{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2026,7,31]],"date-time":"2026-07-31T16:04:11Z","timestamp":1785513851943,"version":"3.56.0"},"reference-count":21,"publisher":"American Veterinary Medical Association (AVMA)","issue":"8","license":[{"start":{"date-parts":[[2026,8,1]],"date-time":"2026-08-01T00:00:00Z","timestamp":1785542400000},"content-version":"unspecified","delay-in-days":0,"URL":"http:\/\/creativecommons.org\/licenses\/by-nc\/4.0\/"}],"content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":["ajvr"],"published-print":{"date-parts":[[2026,8,1]]},"abstract":"<jats:title>Abstract<\/jats:title>\n                  <jats:sec id=\"sc_a1\">\n                    <jats:title>Objective<\/jats:title>\n                    <jats:p>To compare the accuracy of 2 Doppler and 2 oscillometric devices to invasive blood pressure (IBP) measurements in conscious dogs at 2 cuff locations using the American College of Veterinary Internal Medicine (ACVIM) validation criteria.<\/jats:p>\n                  <\/jats:sec>\n                  <jats:sec id=\"sc_a2\">\n                    <jats:title>Methods<\/jats:title>\n                    <jats:p>8 Beagles had dorsal pedal arterial catheters placed under anesthesia on November 13 and 18, 2024. Once conscious, paired non-IBP and IBP measurements were collected at the tail and forelimb for all devices. Bias, SD, percentage of values within 10 and 20 mm Hg of IBP, correlation coefficients, and limits of agreement were calculated for each paired measurement.<\/jats:p>\n                  <\/jats:sec>\n                  <jats:sec id=\"sc_a3\">\n                    <jats:title>Results<\/jats:title>\n                    <jats:p>No device met ACVIM validation criteria at either cuff location. Devices performed better at the tail compared to the forelimb. Doppler devices performed better than oscillometric devices for systolic blood pressure. Oscillometric devices performed best for mean blood pressure (MBP). The MBP measured by the multiparameter monitor at the tail showed the most favorable overall performance, including a bias of 1.75 mm Hg (SD, 8.98); 82% and 96% of measurements within 10 and 20 mm Hg of IBP, respectively; and limits of agreement of \u221215 to 19. The correlation coefficient threshold was not met by any device.<\/jats:p>\n                  <\/jats:sec>\n                  <jats:sec id=\"sc_a4\">\n                    <jats:title>Conclusions<\/jats:title>\n                    <jats:p>None of the devices were validated against IBP in conscious adult dogs using ACVIM criteria. Tail placement improved performance across devices. Oscillometric devices appeared more reliable for MBP, whereas Doppler devices provided more reliable systolic blood pressure measurements.<\/jats:p>\n                  <\/jats:sec>\n                  <jats:sec id=\"sc_a5\">\n                    <jats:title>Clinical Relevance<\/jats:title>\n                    <jats:p>Clinicians should interpret NIBP cautiously in conscious dogs.<\/jats:p>\n                  <\/jats:sec>","DOI":"10.2460\/ajvr.26.03.0108","type":"journal-article","created":{"date-parts":[[2026,5,27]],"date-time":"2026-05-27T12:38:41Z","timestamp":1779885521000},"source":"Crossref","is-referenced-by-count":0,"title":["Two Doppler blood pressure devices and 2 oscillometric blood pressure devices failed American College of Veterinary Internal Medicine validation in conscious dogs"],"prefix":"10.2460","volume":"87","author":[{"given":"Kaylee A.","family":"Quinn","sequence":"first","affiliation":[{"name":"Department of Clinical Sciences, College of Veterinary Medicine and Biomedical Sciences, Colorado State University, Fort Collins, CO"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Casey","family":"Lorge","sequence":"additional","affiliation":[{"name":"Department of Clinical Sciences, College of Veterinary Medicine and Biomedical Sciences, Colorado State University, Fort Collins, CO"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"ORCID":"https:\/\/orcid.org\/0000-0003-0327-8537","authenticated-orcid":true,"given":"Kristin M.","family":"Zersen","sequence":"additional","affiliation":[{"name":"Department of Clinical Sciences, College