{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2025,11,6]],"date-time":"2025-11-06T19:49:00Z","timestamp":1762458540573},"reference-count":19,"publisher":"British Editorial Society of Bone & Joint Surgery","issue":"4","license":[{"start":{"date-parts":[[2007,4,1]],"date-time":"2007-04-01T00:00:00Z","timestamp":1175385600000},"content-version":"tdm","delay-in-days":0,"URL":"https:\/\/online.boneandjoint.org.uk\/TDM"}],"content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":["The Journal of Bone and Joint Surgery. British volume"],"published-print":{"date-parts":[[2007,4]]},"abstract":"<jats:p> During open reduction of an irreducible anterior dislocation of a total hip replacement with an Oxinium femoral head, it was observed that the head had been significantly damaged. Gross and scanning electron microscopic examination revealed cracking, gouging, and delamination of the surface. Because of the risk which this poses for damaging the polyethylene acetabular liner, it is strongly recommended that patients with this type of prosthetic head be carefully monitored after a dislocation. <\/jats:p>","DOI":"10.1302\/0301-620x.89b4.17557","type":"journal-article","created":{"date-parts":[[2007,4,26]],"date-time":"2007-04-26T17:02:50Z","timestamp":1177606970000},"page":"535-537","source":"Crossref","is-referenced-by-count":55,"title":["Surface damage to an Oxinium femoral head prosthesis after dislocation"],"prefix":"10.1302","volume":"89-B","author":[{"given":"G. T.","family":"Evangelista","sequence":"first","affiliation":[{"name":"Musculoskeletal Research Center, NYU-Hospital for Joint Diseases, 301 East 17th Street, New York 10003, USA."}],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"E.","family":"Fulkerson","sequence":"additional","affiliation":[{"name":"Musculoskeletal Research Center, NYU-Hospital for Joint Diseases, 301 East 17th Street, New York 10003, USA."}],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"F.","family":"Kummer","sequence":"additional","affiliation":[{"name":"Musculoskeletal Research Center, NYU-Hospital for Joint Diseases, 301 East 17th Street, New York 10003, USA."}],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"P. E.","family":"Di Cesare","sequence":"additional","affiliation":[{"name":"Department of Orthopaedic Surgery, UC Davis Medical Center, 4860, Y Street, Suite 3800, Sacramento, California 95817, USA."}],"role":[{"role":"author","vocabulary":"crossref"}]}],"member":"42","reference":[{"key":"R1","doi-asserted-by":"publisher","DOI":"10.1302\/0301-620X.86B4.14249"},{"key":"R2","doi-asserted-by":"crossref","unstructured":"Amstutz HC, Campbell P, Kossovsky N, Clarke IC. Mechanism and clinical significance of wear debris induced osteolysis. Clin Orthop 1992;276:7\u201318.","DOI":"10.1097\/00003086-199203000-00003"},{"key":"R3","unstructured":"www.strongasanox.com. Smith &\n Nephew Oxinium femoral heads. Orthopedics technical information (date last accessed 3 April 2007)."},{"key":"R4","doi-asserted-by":"publisher","DOI":"10.1097\/00003086-200010000-00006"},{"key":"R5","doi-asserted-by":"crossref","unstructured":"Urban JA, Garvin KL, Boese CK, et al. Ceramic-on-polyethylene bearing surfaces in total hip arthroplasty: seventeen to twenty-one-year results. J Bone Joint Surg [Am] 2001;83-A:1688\u201394.","DOI":"10.2106\/00004623-200111000-00011"},{"key":"R6","doi-asserted-by":"crossref","unstructured":"Kim YH. Comparison of polyethylene wear associated with cobalt-chromium and zirconia heads after total hip replacement: a prospective, randomized study. J Bone Joint Surg [Am] 2005;87-A:1769\u201376.","DOI":"10.2106\/00004623-200508000-00015"},{"key":"R7","doi-asserted-by":"crossref","unstructured":"Heisel C, Silva M, Schmalzried TP. Bearing surface options for total hip replacement in young patients. Instr Course Lect 2004;53:49\u201365.","DOI":"10.1097\/01.blo.0000149995.84350.d7"},{"key":"R8","doi-asserted-by":"crossref","unstructured":"Good