{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2026,10,9]],"date-time":"2026-10-09T08:08:28Z","timestamp":1791533308560,"version":"4.3.4"},"reference-count":33,"publisher":"JMIR Publications Inc.","content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":["JMIR Cardio"],"abstract":"<jats:title>Abstract<\/jats:title>\n                  <jats:sec sec-type=\"background\">\n                    <jats:title>Background<\/jats:title>\n                    <jats:p>Wearable devices offer a promising solution for remotely monitoring heart rate (HR) during home-based cardiac rehabilitation. However, evidence regarding their accuracy across varying exercise intensities and patient profiles remains limited, particularly in populations with cardiovascular disease (CVD) such as those with heart failure (HF).<\/jats:p>\n                  <\/jats:sec>\n                  <jats:sec sec-type=\"objective\">\n                    <jats:title>Objective<\/jats:title>\n                    <jats:p>The objective of this study was to evaluate the accuracy of HR measurements obtained using the Fitbit Inspire 3 during cardiopulmonary exercise testing (CPX) in patients with CVD, including those with HF.<\/jats:p>\n                  <\/jats:sec>\n                  <jats:sec sec-type=\"methods\">\n                    <jats:title>Methods<\/jats:title>\n                    <jats:p>In this single-center, prospective pilot study, we enrolled 30 patients with CVD undergoing CPX. HR was simultaneously recorded using electrocardiography and the Fitbit Inspire 3 at 1-minute intervals across various CPX phases: rest, exercise below and above the anaerobic threshold (AT), and recovery. The correlation between the two methods was assessed using the Pearson correlation coefficient. Measurement error was quantified by mean absolute error and mean absolute percentage error (MAPE), with a MAPE of \u226410% defined as the threshold for acceptable agreement.<\/jats:p>\n                  <\/jats:sec>\n                  <jats:sec sec-type=\"results\">\n                    <jats:title>Results<\/jats:title>\n                    <jats:p>\n                      All data points were 630 points per minute. The Fitbit Inspire 3 device demonstrated a strong overall correlation with electrocardiography-derived HR (\n                      <jats:italic>r<\/jats:italic>\n                      =0.90; IQR 0.88\u20100.91) and an acceptable MAPE of 5.40% (SD 8.33%). The total error was 14.9% (94\/630), with overestimation and underestimation of 37 (5.8%) points and 57 (9%) points, respectively. The rate of HR underestimation reached 19 (16%) points during exercise above the AT, compared to 1 (3%) point at rest. When stratified by HF stage (B vs C), underestimation was more pronounced in patients with HF (14\/275, 5% points vs 40\/355, 11.2% points).\n                    <\/jats:p>\n                  <\/jats:sec>\n                  <jats:sec sec-type=\"conclusions\">\n                    <jats:title>Conclusions<\/jats:title>\n                    <jats:p>The Fitbit Inspire 3 provides acceptable validity for HR monitoring during CPX in patients with CVD. However, clinicians should interpret HR data with caution during high-intensity exercise, especially in patients with HF.