of Veterinary Medicine and Biomedical Sciences, Colorado State University, Fort Collins, CO"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"ORCID":"https:\/\/orcid.org\/0000-0001-8622-4387","authenticated-orcid":true,"given":"Julien","family":"Guillaumin","sequence":"additional","affiliation":[{"name":"Department of Clinical Sciences, College of Veterinary Medicine and Biomedical Sciences, Colorado State University, Fort Collins, CO"}],"role":[{"vocabulary":"crossref","role":"author"}]}],"member":"1272","reference":[{"key":"B1","series-title":"Advanced Monitoring and Procedures for Small Animal Emergency and Critical Care","first-page":"134","article-title":"Noninvasive arterial blood pressure monitoring","author":"Williamson JA","year":"2012","unstructured":"Williamson JA, Leone S. Noninvasive arterial blood pressure monitoring. In: Burkitt-Creedon JM, Davis H, eds. Advanced Monitoring and Procedures for Small Animal Emergency and Critical Care. John Wiley & Sons Ltd; 2012:134\u2013144."},{"key":"B2","series-title":"Veterinary Anesthesia and Analgesia","first-page":"86","article-title":"Monitoring anesthetized patients","author":"Haskins SC","year":"2015","unstructured":"Haskins SC. Monitoring anesthetized patients. In: Grimm KA, Lamont LA, Tranquilli WJ, Greene SA, Robertson SA, eds. Veterinary Anesthesia and Analgesia. 5th ed. John Wiley & Sons Ltd; 2015:86\u2013113."},{"issue":"4","key":"B3","series-title":"Can Vet J","first-page":"368","article-title":"Non-invasive blood pressure measurement in animals: part 1 - techniques for measurement and validation of non-invasive devices","author":"Skelding A","year":"2020","unstructured":"Skelding A, Valverde A. Non-invasive blood pressure measurement in animals: part 1 - techniques for measurement and validation of non-invasive devices. Can Vet J. 2020;61(4):368\u2013374."},{"issue":"7","key":"B4","series-title":"J Clin Hypertens","first-page":"1096","article-title":"Validation protocols for blood pressure measuring devices in the 21st century","author":"Stergiou GS","year":"2018","unstructured":"Stergiou GS, Alpert BS, Mieke S, Wang J, O\u2019Brien E. Validation protocols for blood pressure measuring devices in the 21st century. J Clin Hypertens. 2018;20(7):1096\u20131099. doi:10.1111\/jch.13294"},{"issue":"3","key":"B5","series-title":"J Vet Intern Med","first-page":"542","article-title":"Guidelines for the identification, evaluation, and management of systemic hypertension in dogs and cats","author":"Brown S","year":"2007","unstructured":"Brown S, Atkins C, Bagley R, et al. Guidelines for the identification, evaluation, and management of systemic hypertension in dogs and cats. J Vet Intern Med. 2007;21(3):542\u2013558. doi:10.1892\/0891-6640(2007)21%5B542:gftiea%5D2.0.co;2"},{"issue":"1","key":"B6","series-title":"Res Vet Sci","first-page":"111","article-title":"Evaluation of oscillometric and Doppler ultrasonic devices for blood pressure measurements in anesthetized and conscious dogs","author":"Vachon C","year":"2014","unstructured":"Vachon C, Belanger MC, Burns PM. Evaluation of oscillometric and Doppler ultrasonic devices for blood pressure measurements in anesthetized and conscious dogs. Res Vet Sci. 2014;97(1):111\u2013117. doi:10.1016\/j.rvsc.2014.05.003"},{"issue":"2","key":"B7","series-title":"J Vet Emerg Crit Care","first-page":"207","article-title":"Comparison of ultrasonic Doppler flow monitor, oscillometric, and direct arterial blood pressure measurements in ill dogs","author":"Bosiack AP","year":"2010","unstructured":"Bosiack AP, Mann FA, Dodam JR, Wagner-Mann CC, Branson KR. Comparison of ultrasonic Doppler flow monitor, oscillometric, and direct arterial blood pressure measurements in ill dogs. J Vet Emerg Crit Care. 2010;20(2):207\u2013215. doi:10.1111\/j.1476-4431.2010.00520.x"},{"issue":"3","key":"B8","series-title":"Can J Vet Res","first-page":"211","article-title":"Evaluation of oscillometric and Doppler ultrasonic methods of indirect blood pressure estimation in conscious dogs","author":"Haberman CE","year":"2006","unstructured":"Haberman CE, Kang CW, Morgan JD, Brown SA. Evaluation of oscillometric and Doppler ultrasonic methods of indirect blood pressure estimation in conscious dogs. Can J Vet Res. 2006;70(3):211\u2013217."