]},"reference":[{"key":"2026052210201980000_jnis-2026-025499v1.1","doi-asserted-by":"crossref","DOI":"10.1016\/j.ebr.2025.100748","article-title":"First experiences with multiple bilateral insertions of a newly developed microcatheter-compatible endovascular electroencephalogram electrode for humans with epilepsy","volume":"29","author":"Masuda","year":"2025","journal-title":"Epilepsy Behav Rep"},{"key":"2026052210201980000_jnis-2026-025499v1.2","doi-asserted-by":"crossref","unstructured":"Araki K , Masuda Y , Fujimoto A , et al . Detection of interictal epileptiform discharges using multiple bilateral insertions of a newly developed microcatheter-compatible endovascular electroencephalogram electrode: a clinical feasibility trial. Epilepsia Open 2025. doi:10.1002\/epi4.70166","DOI":"10.1002\/epi4.70166"},{"key":"2026052210201980000_jnis-2026-025499v1.3","doi-asserted-by":"crossref","DOI":"10.1371\/journal.pone.0332387","article-title":"Endovascular EEG device prospective multicenter single-arm clinical trial to confirm efficacy and safety performance on patients with intractable epilepsy: the EPSILON IE trial protocol","volume":"20","author":"Matsumaru","year":"2025","journal-title":"PLoS One"},{"key":"2026052210201980000_jnis-2026-025499v1.4","doi-asserted-by":"publisher","DOI":"10.3174\/ajnr.A0489"},{"key":"2026052210201980000_jnis-2026-025499v1.5","doi-asserted-by":"publisher","DOI":"10.1177\/1753425915590067"},{"key":"2026052210201980000_jnis-2026-025499v1.6","doi-asserted-by":"crossref","first-page":"547","DOI":"10.5797\/jnet.ra.2020-0086","article-title":"Venous Anatomy of the Cavernous Sinus and Relevant Veins","volume":"14","author":"Tanoue","year":"2020","journal-title":"J Neuroendovasc Ther"},{"key":"2026052210201980000_jnis-2026-025499v1.7","doi-asserted-by":"crossref","first-page":"1103","DOI":"10.1016\/j.acra.2010.04.020","article-title":"Anatomical evaluation of the dural sinuses in the region of the torcular herophili using three dimensional CT venography","volume":"17","author":"Fukusumi","year":"2010","journal-title":"Acad Radiol"},{"key":"2026052210201980000_jnis-2026-025499v1.8","doi-asserted-by":"publisher","DOI":"10.1007\/s00276-013-1223-0"},{"key":"2026052210201980000_jnis-2026-025499v1.9","doi-asserted-by":"crossref","DOI":"10.1097\/MD.0000000000002862","article-title":"Diagnosis of transverse sinus hypoplasia in magnetic resonance venography: new insights based on magnetic resonance imaging in combined dataset of venous outflow impairment case-control studies: post hoc case-control study","volume":"95","author":"Han","year":"2016","journal-title":"Medicine (Baltimore)"},{"key":"2026052210201980000_jnis-2026-025499v1.10","doi-asserted-by":"crossref","DOI":"10.5797\/jnet.sr.2024-0099","article-title":"The Qualification Examination for Specialists and Instructors in the Japanese Society of Neuroendovascular Therapy: History and Current Status","volume":"19","author":"Yoshimura","year":"2025","journal-title":"J Neuroendovasc Ther"}],"container-title":["Journal of NeuroInterventional Surgery"],"original-title":[],"language":"en","link":[{"URL":"https:\/\/syndication.highwire.org\/content\/doi\/10.1136\/jnis-2026-025499","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2026,5,22]],"date-time":"2026-05-22T17:20:32Z","timestamp":1779470432000},"score":1,"resource":{"primary":{"URL":"https:\/\/jnis.bmj.com\/lookup\/doi\/10.1136\/jnis-2026-025499"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[2026,5,22]]},"references-count":10,"alternative-id":["10.1136\/jnis-2026-025499"],"URL":"https:\/\/doi.org\/10.1136\/jnis-2026-025499","relation":{},"ISSN":["1759-8478","1759-8486"],"issn-type":[{"value":"1759-8478","type":"print"},{"value":"1759-8486","type":"electronic"}],"subject":[],"published":{"date-parts":[[2026,5,22]]}}}