<\/jats:p>\n                  <\/jats:sec>","DOI":"10.2196\/77911","type":"journal-article","created":{"date-parts":[[2025,10,6]],"date-time":"2025-10-06T20:25:56Z","timestamp":1759782356000},"page":"e77911-e77911","source":"Crossref","is-referenced-by-count":2,"title":["Validity of Heart Rate Measurement Using Wearable Devices During Cardiopulmonary Exercise Testing in Patients With Cardiovascular Disease: Prospective Pilot Validation Study"],"prefix":"10.2196","volume":"9","author":[{"ORCID":"https:\/\/orcid.org\/0000-0003-1955-257X","authenticated-orcid":false,"given":"Kazufumi","family":"Kitagaki","sequence":"first","affiliation":[],"role":[{"vocabulary":"crossref","role":"author"}]},{"ORCID":"https:\/\/orcid.org\/0009-0003-8040-1778","authenticated-orcid":false,"given":"Yuji","family":"Hongo","sequence":"additional","affiliation":[],"role":[{"vocabulary":"crossref","role":"author"}]},{"ORCID":"https:\/\/orcid.org\/0009-0003-6481-0977","authenticated-orcid":false,"given":"Rie","family":"Futai","sequence":"additional","affiliation":[],"role":[{"vocabulary":"crossref","role":"author"}]},{"ORCID":"https:\/\/orcid.org\/0009-0001-5092-3503","authenticated-orcid":false,"given":"Takeshi","family":"Hasegawa","sequence":"additional","affiliation":[],"role":[{"vocabulary":"crossref","role":"author"}]},{"ORCID":"https:\/\/orcid.org\/0009-0001-0420-8154","authenticated-orcid":false,"given":"Hiroshi","family":"Morikawa","sequence":"additional","affiliation":[],"role":[{"vocabulary":"crossref","role":"author"}]},{"ORCID":"https:\/\/orcid.org\/0000-0002-6577-6681","authenticated-orcid":false,"given":"Hisashi","family":"Shimoyama","sequence":"additional","affiliation":[],"role":[{"vocabulary":"crossref","role":"author"}]}],"member":"1010","published-online":{"date-parts":[[2025,10,6]]},"reference":[{"issue":"1","key":"R1","doi-asserted-by":"publisher","first-page":"155","DOI":"10.1253\/circj.CJ-22-0234","article-title":"JCS\/JACR 2021 Guideline on rehabilitation in patients with cardiovascular disease","volume":"87","author":"Makita","journal-title":"Circ J"},{"issue":"10","key":"R2","doi-asserted-by":"publisher","first-page":"1860","DOI":"10.1253\/circj.CJ-20-1154","article-title":"Estimating incidence of acute heart failure syndromes in Japan - an analysis from the KUNIUMI Registry","volume":"85","author":"Fujimoto","journal-title":"Circ J"},{"issue":"7","key":"R3","doi-asserted-by":"publisher","first-page":"1546","DOI":"10.1253\/circj.CJ-19-0241","article-title":"Nationwide survey of multidisciplinary care and cardiac rehabilitation for patients with heart failure in Japan - an analysis of the AMED-CHF Study","volume":"83","author":"Kamiya","journal-title":"Circ J"},{"key":"R4","doi-asserted-by":"publisher","DOI":"10.2147\/VHRM.S425505","article-title":"Barriers to participation in cardiac rehabilitation among patients with coronary heart disease after reperfusion therapy: a scoping review","volume":"19","author":"Sugiharto","journal-title":"Vasc Health Risk Manag"},{"issue":"12","key":"R5","doi-asserted-by":"publisher","first-page":"2263","DOI":"10.1002\/ejhf.3046","article-title":"Exercise-based cardiac rehabilitation for adults with heart failure - 2023 Cochrane systematic review and meta-analysis","volume":"25","author":"Molloy","journal-title":"Eur J Heart Fail"},{"issue":"9","key":"R6","doi-asserted-by":"publisher","first-page":"865","DOI":"10.1007\/s40279-013-0068-3","article-title":"Estimation of maximal oxygen uptake via submaximal exercise testing in sports, clinical, and home settings","volume":"43","author":"Sartor","journal-title":"Sports Med"},{"issue":"4","key":"R7","doi-asserted-by":"publisher","first-page":"209","DOI":"10.5535\/arm.22050","article-title":"Accuracy and validity of commercial smart bands for heart rate measurements during cardiopulmonary exercise test","volume":"46","author":"Kim","journal-title":"Ann Rehabil Med"},{"issue":"22","key":"R8","doi-asserted-by":"publisher","first-page":"2100","DOI":"10.1056\/NEJMe2303219","article-title":"Wearable digital health technology","volume":"389","author":"Friend","journal-title":"N Engl J Med"},{"issue":"3","key":"R9","doi-asserted-by":"publisher","first-page":"1727","DOI":"10.1002\/ehf2.15226","article-title":"Smartwatches in the assessment of heart failure patients in epidemiology and pathophysiology studies: a scoping review","volume":"12","author":"Fabien","journal-title":"ESC