},{"issue":"6","key":"B9","series-title":"J Vet Intern Med","first-page":"1803","article-title":"ACVIM consensus statement: guidelines for the identification, evaluation, and management of systemic hypertension in dogs and cats","author":"Acierno MJ","year":"2018","unstructured":"Acierno MJ, Brown S, Coleman AE, et al. ACVIM consensus statement: guidelines for the identification, evaluation, and management of systemic hypertension in dogs and cats. J Vet Intern Med. 2018;32(6):1803\u20131822. doi:10.1111\/jvim.15331"},{"issue":"2","key":"B10","series-title":"Vet Anaesth Analg","first-page":"156","article-title":"Validation of oscillometric blood pressure measurement using a Datex S\/5 compact multiparameter monitor in anaesthetized adult dogs","author":"Felisberto R","year":"2022","unstructured":"Felisberto R, Gonzalez AG, Flaherty D, Casoria V, Auckburally A. Validation of oscillometric blood pressure measurement using a Datex S\/5 compact multiparameter monitor in anaesthetized adult dogs. Vet Anaesth Analg. 2022;49(2):156\u2013164. doi:10.1016\/j.vaa.2021.12.001"},{"issue":"4","key":"B11","series-title":"N Z Vet J","first-page":"185","article-title":"Evaluation of an indirect oscillometric blood pressure monitor in anaesthetised dogs at three different anatomical locations","author":"Fujiyama M","year":"2017","unstructured":"Fujiyama M, Sano H, Chambers J, Gieseg M. Evaluation of an indirect oscillometric blood pressure monitor in anaesthetised dogs at three different anatomical locations. N Z Vet J. 2017;65(4):185\u2013191. doi:10.1080\/00480169.2017.1295000"},{"issue":"1","key":"B12","series-title":"Vet Anaesth Analg","first-page":"21","article-title":"Comparison of high definition oscillometric and Doppler ultrasound devices with invasive blood pressure in anaesthetized dogs","author":"Seli\u0161kar A","year":"2013","unstructured":"Seli\u0161kar A, Zrim\u0161ek P, Sreden\u0161ek J, Petri\u010d AD. Comparison of high definition oscillometric and Doppler ultrasound devices with invasive blood pressure in anaesthetized dogs. Vet Anaesth Analg. 2013;40(1):21\u201327. doi:10.1111\/j.1467-2995.2012.00774.x"},{"issue":"5","key":"B13","series-title":"Vet Anaesth Analg","first-page":"503","article-title":"A clinical comparison between a non\u2013invasive blood pressure monitor using high definition oscillometry (Memodiagnostic MD 15\/90 Pro) and invasive arterial blood pressure measurement in anaesthetized dogs","author":"Rysnik MK","year":"2013","unstructured":"Rysnik MK, Cripps P, Iff I. A clinical comparison between a non\u2013invasive blood pressure monitor using high definition oscillometry (Memodiagnostic MD 15\/90 Pro) and invasive arterial blood pressure measurement in anaesthetized dogs. Vet Anaesth Analg. 2013;40(5):503\u2013511. doi:10.1111\/vaa.12035"},{"issue":"5","key":"B14","series-title":"J Vet Emerg Crit Care","first-page":"523","article-title":"Comparison of invasive versus noninvasive blood pressure measurements before and after hemorrhage in anesthetized greyhounds using the Surgivet V9203","author":"Drynan EA","year":"2013","unstructured":"Drynan EA, Raisis AL. Comparison of invasive versus noninvasive blood pressure measurements before and after hemorrhage in anesthetized greyhounds using the Surgivet V9203. J Vet Emerg Crit Care. 2013;23(5):523\u2013531. doi:10.1111\/vec.12075"},{"issue":"3","key":"B15","series-title":"Vet Res Commun","first-page":"217","article-title":"Comparison of invasive and non-invasive blood pressure monitoring during clinical anaesthesia in dogs","author":"MacFarlane PD","year":"2010","unstructured":"MacFarlane PD, Grint N, Dugdale A. Comparison of invasive and non-invasive blood pressure monitoring during clinical anaesthesia in dogs. Vet Res Commun. 