{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2026,9,26]],"date-time":"2026-09-26T16:55:37Z","timestamp":1790441737693,"version":"4.1.0"},"reference-count":23,"publisher":"American Academy of Implant Dentistry","issue":"5","content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":[],"published-print":{"date-parts":[[2024,10]]},"abstract":"<jats:p>During implant maintenance, preserving a smooth surface on the machined transmucosal abutment is critical to reduce biofilm attachment and colonization. The present study compared the surface roughness and bacterial colonization of machined titanium surfaces after instrumentation with various materials. Forty-four machined grade 23 titanium discs were instrumented with a round polyether ether ketone (PEEK) tip, a plastic curette tip, or a pure titanium curette tip with piezoelectric devices. Before and after instrumentation, the surface roughness (Ra and Rz) values were analyzed with a profilometer and scanning electron microscopy (SEM). <jats:italic>Streptococcus sanguinis<\/jats:italic> was cultured and incubated for 24 hours on the instrumented discs, and colony-forming units per milliliter were obtained for each group. Samples instrumented with the metal ultrasonic tip significantly increased surface roughness compared with the other groups. This resulted in greater colonization by <jats:italic>S. sanguinis<\/jats:italic> than surfaces instrumented with PEEK tips or the negative control. Samples instrumented with PEEK and plastic tips did not exhibit any statistically significant increase in surface roughness, and SEM analysis revealed a significantly rougher surface of discs instrumented with metal compared with discs instrumented with plastic or PEEK tips despite the possibility of debris from tip dissolution. Our results suggest that instrumentation with metal ultrasonic tips with piezoelectric devices significantly increased machined titanium\u2019s surface roughness and elicited higher biofilm formation in vitro. Meanwhile, instrumentation of machined titanium with PEEK or plastic ultrasonic tips did not affect the surface roughness or bacterial adhesion.<\/jats:p>","DOI":"10.1563\/aaid-joi-d-24-00049","type":"journal-article","created":{"date-parts":[[2024,7,18]],"date-time":"2024-07-18T11:18:20Z","timestamp":1721301500000},"page":"537-543","source":"Crossref","is-referenced-by-count":4,"title":["In Vitro Comparison of Titanium Disc Surface Roughness and Bacterial Colonization After Ultrasonic Instrumentation With Three Different Tips"],"prefix":"10.1563","volume":"50","author":[{"given":"Aleksandr","family":"Kitaygorodskiy","sequence":"first","affiliation":[{"name":"Private practice, Philadelphia, Pennsylvania"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Richard L.","family":"Gregory","sequence":"additional","affiliation":[{"name":"Department of Biomedical Sciences and Comprehensive Care, Indiana University School of Dentistry, Indianapolis, Indiana"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Glendale","family":"Lim","sequence":"additional","affiliation":[{"name":"Private practice, Indianapolis, Indiana"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Yusuke","family":"Hamada","sequence":"additional","affiliation":[{"name":"Section of Periodontics, UCLA School of Dentistry, Los Angeles, California"}],"role":[{"vocabulary":"crossref","role":"author"}]}],"member":"434","reference":[{"key":"i1548-1336-50-5-537-b01","series-title":"J Clin Periodontol","first-page":"S158","article-title":"Peri-implant health and disease. A systematic review of current epidemiology","volume":"42","author":"Derks","year":"2015","unstructured":"Derks\n              J,\n            \n            \n              Tomasi\n              C.\n            \n          \n          Peri-implant health and disease. A systematic review of current epidemiology. J Clin Periodontol. 2015;42:S158\u2013S171."},{"key":"i1548-1336-50-5-537-b02","series-title":"J Dent","first-page":"1","article-title":"Prevalences of peri-implantitis and peri-implant mucositis: systematic review and meta-analysis","volume":"62","author":"Lee","year":"2017","unstructured":"Lee\n              CT,\n            \n            \n              Huang\n              YW,\n            \n            \n              Zhu\n              L,\n            \n            \n              Weltman\n              R.\n            \n          \n          Prevalences of peri-implantitis and peri-implant mucositis: systematic review and meta-analysis. J Dent. 2017;62:1\u201312."},{"key":"i1548-1336-50-5-537-b03","series-title":"J Periodontal Res","first-page":"657","article-title":"Epidemiology and risk factors of peri-implantitis: a systematic review","volume":"53","author":"Dreyer","year":"2018","unstructured":"Dreyer\n              H,\n            \n            \n              Grischke\n              J,\n            \n            \n              Tiede\n              C,\n            \n            et al.