{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2025,10,27]],"date-time":"2025-10-27T15:23:11Z","timestamp":1761578591687,"version":"3.28.0"},"reference-count":3,"publisher":"Texas Heart Institute Journal","issue":"6","content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":[],"published-print":{"date-parts":[[2014,12]]},"abstract":"<jats:p>We report the case of an 85-year-old woman with severe aortic stenosis who underwent transcatheter aortic valve replacement with use of the Edwards S<jats:sc>apien<\/jats:sc><jats:sup>\u00ae<\/jats:sup> valve system. The procedure was complicated by rupture of the valve-deployment balloon, with separation and retention of the nose cone of the RetroFlex 3<jats:sup>\u00ae<\/jats:sup> delivery system in the iliac artery. Our endovascular retrieval of the equipment was successful, and we achieved access-site hemostasis by deploying a covered stent. To our knowledge, this is the first report of the endovascular retrieval of a malfunctioning delivery system during transcatheter aortic valve replacement.<\/jats:p>","DOI":"10.14503\/thij-13-3447","type":"journal-article","created":{"date-parts":[[2014,12,1]],"date-time":"2014-12-01T16:53:08Z","timestamp":1417452788000},"page":"641-644","source":"Crossref","is-referenced-by-count":6,"title":["Percutaneous Management of RetroFlex 3 Balloon Rupture and Separation of the Edwards Sapien Delivery System"],"prefix":"10.14503","volume":"41","author":[{"given":"Kelly D.","family":"Green","sequence":"first","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Alain","family":"Waked","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Usman","family":"Majeed","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Marat","family":"Fudim","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Mark A.","family":"Robbins","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Marshall","family":"Crenshaw","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"David","family":"Zhao","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]}],"member":"5628","reference":[{"issue":"2","key":"i0730-2347-41-6-641-b1","series-title":"Eur J Cardiothorac Surg","first-page":"437","article-title":"Rupture of the NovaFlex balloon during transcatheter aortic valve implantation and subsequent dissection of the right iliac arteries","volume":"43","year":"2013","unstructured":"Gul\n              M\n            , AksuHU, OzK, BakirI.\n          Rupture of the NovaFlex balloon during transcatheter aortic valve implantation and subsequent dissection of the right iliac arteries. Eur J Cardiothorac Surg2013;43(2):437\u20138."},{"key":"i0730-2347-41-6-641-b2","unstructured":"U.S. Food and Drug Administration\n          . MAUDE adverse event report: Edwards Lifesciences Edwards Sapien transcatheter RetroFlex 3 delivery system [Internet]. Available at: http:\/\/www.accessdata.fda.gov\/scripts\/cdrh\/Cfdocs\/cfmaude\/detail.cfm?mdrfoi__id=2414428 [cited 2013 Feb 14].","article-title":"MAUDE adverse event report: Edwards Lifesciences Edwards Sapien transcatheter RetroFlex 3 delivery system [Internet]"},{"issue":"5","key":"i0730-2347-41-6-641-b3","series-title":"JACC Cardiovasc Interv","first-page":"515","article-title":"Percutaneous management of vascular complications in patients undergoing transcatheter aortic valve implantation","volume":"5","year":"2012","unstructured":"Stortecky\n              S\n            , WenaweserP, DiehmN, PilgrimT, HuberC, RosskopfAB, et al. Percutaneous management of vascular complications in patients undergoing transcatheter aortic valve implantation. JACC Cardiovasc Interv2012;5(5):515\u201324."}],"container-title":["Texas Heart Institute Journal"],"original-title":[],"link":[{"URL":"https:\/\/thij.kglmeridian.com\/view\/journals\/thij\/41\/6\/article-p641.xml","content-type":"text\/html","content-version":"vor","intended-application":"text-mining"},{"URL":"http:\/\/meridian.allenpress.com\/thij\/article-pdf\/41\/6\/641\/1968190\/thij-13-3447.pdf","content-type":"application\/pdf","content-version":"vor","intended-application":"syndication"},{"URL":"https:\/\/thij.kglmeridian.com\/downloadpdf\/journals\/thij\/41\/6\/article-p641.xml","content-type":"text\/html","content-version":"vor","intended-application":"syndication"},{"URL":"https:\/\/thij.kglmeridian.com\/downloadpdf\/journals\/thij\/41\/6\/article-p641.xml","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2024,11,7]],"date-time":"2024-11-07T15:29:26Z","timestamp":1730993366000},"score":1,"resource":{"primary":{"URL":"https:\/\/thij.kglmeridian.com\/view\/journals\/thij\/41\/6\/article-p641.xml"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[2014,12]]},"references-count":3,"journal-issue":{"issue":"6"},"URL":"https:\/\/doi.org\/10.14503\/thij-13-3447","relation":{},"ISSN":["0730-2347","1526-6702"],"issn-type":[{"type":"print","value":"0730-2347"},{"type":"electronic","value":"1526-6702"}],"subject":[],"published":{"date-parts":[[2014,12]]}}}