{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2026,11,4]],"date-time":"2026-11-04T11:38:01Z","timestamp":1793792281107,"version":"4.6.1"},"reference-count":25,"publisher":"Oxford University Press (OUP)","issue":"5","content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":[],"published-print":{"date-parts":[[2013,5,1]]},"abstract":"<jats:title>Abstract<\/jats:title><jats:p>Hematology analyzers generate suspect flags that involve microscopic reviews to confirm the presence of pathologic cells. This study investigated the reliability of the blast flag in a side-by-side evaluation of 3 Sysmex XE-5000 instruments (Sysmex, Kobe, Japan). The repeatability of the Q values reported by each instrument for 10 replicates of the same blood samples was low (intraclass correlation coefficient [ICC] values, 0.62\u20130.74). The reproducibility of the Q values obtained by analyzing 408 samples on all 3 instruments was reasonable (ICC value, 0.85). In addition, a systematic difference was observed among the instruments in the level of reported Q values. With cutoff commonly being 100, the observed reproducibility of the blast flagging among the instruments was evaluated as poor (\u03ba = 0.73). Based on the observed low performances, we question the usefulness of the Q value as a predictor of blasts and whether a blast flag reported by the XE-5000 is sufficient as a criterion for performing a microscopic review.<\/jats:p>","DOI":"10.1309\/ajcpduzvrn5vy9wz","type":"journal-article","created":{"date-parts":[[2013,4,17]],"date-time":"2013-04-17T20:33:19Z","timestamp":1366230799000},"page":"633-640","source":"Crossref","is-referenced-by-count":17,"title":["The Usefulness of Blast Flags on the Sysmex XE-5000 Is Questionable"],"prefix":"10.1093","volume":"139","author":[{"given":"Heidi","family":"Eilertsen","sequence":"first","affiliation":[{"name":"1Division of Diagnostics and Technology, Akershus University Hospital, Lorenskog, Norway"},{"name":"3Faculty of Health Sciences, Oslo and Akershus University College of Applied Sciences, Oslo"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Nina K.","family":"V\u00f8llestad","sequence":"additional","affiliation":[{"name":"4Department of Health Sciences, Institute of Health and Society, University of Oslo, Oslo"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Tor-Arne","family":"Hagve","sequence":"additional","affiliation":[{"name":"1Division of Diagnostics and Technology, Akershus University Hospital, Lorenskog, Norway"},{"name":"2Institute of Clinical Medicine, University of Oslo, Oslo, Norway"}],"role":[{"vocabulary":"crossref","role":"author"}]}],"member":"286","published-online":{"date-parts":[[2013,5,1]]},"reference":[{"key":"2023012006411365500_b1-1390633","doi-asserted-by":"crossref","first-page":"83","DOI":"10.1532\/LH96.05019","article-title":"The International Consensus Group for Hematology review: suggested criteria for action following automated CBC and WBC differential analysis","volume":"11","author":"Barnes","year":"2005","journal-title":"Lab Hematol"},{"key":"2023012006411365500_b2-1390633","first-page":"8","article-title":"Precision and accuracy of the leukocyte differential on the Sysmex XE-2100","volume":"11","author":"Herklotz","year":"2001","journal-title":"Sysmex J Int"},{"key":"2023012006411365500_b3-1390633","first-page":"18","article-title":"Optimisation of the flagging criteria on Sysmex SE-9500","volume":"10","author":"Gossens","year":"2000","journal-title":"Sysmex J Int"},{"key":"2023012006411365500_b4-1390633","doi-asserted-by":"crossref","first-page":"1528","DOI":"10.5858\/2009-0531-OA.1","article-title":"A method for optimizing and validating institution-specific flagging criteria for automated cell counters","volume":"134","author":"Sireci","year":"2010","journal-title":"Arch Pathol Lab Med"},{"key":"2023012006411365500_b5-1390633","first-page":"22","article-title":"Comparison of the Sysmex XE-2100 to the Abbott Cell-Dyn 4000, automated hematology analyzer","volume":"11","author":"Uptmore","year":"2001","journal-title":"Sysmex J Int"},{"key":"2023012006411365500_b6-1390633","first-page":"58","article-title":"Overview of automated hematology analyzer XE-2100TM","volume":"9","author":"Hiroyuki","year":"1999","journal-title":"Sysmex J Int"},{"key":"2023012006411365500_b7-1390633","author":"Sysmex America Inc"},{"key":"2023012006411365500_b8-1390633","volume-title":"Reference Leukocytes (WBC) Differential Count (Proportional) and Evaluation of Instrumental Methods","author":"Koepke","year":"2007","edition":"2nd ed"},{"key":"2023012006411365500_b9-1390633","article-title":"Consensus rules: consensus guidelines\u2014positive smear findings","author":"International Society for Laboratory Hematology"},{"key":"2023012006411365500_b10-1390633","article-title":"Online kappa calculator","author":"Randolph"},{"key":"2023012006411365500_b11-1390633","volume-title":"Practical Statistics for Medical Research","author":"Altman","year":"1999"},{"key":"2023012006411365500_b12-1390633","first-page":"39","volume-title":"Measures of Interobserver