\n          \n          Epidemiology and risk factors of peri-implantitis: a systematic review. J Periodontal Res. 2018;53:657\u2013681."},{"key":"i1548-1336-50-5-537-b04","series-title":"Clin Oral Implants Res","first-page":"1","article-title":"Supportive peri-implant therapy following anti-infective surgical peri-implantitis treatment: 5-year survival and success","volume":"29","author":"Heitz-Mayfield","year":"2018","unstructured":"Heitz-Mayfield\n              LJA,\n            \n            \n              Salvi\n              GE,\n            \n            \n              Mombelli\n              A,\n            \n            et al.\n          \n          Supportive peri-implant therapy following anti-infective surgical peri-implantitis treatment: 5-year survival and success. Clin Oral Implants Res. 2018;29:1\u20136. 27335316"},{"key":"i1548-1336-50-5-537-b05","series-title":"Heliyon","first-page":"e10263","article-title":"Comparison of surface aspects of turned and anodized titanium dental implant, or abutment material for an optimal soft tissue integration","volume":"8","author":"Muhl","year":"2022","unstructured":"Muhl\n              A,\n            \n            \n              Szabo\n              P,\n            \n            \n              Krafcsik\n              O,\n            \n            et al.\n          \n          Comparison of surface aspects of turned and anodized titanium dental implant, or abutment material for an optimal soft tissue integration. Heliyon. 2022;8:e10263."},{"key":"i1548-1336-50-5-537-b06","series-title":"J Periodontol","first-page":"S267","article-title":"Peri-implantitis","volume":"89","author":"Schwarz","year":"2018","unstructured":"Schwarz\n              F,\n            \n            \n              Derks\n              J,\n            \n            \n              Monje\n              A,\n            \n            \n              Wang\n              HL.\n            \n          \n          Peri-implantitis. J Periodontol. 2018;89:S267\u2013S290. 29926957"},{"key":"i1548-1336-50-5-537-b07","series-title":"J Periodontol","first-page":"1030","article-title":"Association of Preventive Maintenance Therapy compliance and peri-implant diseases: a cross-sectional study","volume":"88","author":"Monje","year":"2017","unstructured":"Monje\n              A,\n            \n            \n              Wang\n              HL,\n            \n            \n              Nart\n              J.\n            \n          \n          Association of Preventive Maintenance Therapy compliance and peri-implant diseases: a cross-sectional study. J Periodontol. 2017;88:1030\u20131041."},{"key":"i1548-1336-50-5-537-b08","series-title":"J Clin Periodontol","first-page":"S152","article-title":"Primary prevention of peri-implantitis: managing peri-implant mucositis","volume":"42","author":"Jepsen","year":"2015","unstructured":"Jepsen\n              S,\n            \n            \n              Berglundh\n              T,\n            \n            \n              Genco\n              R,\n            \n            et al.\n          \n          Primary prevention of peri-implantitis: managing peri-implant mucositis. J Clin Periodontol. 2015;42:S152\u2013S157."},{"key":"i1548-1336-50-5-537-b09","series-title":"Biology (Basel)","first-page":"10","article-title":"Impact of implant surface material and microscale roughness on the initial attachment and proliferation of primary human gingival fibroblasts","author":"Rausch","year":"2021","unstructured":"Rausch\n              MA,\n            \n            \n              Shokoohi-Tabrizi\n              H,\n            \n            \n              Wehner\n              C,\n            \n            et al.\n          \n          Impact of implant surface material and microscale roughness on the initial attachment and proliferation of primary human gingival fibroblasts. Biology (Basel). 2021;10."},{"key":"i1548-1336-50-5-537-b10","series-title":"Clin Oral Implants Res","first-page":"201","article-title":"The influence of abutment surface roughness on plaque accumulation and peri-implant mucositis","volume":"7","author":"Bollen","year":"1996","unstructured":"Bollen\n              CM,\n            \n            \n              Papaioanno\n              W,\n            \n            \n              Van Eldere\n              J,\n            \n            et al.\n          \n          The influence of abutment surface roughness on plaque accumulation and peri-implant mucositis. Clin Oral Implants Res. 1996;7:201\u2013211."