Heart Fail"},{"issue":"9","key":"R10","doi-asserted-by":"publisher","DOI":"10.2196\/18694","article-title":"Reliability and validity of commercially available wearable devices for measuring steps, energy expenditure, and heart rate: systematic review","volume":"8","author":"Fuller","journal-title":"JMIR Mhealth Uhealth"},{"issue":"3","key":"R11","first-page":"540","volume":"15","author":"Jo","journal-title":"J Sports Sci Med"},{"issue":"4","key":"R12","doi-asserted-by":"publisher","first-page":"261","DOI":"10.5535\/arm.23019","article-title":"Validation of wearable digital devices for heart rate measurement during exercise test in patients with coronary artery disease","volume":"47","author":"Kim","journal-title":"Ann Rehabil Med"},{"issue":"12","key":"R13","doi-asserted-by":"publisher","DOI":"10.2196\/42359","article-title":"Accuracy and systematic biases of heart rate measurements by consumer-grade fitness trackers in postoperative patients: prospective clinical trial","volume":"24","author":"Helmer","journal-title":"J Med Internet Res"},{"issue":"2","key":"R14","doi-asserted-by":"publisher","first-page":"280","DOI":"10.1249\/MSS.0000000000003567","article-title":"Validity of wrist-worn activity tracker heart rate detection in Fontan patients during exercise","volume":"57","author":"Pierick","journal-title":"Med Sci Sports Exerc"},{"issue":"10","key":"R15","doi-asserted-by":"publisher","DOI":"10.2196\/mhealth.8233","article-title":"Fitbit Charge HR wireless heart rate monitor: validation study conducted under free-living conditions","volume":"5","author":"Gorny","journal-title":"JMIR Mhealth Uhealth"},{"issue":"18","key":"R16","doi-asserted-by":"publisher","first-page":"e895","DOI":"10.1161\/CIR.0000000000001063","article-title":"2022 AHA\/ACC\/HFSA Guideline for the Management of Heart Failure: a report of the American College of Cardiology\/American Heart Association Joint Committee on Clinical Practice Guidelines","volume":"145","author":"Heidenreich","journal-title":"Circulation"},{"issue":"3","key":"R17","doi-asserted-by":"publisher","first-page":"352","DOI":"10.1002\/ejhf.2115","article-title":"Universal definition and classification of heart failure: a report of the Heart Failure Society of America, Heart Failure Association of the European Society of Cardiology, Japanese Heart Failure Society and Writing Committee of the Universal Definition of Heart Failure: endorsed by the Canadian Heart Failure Society, Heart Failure Association of India, Cardiac Society of Australia and New Zealand, and Chinese Heart Failure Association, Heart Failure Association","volume":"23","author":"Bozkurt","journal-title":"Eur J Heart Fail"},{"issue":"3","key":"R18","doi-asserted-by":"publisher","first-page":"624","DOI":"10.1249\/MSS.0000000000001471","article-title":"Validity of wearable activity monitors during cycling and resistance exercise","volume":"50","author":"Boudreaux","journal-title":"Med Sci Sports Exerc"},{"issue":"3","key":"R19","doi-asserted-by":"publisher","DOI":"10.2196\/10828","article-title":"Accuracy of consumer wearable heart rate measurement during an ecologically valid 24-hour period: intraindividual validation study","volume":"7","author":"Nelson","journal-title":"JMIR Mhealth Uhealth"},{"issue":"1","key":"R20","doi-asserted-by":"publisher","DOI":"10.1136\/bmjsem-2015-000106","article-title":"How accurate are the wrist-based heart rate monitors during walking and running activities? Are they accurate enough?","volume":"2","author":"Stahl","journal-title":"BMJ Open Sport Exerc Med"},{"issue":"1","key":"R21","doi-asserted-by":"publisher","first-page":"104","DOI":"10.1001\/jamacardio.2016.3340","article-title":"Accuracy of wrist-worn heart rate monitors","volume":"2","author":"Wang","journal-title":"JAMA