Agreement","author":"Shoukri","year":"2004"},{"key":"2023012006411365500_b13-1390633","doi-asserted-by":"crossref","first-page":"68","DOI":"10.1016\/j.cca.2008.07.018","article-title":"Stability of hematological analytes depends on the hematology analyser used: a stability study with Bayer Advia 120, Beckman Coulter LH 750 and Sysmex XE 2100","volume":"397","author":"Imeri","year":"2008","journal-title":"Clin Chim Acta"},{"key":"2023012006411365500_b14-1390633","doi-asserted-by":"crossref","first-page":"28","DOI":"10.1309\/1EERK1M02QFJRX6P","article-title":"Effects of storage of blood at room temperature on hematologic parameters measured on Sysmex XE-2100","volume":"37","author":"de Baca","year":"2006","journal-title":"Lab Med"},{"key":"2023012006411365500_b15-1390633","doi-asserted-by":"crossref","first-page":"285","DOI":"10.1532\/LH96.05036","article-title":"Performance evaluation of the latest fully automated hematology analyzers in a large, commercial laboratory setting: a 4-way, side-by-side study","volume":"11","author":"Bourner","year":"2005","journal-title":"Lab Hematol"},{"key":"2023012006411365500_b16-1390633","first-page":"230","article-title":"Three-way evaluation of high-throughput hematology analyzers: Beckman Coulter LH 750, Abbott Cell-Dyn 4000, and Sysmex XE-2100","volume":"8","author":"Johnson","year":"2002","journal-title":"Lab Hematol"},{"key":"2023012006411365500_b17-1390633","doi-asserted-by":"crossref","first-page":"347","DOI":"10.1046\/j.1365-2257.1999.00229.x","article-title":"Performance characteristics of blast flagging on the Cell Dyn 4000 haematology analyser","volume":"24","author":"Hoedemakers","year":"1999","journal-title":"Clin Lab Haematol"},{"key":"2023012006411365500_b18-1390633","doi-asserted-by":"crossref","first-page":"480","DOI":"10.1111\/j.1751-553X.2007.00981.x","article-title":"Comparison of four hematology analyzers, CELL-DYN Sapphire, ADVIA 120, Coulter LH 750 and Sysmex XE-2100, in terms of clinical usefulness","volume":"30","author":"Kang","year":"2008","journal-title":"Int J Lab Hematol"},{"key":"2023012006411365500_b19-1390633","doi-asserted-by":"crossref","first-page":"391","DOI":"10.5858\/2001-125-0391-TNHASX","article-title":"The new hematology analyzer Sysmex XE-2100: performance evaluation of a novel white blood cell differential technology","volume":"125","author":"Ruzicka","year":"2001","journal-title":"Arch Pathol Lab Med"},{"key":"2023012006411365500_b20-1390633","doi-asserted-by":"crossref","first-page":"271","DOI":"10.1046\/j.1365-2257.2002.00458.x","article-title":"Performance of the XE-2100 leucocyte differential","volume":"24","author":"Stamminger","year":"2002","journal-title":"Clin Lab Haematol"},{"key":"2023012006411365500_b21-1390633","doi-asserted-by":"crossref","first-page":"309","DOI":"10.1309\/AJCPDLR4KGKAFW4W","article-title":"Improved flagging rates on the Sysmex XE-5000 compared with the XE-2100 reduce the number of manual film reviews and increase laboratory productivity","volume":"136","author":"Briggs","year":"2011","journal-title":"Am J Clin Pathol"},{"key":"2023012006411365500_b22-1390633","first-page":"373","article-title":"Automated five-part white blood cell differential counts","volume":"121","author":"Thalhammer-Scherrer","year":"1997","journal-title":"Arch Pathol Lab Med"},{"key":"2023012006411365500_b23-1390633","first-page":"173","article-title":"A critical evaluation of the manual\/visual differential leukocyte counting method","volume":"11","author":"Koepke","year":"1985","journal-title":"Blood Cells"},{"key":"2023012006411365500_b24-1390633","first-page":"141","article-title":"Statistical reflections on finding atypical cells","volume":"11","author":"Rumke","year":"1985","journal-title":"Blood Cells"},{"key":"2023012006411365500_b25-1390633","doi-asserted-by":"crossref","first-page":"252","DOI":"10.1093\/ajcp\/93.2.252","article-title":"Evaluating diagnostic tests with imperfect standards","volume":"93","author":"Valenstein","year":"1990","journal-title":"Am J Clin Pathol"}],"container-title":["American Journal of Clinical Pathology"],"original-title":[],"language":"en","link":[{"URL":"https:\/\/academic.oup.com\/ajcp\/article-pdf\/139\/5\/633\/48795042\/ajcp_139_5_633.pdf","content-type":"application\/pdf","content-version":"vor","intended-application":"syndication"},{"URL":"https:\/\/academic.oup.com\/ajcp\/article-pdf\/139\/5\/633\/48795042\/ajcp_139_5_633.pdf","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2024,5,9]],"date-time":"2024-05-09T04:54:49Z","timestamp":1715230489000},"score":1,"resource":{"primary":{"URL":"https:\/\/academic.oup.com\/ajcp\/article\/139\/5\/633\/1760975"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[2013,5,1]]},"references-count":25,"journal-issue":{"issue":"5","published-online":{"date-parts":[[2013,5,1]]},"published-print":{"date-parts":[[2013,5,1]]}},"URL":"https:\/\/doi.org\/10.1309\/ajcpduzvrn5vy9wz","relation":{},"ISSN":["1943-7722","0002-9173"],"issn-type":[{"value":"1943-7722","type":"electronic"},{"value":"0002-9173","type":"print"}],"subject":[],"published-other":{"date-parts":[[2013,5,1]]},"published":{"date-parts":[[2013,5,1]]}}}