},{"key":"i1548-1336-50-5-537-b11","series-title":"Int J Oral Maxillofac Implants","first-page":"169","article-title":"The influence of titanium abutment surface roughness on plaque accumulation and gingivitis: short-term observations","volume":"11","author":"Quirynen","year":"1996","unstructured":"Quirynen\n              M,\n            \n            \n              Bollen\n              CM,\n            \n            \n              Papaioannou\n              W,\n            \n            \n              Van Eldere\n              J,\n            \n            \n              van Steenberghe\n              D.\n            \n          \n          The influence of titanium abutment surface roughness on plaque accumulation and gingivitis: short-term observations. Int J Oral Maxillofac Implants. 1996;11:169\u2013178."},{"key":"i1548-1336-50-5-537-b12","series-title":"Clin Oral Implants Res","first-page":"68","article-title":"Effect of material characteristics and\/or surface topography on biofilm development","volume":"17","author":"Teughels","year":"2006","unstructured":"Teughels\n              W,\n            \n            \n              Van Assche\n              N,\n            \n            \n              Sliepen\n              I,\n            \n            \n              Quirynen\n              M.\n            \n          \n          Effect of material characteristics and\/or surface topography on biofilm development. Clin Oral Implants Res. 2006;17:68\u201381."},{"key":"i1548-1336-50-5-537-b13","series-title":"J Periodontol","first-page":"485","article-title":"The effects of scaling a titanium implant surface with metal and plastic instruments: an in vitro study","volume":"61","author":"Fox","year":"1990","unstructured":"Fox\n              SC,\n            \n            \n              Moriarty\n              JD,\n            \n            \n              Kusy\n              RP.\n            \n          \n          The effects of scaling a titanium implant surface with metal and plastic instruments: an in vitro study. J Periodontol. 1990;61:485\u2013490."},{"key":"i1548-1336-50-5-537-b14","series-title":"Clin Oral Implants Res","first-page":"1149","article-title":"The effects of mechanical instruments on contaminated titanium dental implant surfaces: a systematic review","volume":"25","author":"Louropoulou","year":"2014","unstructured":"Louropoulou\n              A,\n            \n            \n              Slot\n              DE,\n            \n            \n              Van der Weijden\n              F.\n            \n          \n          The effects of mechanical instruments on contaminated titanium dental implant surfaces: a systematic review. Clin Oral Implants Res. 2014;25:1149\u20131160."},{"key":"i1548-1336-50-5-537-b15","series-title":"Int Dent Res","first-page":"84","article-title":"Areas for use of PEEK material in dentistry","volume":"8","author":"Tekin","year":"2018","unstructured":"Tekin\n              SCS,\n            \n            \n              Ad\u0131g\u00fczel\n              \u00d6,\n            \n            \n              De\u011fer\n              Y.\n            \n          \n          Areas for use of PEEK material in dentistry. Int Dent Res. 2018;8:84\u201392."},{"key":"i1548-1336-50-5-537-b16","series-title":"J Clin Periodontol","first-page":"88","article-title":"Some soft tissue characteristics at implant abutments with different surface topography. A study in humans","volume":"30","author":"Wennerberg","year":"2003","unstructured":"Wennerberg\n              A,\n            \n            \n              Sennerby\n              L,\n            \n            \n              Kultje\n              C,\n            \n            \n              Lekholm\n              U.\n            \n          \n          Some soft tissue characteristics at implant abutments with different surface topography. A study in humans. J Clin Periodontol. 2003;30:88\u201394."},{"key":"i1548-1336-50-5-537-b17","series-title":"J Dent Res","first-page":"21","article-title":"Microbial community-driven etiopathogenesis of peri-implantitis","volume":"100","author":"Belibasakis","year":"2021","unstructured":"Belibasakis\n              GN,\n            \n            \n              Manoil\n              D.\n            \n          \n          Microbial community-driven etiopathogenesis of peri-implantitis. J Dent Res. 2021;100:21\u201328."},{"key":"i1548-1336-50-5-537-b18","series-title":"BMC Oral Health","first-page":"25","article-title":"Assessment of implant surface and instrument insert changes due to instrumentation with different tips for ultrasonic-driven debridement","volume":"21","author":"Sahrmann","year":"2021","unstructured":"Sahrmann\n              P,\n            \n            \n              Winkler\n              S,\n            \n            \n              Gubler\n              A,\n            \n            \n              Attin\n              T.\n            \n          \n          Assessment of implant surface and instrument insert changes due to instrumentation with different tips for ultrasonic-driven debridement. BMC Oral Health. 2021;21:25."