Cardiol"},{"issue":"3","key":"R22","doi-asserted-by":"publisher","DOI":"10.2196\/33635","article-title":"Heart rate measurement accuracy of Fitbit Charge 4 and Samsung Galaxy Watch Active 2: device evaluation study","volume":"6","author":"Nissen","journal-title":"JMIR Form Res"},{"issue":"1","key":"R23","doi-asserted-by":"publisher","DOI":"10.2196\/30863","article-title":"Continuous monitoring of vital signs with wearable sensors during daily life activities: validation study","volume":"6","author":"Haveman","journal-title":"JMIR Form Res"},{"issue":"3","key":"R24","doi-asserted-by":"publisher","DOI":"10.2196\/25313","article-title":"Measurement of heart rate using the Polar OH1 and Fitbit Charge 3 wearable devices in healthy adults during light, moderate, vigorous, and sprint-based exercise: validation study","volume":"9","author":"Muggeridge","journal-title":"JMIR Mhealth Uhealth"},{"issue":"110938","key":"R25","doi-asserted-by":"publisher","first-page":"110938","DOI":"10.1016\/j.exger.2020.110938","article-title":"Pulse transit time as a proxy for vasoconstriction in younger and older adults","volume":"135","author":"Mol","journal-title":"Exp Gerontol"},{"key":"R26","doi-asserted-by":"crossref","unstructured":"Johnson PC . Overview of the microcirculation. In: Tuma RF Duran WN Ley K , editors. Microcirculation. Academic Press; 2008:xi-xxiv. [doi: 10.1016\/b978-0-12-374530-9.00022-x] ISBN: 9780123745309","DOI":"10.1016\/B978-0-12-374530-9.00022-X"},{"issue":"3","key":"R27","doi-asserted-by":"publisher","first-page":"382","DOI":"10.1097\/00005768-199803000-00007","article-title":"Physical training and the control of skin blood flow","volume":"30","author":"Johnson","journal-title":"Med Sci Sports Exerc"},{"issue":"3","key":"R28","doi-asserted-by":"publisher","first-page":"R1","DOI":"10.1088\/0967-3334\/28\/3\/R01","article-title":"Photoplethysmography and its application in clinical physiological measurement","volume":"28","author":"Allen","journal-title":"Physiol Meas"},{"issue":"3","key":"R29","doi-asserted-by":"publisher","first-page":"125","DOI":"10.1080\/03091902.2020.1753836","article-title":"Measurement latency significantly contributes to reduced heart rate measurement accuracy in wearable devices","volume":"44","author":"St\u00f8ve","journal-title":"J Med Eng Technol"},{"issue":"1","key":"R30","doi-asserted-by":"publisher","DOI":"10.1038\/s41746-020-0226-6","article-title":"Investigating sources of inaccuracy in wearable optical heart rate sensors","volume":"3","author":"Bent","journal-title":"NPJ Digit Med"},{"issue":"3","key":"R31","doi-asserted-by":"publisher","first-page":"245","DOI":"10.1016\/j.jacc.2023.04.054","article-title":"Consumer wearable health and fitness technology in cardiovascular medicine: JACC state-of-the-art review","volume":"82","author":"Petek","journal-title":"J Am Coll Cardiol"},{"issue":"6","key":"R32","first-page":"948","volume":"38","author":"Nappi","journal-title":"J Nucl Med"},{"key":"R33","doi-asserted-by":"publisher","DOI":"10.1016\/j.ijcard.2019.11.120","article-title":"Accuracy of wrist-worn heart rate monitors for rate control assessment in atrial fibrillation","volume":"300","author":"Al-Kaisey","journal-title":"Int J Cardiol"}],"container-title":["JMIR Cardio"],"original-title":[],"language":"en","link":[{"URL":"https:\/\/cardio.jmir.org\/2025\/1\/e77911","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2026,6,30]],"date-time":"2026-06-30T18:10:16Z","timestamp":1782843016000},"score":1,"resource":{"primary":{"URL":"https:\/\/cardio.jmir.org\/2025\/1\/e77911"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[2025,10,6]]},"references-count":33,"URL":"https:\/\/doi.org\/10.2196\/77911","relation":{},"ISSN":["2561-1011"],"issn-type":[{"value":"2561-1011","type":"electronic"}],"subject":[],"published":{"date-parts":[[2025,10,6]]},"article-number":"v9i1e77911"}}