2010;34(3):217\u2013227. doi:10.1007\/s11259-010-9346-9"},{"issue":"1","key":"B16","series-title":"Am J Vet Res","first-page":"59","article-title":"Comparison of blood pressure measurements of anesthetized dogs obtained noninvasively with a cylindrical blood pressure cuff and an anatomically modified conical blood pressure cuff","author":"Ramos SJ","year":"2016","unstructured":"Ramos SJ, Da Cunha AF, Domingues M, Shelby AM, Stout RW, Acierno MJ. Comparison of blood pressure measurements of anesthetized dogs obtained noninvasively with a cylindrical blood pressure cuff and an anatomically modified conical blood pressure cuff. Am J Vet Res. 2016;77(1):59\u201364. doi:10.2460\/ajvr.77.1.59"},{"issue":"2","key":"B17","series-title":"Vet Anaesth Analg","first-page":"199","article-title":"Agreement between two oscillometric blood pressure technologies and invasively measured arterial pressure in the dog","author":"Da Cunha AF","year":"2016","unstructured":"Da Cunha AF, Ramos SJ, Domingues M, et al. Agreement between two oscillometric blood pressure technologies and invasively measured arterial pressure in the dog. Vet Anaesth Analg. 2016;43(2):199\u2013203. doi:10.1111\/vaa.12312"},{"issue":"3","key":"B18","series-title":"Am J Vet Res","first-page":"266","article-title":"Comparison of directly measured arterial blood pressure at various anatomic locations in anesthetized dogs","author":"Acierno MJ","year":"2015","unstructured":"Acierno MJ, Domingues ME, Ramos SJ, Shelby AM, Da Cunha AF. Comparison of directly measured arterial blood pressure at various anatomic locations in anesthetized dogs. Am J Vet Res. 2015;76(3):266\u2013271. doi:10.2460\/ajvr.76.3.266"},{"issue":"6","key":"B19","series-title":"Am J Vet Res","first-page":"ajvr.23.11.0263","article-title":"Doppler ultrasound is more accurate than pulse oximeter plethysmography in the measurement of systolic arterial pressure from the median caudal artery in anesthetized dogs","author":"Whittaker JH","year":"2024","unstructured":"Whittaker JH, Kapaldo N, Bortoluzzi EM, Rankin DC. Doppler ultrasound is more accurate than pulse oximeter plethysmography in the measurement of systolic arterial pressure from the median caudal artery in anesthetized dogs. Am J Vet Res. 2024;85(6):ajvr.23.11.0263. doi:10.2460\/ajvr.23.11.0263"},{"issue":"2","key":"B20","series-title":"Vet Anaesth Analg","first-page":"135","article-title":"Validity and reliability of Doppler ultrasonography and direct arterial blood pressure measurements in anaesthetized dogs weighing less than 5 kg","author":"Moll X","year":"2018","unstructured":"Moll X, Aguilar A, Garc\u00eda F, Ferrer R, Andaluz A. Validity and reliability of Doppler ultrasonography and direct arterial blood pressure measurements in anaesthetized dogs weighing less than 5 kg. Vet Anaesth Analg. 2018;45(2):135\u2013144. doi:10.1016\/j.vaa.2017.08.006"},{"issue":"8","key":"B21","series-title":"J Small Anim Pract","first-page":"474","article-title":"Evaluation of oscillometric blood pressure monitor BLT M9000 VET in anaesthetised healthy adult dogs","author":"Zrim\u0161ek P","year":"2018","unstructured":"Zrim\u0161ek P, Sreden\u0161ek J, Vengu\u0161t M, Seli\u0161kar A. Evaluation of oscillometric blood pressure monitor BLT M9000 VET in anaesthetised healthy adult dogs. J Small Anim Pract. 2018;59(8):474\u2013479. doi:10.1111\/jsap.12811"}],"container-title":["American Journal of Veterinary Research"],"original-title":[],"link":[{"URL":"https:\/\/avmajournals.avma.org\/view\/journals\/ajvr\/87\/8\/ajvr.26.03.0108.xml","content-type":"text\/html","content-version":"vor","intended-application":"text-mining"},{"URL":"https:\/\/avmajournals.avma.org\/downloadpdf\/journals\/ajvr\/87\/8\/ajvr.26.03.0108.xml","content-type":"text\/html","content-version":"vor","intended-application":"syndication"},{"URL":"https:\/\/avmajournals.avma.org\/downloadpdf\/journals\/ajvr\/87\/8\/ajvr.26.03.0108.xml","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2026,7,31]],"date-time":"2026-07-31T15:25:32Z","timestamp":1785511532000},"score":1,"resource":{"primary":{"URL":"https:\/\/avmajournals.avma.org\/view\/journals\/ajvr\/87\/8\/ajvr.26.03.0108.xml"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[2026,8,1]]},"references-count":21,"journal-issue":{"issue":"8"},"URL":"https:\/\/doi.org\/10.2460\/ajvr.26.03.0108","relation":{},"ISSN":["0002-9645","1943-5681"],"issn-type":[{"value":"0002-9645","type":"print"},{"value":"1943-5681","type":"electronic"}],"subject":[],"published":{"date-parts":[[2026,8,1]]}}}