},{"key":"i1548-1336-50-5-537-b19","series-title":"J Periodontol","first-page":"556","article-title":"The effect of surface roughness on early in vivo plaque colonization on titanium","volume":"68","author":"Rimondini","year":"1997","unstructured":"Rimondini\n              L,\n            \n            \n              Fare\n              S,\n            \n            \n              Brambilla\n              E,\n            \n            et al.\n          \n          The effect of surface roughness on early in vivo plaque colonization on titanium. J Periodontol. 1997;68:556\u2013562."},{"key":"i1548-1336-50-5-537-b20","series-title":"J Dent Res","first-page":"1304","article-title":"An in vivo study of the influence of the surface roughness of implants on the microbiology of supra- and subgingival plaque","volume":"72","author":"Quirynen","year":"1993","unstructured":"Quirynen\n              M,\n            \n            \n              van der Mei\n              HC,\n            \n            \n              Bollen\n              CM,\n            \n            et al.\n          \n          An in vivo study of the influence of the surface roughness of implants on the microbiology of supra- and subgingival plaque. J Dent Res. 1993;72:1304\u20131309."},{"key":"i1548-1336-50-5-537-b21","series-title":"Int J Appl Dent","first-page":"643","article-title":"Comparative evaluation of surface roughness of three different dental implant abutments using atomic force microscopy: an in vitro study","volume":"6","author":"Femiljilta","year":"2020","unstructured":"Femiljilta\n              C,\n            \n            \n              Ravichandran\n              R,\n            \n            \n              Karunakaran\n              HK,\n            \n            \n              Nair\n              VV,\n            \n            \n              Janardanan\n              K,\n            \n            \n              Zeenath\n              H.\n            \n          \n          Comparative evaluation of surface roughness of three different dental implant abutments using atomic force microscopy: an in vitro study. Int J Appl Dent. 2020;6:643\u2013647."},{"key":"i1548-1336-50-5-537-b22","series-title":"Int J Oral Maxillofac Implants","first-page":"308","article-title":"Effects of 10 cleaning instruments on four different implant surfaces","volume":"27","author":"Schmage","year":"2012","unstructured":"Schmage\n              P,\n            \n            \n              Thielemann\n              J,\n            \n            \n              Nergiz\n              I,\n            \n            \n              Scorziello\n              TM,\n            \n            \n              Pfeiffer\n              P.\n            \n          \n          Effects of 10 cleaning instruments on four different implant surfaces. Int J Oral Maxillofac Implants. 2012;27:308\u2013317."},{"key":"i1548-1336-50-5-537-b23","series-title":"Clin Oral Implants Res","first-page":"483","article-title":"Influence of different instrumentation modalities on the surface characteristics and biofilm formation on dental implant neck, in vitro","volume":"28","author":"Schmidt","year":"2017","unstructured":"Schmidt\n              KE,\n            \n            \n              Auschill\n              TM,\n            \n            \n              Heumann\n              C,\n            \n            et al.\n          \n          Influence of different instrumentation modalities on the surface characteristics and biofilm formation on dental implant neck, in vitro. Clin Oral Implants Res. 2017;28:483\u2013490."}],"container-title":["Journal of Oral Implantology"],"original-title":[],"link":[{"URL":"https:\/\/joi.kglmeridian.com\/view\/journals\/orim\/50\/5\/article-p537.xml","content-type":"text\/html","content-version":"vor","intended-application":"text-mining"},{"URL":"https:\/\/meridian.allenpress.com\/joi\/article-pdf\/50\/5\/537\/3440146\/i1548-1336-50-5-537.pdf","content-type":"application\/pdf","content-version":"vor","intended-application":"syndication"},{"URL":"https:\/\/joi.kglmeridian.com\/downloadpdf\/journals\/orim\/50\/5\/article-p537.xml","content-type":"text\/html","content-version":"vor","intended-application":"syndication"},{"URL":"https:\/\/joi.kglmeridian.com\/downloadpdf\/journals\/orim\/50\/5\/article-p537.xml","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2025,7,11]],"date-time":"2025-07-11T00:22:30Z","timestamp":1752193350000},"score":1,"resource":{"primary":{"URL":"https:\/\/joi.kglmeridian.com\/view\/journals\/orim\/50\/5\/article-p537.xml"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[2024,10]]},"references-count":23,"journal-issue":{"issue":"5"},"URL":"https:\/\/doi.org\/10.1563\/aaid-joi-d-24-00049","relation":{},"ISSN":["1548-1336","0160-6972"],"issn-type":[{"value":"1548-1336","type":"electronic"},{"value":"0160-6972","type":"print"}],"subject":[],"published":{"date-parts":[[2024,10]]}}}