{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2025,6,28]],"date-time":"2025-06-28T00:10:07Z","timestamp":1751069407154,"version":"3.41.0"},"reference-count":23,"publisher":"ASTM International","issue":"1","content-domain":{"domain":["asmedigitalcollection.asme.org"],"crossmark-restriction":true},"short-container-title":[],"published-print":{"date-parts":[[2022,7,19]]},"abstract":"<jats:title>ABSTRACT<\/jats:title>\n               <jats:p>Pressure gauge calibration refers to comparing the standard and test gauge to determine the error percentage. The reading accuracy of the analog dial pressure gauge is found in the conventional method through manual calculations and then tabulating the values in the calibration certificate. The present study is undertaken to increase the accuracy level of the analog dial pressure gauges. A new automated computer-based method is proposed to calibrate the meters more accurately than the conventional method. The indicated value of the pressure gauge is obtained through the angular position of the pointer and the error value is identified by comparing both standard and test gauge values. In this work, Red, Green, Blue (RGB) images of standard and test pressure gauges are acquired through a high-definition camera. After the image processing operations, the pointer orientation of the indicatoris identified, and the indicated values are calculated. The computer-vision-based automatic calibration system is applied in a 0\u2013100-psi (0\u2013689.476 kPa) analog dial pressure gauge, and three trials were performed to determine the accuracy. The computer vision technique improved accuracy, varying from 97\u00a0% to 98\u00a0% compared with the conventional manual observation method. The observed results have improved repeatability and accuracy with the proposed computer-vision-based system.<\/jats:p>","DOI":"10.1520\/ssms20210019","type":"journal-article","created":{"date-parts":[[2022,7,15]],"date-time":"2022-07-15T05:28:26Z","timestamp":1657862906000},"page":"131-147","update-policy":"https:\/\/doi.org\/10.1115\/crossmarkpolicy-asme","source":"Crossref","is-referenced-by-count":1,"title":["Design and Development of Novel Computer Vision-Based Automatic Calibration System for Analog\u00a0Dial Pressure Gauge"],"prefix":"10.1520","volume":"6","author":[{"ORCID":"https:\/\/orcid.org\/0000-0003-2315-3555","authenticated-orcid":false,"given":"S. J. Suji","family":"Prasad","sequence":"first","affiliation":[{"name":"Department of Electronics and Instrumentation Engineering, Kongu Engineering College (Autonomous), Perundurai 1 , Erode 638060 ,","place":["India"]}],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"J.","family":"Indra","sequence":"additional","affiliation":[{"name":"Department of Electronics and Communication Engineering, KPR Institute of Engineering and Technology (Autonomous), Arasur 2 , Coimbatore 641407 ,","place":["India"]}],"role":[{"role":"author","vocabulary":"crossref"}]},{"ORCID":"https:\/\/orcid.org\/0000-0002-9122-9024","authenticated-orcid":false,"given":"M.","family":"Thangatamilan","sequence":"additional","affiliation":[{"name":"Department of Electronics and Instrumentation Engineering, Kongu Engineering College (Autonomous), Perundurai 3 , Erode 638060 ,","place":["India"]}],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Radhey Shyam","family":"Meena","sequence":"additional","affiliation":[{"name":"Ministry of New and Renewable Energy 4 , New Delhi ,","place":["India"]}],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Suresh","family":"Muthusamy","sequence":"additional","affiliation":[{"name":"Department of Electronics and Communication Engineering (Autonomous), Kongu Engineering College, Perundurai 5 , Erode 638052 ,","place":["India"]}],"role":[{"role":"author","vocabulary":"crossref"}]},{"ORCID":"https:\/\/orcid.org\/0000-0002-3787-9712","authenticated-orcid":false,"given":"Hitesh","family":"Panchal","sequence":"additional","affiliation":[{"name":"Department of Mechanical Engineering, Government Engineering College 6 , Patan, Gujarat 384002 ,","place":["India"]}],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Mahendran","family":"Krishnamoorthy","sequence":"additional","affiliation":[{"name":"Department of Electrical and Electronics Engineering, Jansons Institute of Technology, Karumathampatti 7 , Coimbatore 641659 ,","place":["India"]}],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Kishor Kumar","family":"Sadasivuni","sequence":"additional","affiliation":[{"name":"Centre for Advanced Materials, Qatar University 8 , PO Box 2713, Doha 8943 ,","place":["Qatar"]}],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Manish","family":"Doshi","sequence":"additional","affiliation":[{"name":"Member of Institution of Engineers, Ahmedabad 9 , Gujarat ,","place":["India"]}],"role":[{"role":"author","vocabulary":"crossref"}]}],"member":"381","published-online":{"date-parts":[[2022,7,19]]},"reference":[{"key":"2025062719422037400_B1","doi-asserted-by":"crossref","unstructured":"J.\u00a0Chi, L.\u00a0Liu, J.\u00a0Liu, Z.\u00a0Jiang, and G.\u00a0Zhang, \u201cMachine Vision Based Automatic Detection Method of Indicating Values of a Pointer Gauge,\u201d Mathematical Problems in Engineering2015 (October 2015): 283629, https:\/\/doi.org\/10.1155\/2015\/283629","DOI":"10.1155\/2015\/283629"},{"issue":"3","key":"2025062719422037400_B2","doi-asserted-by":"publisher","first-page":"79","DOI":"10.1177\/0020294014526870a","article-title":"Tech Talk: (1) Instrument Measurement Basics","volume":"47","year":"2014","journal-title":"Measurement and Control"},{"key":"2025062719422037400_B3","doi-asserted-by":"crossref","first-page":"841","DOI":"10.1109\/RAICS.2011.6069428","article-title":"Real Time Visual Inspection System(RTVIS) for Calibration of Mechanical Gauges","volume-title":"2011 IEEE Recent Advances in Intelligent Computational Systems","year":"2011"},{"volume-title":"Instrument Engineers\u2019 Handbook: Process Measurement and Analysis","year":"2003","key":"2025062719422037400_B4"},{"volume-title":"Measurement and Instrumentation Principles","year":"2001","key":"2025062719422037400_B5"},{"issue":"5","key":"2025062719422037400_B6","doi-asserted-by":"publisher","first-page":"459","DOI":"10.21307\/ijssis-2017-263","article-title":"Patrol Inspection and Time Study to Improve Quality","volume":"10","year":"2017","journal-title":"International Journal on Smart Sensing and Intelligent Systems"},{"key":"2025062719422037400_B7","doi-asserted-by":"publisher","first-page":"012008","DOI":"10.1088\/1757-899x\/200\/1\/012008","article-title":"LabVIEW Application for Motion Tracking Using USB Camera","volume":"200","year":"2017","journal-title":"IOP Conference Series: Materials Science and Engineering"},{"key":"2025062719422037400_B8","first-page":"1","article-title":"Computer Vision-Based Pressure Gauge Measurement for Fire Extinguisher Inspection","volume-title":"2017 International Conference on Platform Technology and Service (PlatCon)","year":"2017"},{"key":"2025062719422037400_B9","first-page":"416","article-title":"Automated Aircraft Instrument Reading Using Real Time Video Analysis","volume-title":"2016 IEEE Eighth International Conference on Intelligent Systems (IS)","year":"2016"},{"key":"2025062719422037400_B10","doi-asserted-by":"publisher","first-page":"1","DOI":"10.1016\/j.compind.2014.10.006","article-title":"Automated Visual Inspection in the Semiconductor Industry: A Survey","volume":"66","year":"2015","journal-title":"Computers in Industry"},{"issue":"4","key":"2025062719422037400_B11","doi-asserted-by":"publisher","first-page":"1236","DOI":"10.1109\/jsen.2012.2231940","article-title":"Architecture of Noninvasive Real Time Visual Monitoring System for Dial Type Measuring Instrument","volume":"13","year":"2013","journal-title":"IEEE Sensors Journal"},{"issue":"9\u201312","key":"2025062719422037400_B12","doi-asserted-by":"publisher","first-page":"2099","DOI":"10.1007\/s00170-013-5167-7","article-title":"An Image Acquisition System for Real-Time Seam Tracking","volume":"69","year":"2013","journal-title":"The International Journal of Advanced Manufacturing Technology"},{"issue":"5","key":"2025062719422037400_B13","first-page":"1309","article-title":"Automatic