V, Ries M, Barrack RL, et al. Reduced wear with oxidized zirconium femoral heads. J Bone Joint Surg [Am] 2003;85-A:105\u201310.","DOI":"10.2106\/00004623-200300004-00013"},{"key":"R9","doi-asserted-by":"crossref","unstructured":"Hermida JC, Bergula A, Chen P, Colwell CW Jr, D\u2019Lima DD. Comparison of the wear rates of twenty-eight and thirty-two-millimeter femoral heads on cross-linked polyethylene acetabular cups in a wear simulator. J Bone Joint Surg [Am] 2003;85-A:2325\u201331.","DOI":"10.2106\/00004623-200312000-00009"},{"key":"R10","doi-asserted-by":"publisher","DOI":"10.2106\/00004623-199072040-00008"},{"key":"R11","unstructured":"Davidson JA, Asgian CM, Mishra AK, Kovacs P. Zirconia (ZrO2)-coated zirconium-2.5Nb alloy for prosthetic knee bearing applications. In: Yamamuro T, Kokubo T, Nakamura T, eds. Bioceramics. Vol 5. Kyoto: Kobunshi Kankokai, 1992:389\u2013401."},{"key":"R12","unstructured":"Hunter G, Long M. Abrasive wear of oxidized Xr-2.5b, CoCrMo, and Ti-6A1-4V against bone cement. In: Procs sixth world biomaterials congress. 2000:835."},{"key":"R13","unstructured":"Hunter G. Adhesion testing of oxidized zirconium. Trans Soc Biomater 2001;24:540."},{"key":"R14","doi-asserted-by":"crossref","unstructured":"Spector BM, Ries MD, Bourne RB, et al. Wear performance of ultra-high molecular weight polyethylene on oxidized zirconium total knee femoral components. J Bone Joint Surg [Am] 2001;83-A(Suppl 2):80\u20136.","DOI":"10.2106\/00004623-200100022-00004"},{"key":"R15","doi-asserted-by":"publisher","DOI":"10.1097\/01.blo.0000148576.70780.13"},{"key":"R16","doi-asserted-by":"crossref","unstructured":"Ries MD, Salehi A, Widding K, Hunter G. Polyethylene wear performance of oxidized zirconium and cobalt-chromium knee components under abrasive conditions. J Bone Joint Surg [Am] 2002;84-A:129\u201335.","DOI":"10.2106\/00004623-200200002-00018"},{"key":"R17","doi-asserted-by":"publisher","DOI":"10.1097\/01.blo.0000193813.08458.e2"},{"key":"R18","doi-asserted-by":"crossref","unstructured":"Yamamoto T, Saito M, Ueno M, et al. Wear analysis of retrieved ceramic-on-ceramic articulations in total hip arthroplasty: femoral head makes contact with the rim of the socket outside of the bearing surface. J Biomed Mater Res B Appl Biomater 2005;73:301\u20137.","DOI":"10.1002\/jbm.b.30215"},{"key":"R19","doi-asserted-by":"crossref","unstructured":"Kim YH, Ritchie A, Hardaker C. Surface roughness of ceramic femoral heads after in vivo transfer of metal: correlation to polyethylene wear. J Bone Joint Surg [Am] 2005;87-A:577\u201382.","DOI":"10.2106\/JBJS.D.01790"}],"container-title":["The Journal of Bone and Joint Surgery. British volume"],"original-title":[],"language":"en","link":[{"URL":"https:\/\/online.boneandjoint.org.uk\/doi\/pdf\/10.1302\/0301-620X.89B4.17557","content-type":"application\/pdf","content-version":"vor","intended-application":"text-mining"},{"URL":"https:\/\/online.boneandjoint.org.uk\/doi\/full-xml\/10.1302\/0301-620X.89B4.17557","content-type":"application\/xml","content-version":"vor","intended-application":"text-mining"},{"URL":"https:\/\/online.boneandjoint.org.uk\/doi\/pdf\/10.1302\/0301-620X.89B4.17557","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2021,8,16]],"date-time":"2021-08-16T20:36:29Z","timestamp":1629146189000},"score":1,"resource":{"primary":{"URL":"https:\/\/online.boneandjoint.org.uk\/doi\/10.1302\/0301-620X.89B4.17557"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[2007,4]]},"references-count":19,"journal-issue":{"issue":"4","published-print":{"date-parts":[[2007,4]]}},"alternative-id":["10.1302\/0301-620X.89B4.17557"],"URL":"https:\/\/doi.org\/10.1302\/0301-620x.89b4.17557","relation":{},"ISSN":["0301-620X","2044-5377"],"issn-type":[{"value":"0301-620X","type":"print"},{"value":"2044-5377","type":"electronic"}],"subject":[],"published":{"date-parts":[[2007,4]]}}}