Value Identification of Pointer-Type Pressure Gauge Based on Machine Vision","volume":"8","year":"2013","journal-title":"Journal of Computers"},{"issue":"3","key":"2025062719422037400_B14","doi-asserted-by":"publisher","first-page":"185","DOI":"10.1016\/s0263-2241(00)00011-7","article-title":"Computer Vision Applied to the Automatic Calibration of Measuring Instruments","volume":"28","year":"2000","journal-title":"Measurement"},{"key":"2025062719422037400_B15","first-page":"32","article-title":"Automatic Pointer Meter Reading Based on Machine Vision","volume-title":"2019 IEEE Fourth International Conference on Image, Vision and Computing (ICIVC)","year":"2019"},{"key":"2025062719422037400_B16","doi-asserted-by":"crossref","first-page":"373","DOI":"10.5220\/0007386000002108","article-title":"Reading Circular Analogue Gauges Using Digital Image Processing","volume-title":"Proceedings of the 14th International Joint Conference on Computer Vision, Imaging and Computer Graphics Theory and Applications","year":"2019"},{"issue":"20","key":"2025062719422037400_B17","doi-asserted-by":"publisher","first-page":"5946","DOI":"10.3390\/s20205946","article-title":"A High-Robust Automatic Reading Algorithm of Pointer Meters Based on Text Detection","volume":"20","year":"2020","journal-title":"Sensors"},{"issue":"1","key":"2025062719422037400_B18","doi-asserted-by":"publisher","first-page":"119","DOI":"10.24846\/v30i1y202111","article-title":"Moving Human Target Detection and Tracking in Video Frames","volume":"30","year":"2021","journal-title":"Studies in Informatics and Control"},{"issue":"3","key":"2025062719422037400_B19","doi-asserted-by":"publisher","first-page":"272","DOI":"10.3390\/e23030272","article-title":"Computer Vision Based Automatic Recognition of Pointer Instruments: Data Set Optimization and Reading","volume":"23","year":"2021","journal-title":"Entropy"},{"issue":"2","key":"2025062719422037400_B20","doi-asserted-by":"publisher","first-page":"13","DOI":"10.1080\/2151237x.2007.10129236","article-title":"Adaptive Thresholding Using the Integral Image","volume":"12","year":"2007","journal-title":"Journal of Graphics Tools"},{"key":"2025062719422037400_B21","doi-asserted-by":"publisher","first-page":"62","DOI":"10.1016\/j.robot.2018.10.004","article-title":"Traffic Light Recognition Using High-Definition Map Features","volume":"111","year":"2019","journal-title":"Robotics and Autonomous Systems"},{"issue":"2","key":"2025062719422037400_B22","doi-asserted-by":"publisher","first-page":"133","DOI":"10.1049\/iet-spr:20070117","article-title":"Error Reduction Technique for Four-Quadrant Arctangent Approximations","volume":"2","year":"2008","journal-title":"IET Signal Processing"},{"key":"2025062719422037400_B23","doi-asserted-by":"publisher","first-page":"164226","DOI":"10.1016\/j.ijleo.2020.164226","article-title":"A Calibration and Correction Method for the Measurement System Based on Four-Quadrant Detector","volume":"204","year":"2020","journal-title":"Optik"}],"container-title":["Smart and Sustainable Manufacturing Systems"],"original-title":[],"language":"en","link":[{"URL":"https:\/\/dl.astm.org\/ssms\/article-pdf\/6\/1\/131\/157511\/ssms20210019.pdf","content-type":"application\/pdf","content-version":"vor","intended-application":"syndication"},{"URL":"https:\/\/dl.astm.org\/ssms\/article-pdf\/6\/1\/131\/157511\/ssms20210019.pdf","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2025,6,27]],"date-time":"2025-06-27T23:42:32Z","timestamp":1751067752000},"score":1,"resource":{"primary":{"URL":"https:\/\/dl.astm.org\/ssms\/article\/6\/1\/131\/17381\/Design-and-Development-of-Novel-Computer-Vision"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[2022,7,19]]},"references-count":23,"journal-issue":{"issue":"1","published-print":{"date-parts":[[2022,7,19]]}},"URL":"https:\/\/doi.org\/10.1520\/ssms20210019","relation":{},"ISSN":["2520-6478","2572-3928"],"issn-type":[{"type":"print","value":"2520-6478"},{"type":"electronic","value":"2572-3928"}],"subject":[],"published":{"date-parts":[[2022,7,19]]}}}