{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2026,6,3]],"date-time":"2026-06-03T14:26:31Z","timestamp":1780496791524,"version":"3.54.1"},"reference-count":17,"publisher":"Journal of Neurosurgery Publishing Group (JNSPG)","issue":"3","content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":["SPI"],"published-print":{"date-parts":[[2007,9]]},"abstract":"<jats:p>\u2713Instrumented spinal fusion in patients with osteoporosis is challenging because of the poor bone quality and is complicated by an elevated risk of delayed hardware failure. The author treated two patients presenting with severe osteoporosis, spinal stenosis, and degenerative spondylolisthesis. He performed decompressive laminectomy, posterolateral fusion, and pedicle screw (PS) fixation involving screws with side openings that allow cement to be injected through the implant. The cement injection was conducted under fluoroscopic control without complications.<\/jats:p>\n<jats:p>Although this technique needs validation in a larger population of patients, the author believes that the injection of cement through these PSs can be performed safely in carefully selected patients. This technique creates not only a vertebroplasty-like effect that strengthens the vertebral body but also provides the additional stability afforded by the immediate anchoring of the screw, which may allow a shorter-length construct, save mobile segments, and finally reduce the risk of hardware failure.<\/jats:p>","DOI":"10.3171\/spi-07\/09\/366","type":"journal-article","created":{"date-parts":[[2008,1,24]],"date-time":"2008-01-24T12:23:52Z","timestamp":1201177432000},"page":"366-369","source":"Crossref","is-referenced-by-count":52,"title":["Increasing pedicle screw anchoring in the osteoporotic spine by cement injection through the implant"],"prefix":"10.3171","volume":"7","author":[{"given":"Patrick","family":"Fransen","sequence":"first","affiliation":[],"role":[{"vocabulary":"crossref","role":"author"}]}],"member":"1754","reference":[{"key":"b1-0070366","series-title":"J Orthop Res","doi-asserted-by":"publisher","first-page":"1217","DOI":"10.1016\/S0736-0266(02)00069-4","article-title":"Pedicular fixation in the osteoporotic spine: a pilot in vivo study on long-term ovariectomized sheep","volume":"20","author":"Aldini","year":"2002"},{"key":"b2-0070366","series-title":"J Biomech Eng","doi-asserted-by":"publisher","first-page":"391","DOI":"10.1115\/1.2796022","article-title":"Factors affecting the pullout strength of cancellous bone screws","volume":"118","author":"Chapman","year":"1996"},{"key":"b3-0070366","series-title":"Spine J","doi-asserted-by":"publisher","first-page":"109","DOI":"10.1016\/S1529-9430(01)00020-1","article-title":"Lumbosacral fixation using expandable pedicle screws. An alternative to reoperation in osteoporosis","volume":"1","author":"Cook","year":"2001"},{"key":"b4-0070366","series-title":"Spine J","doi-asserted-by":"publisher","first-page":"402","DOI":"10.1016\/j.spinee.2003.11.010","article-title":"Bio-mechanical study of pedicle screw fixation in severely osteoporotic bone","volume":"4","author":"Cook","year":"2004"},{"key":"b5-0070366","series-title":"Arch Orthop Trauma Surg","doi-asserted-by":"publisher","first-page":"554","DOI":"10.1007\/s00402-006-0185-7","article-title":"Osseointegration of hollow cylinder based implants in normal and osteoporotic vertebrae: a sheep study","volume":"126","author":"Goldhahn","year":"2006"},{"key":"b6-0070366","series-title":"Spine","doi-asserted-by":"publisher","first-page":"2516","DOI":"10.1097\/01.brs.0000144833.63581.c1","article-title":"Reoperation rate after instrumented posterior lumbar interbody fusion: a report on 1680 cases","volume":"29","author":"Greiner-Perth","year":"2004"},{"key":"b7-0070366","series-title":"Spine J","doi-asserted-by":"publisher","first-page":"239","DOI":"10.1016\/j.spinee.2004.11.010","article-title":"Hydroxyapatite coating of pedicle screws improves resistance against pull-out force in the osteoporotic canine lumbar spine model: a pilot study","volume":"5","author":"Hasegawa","year":"2005"},{"key":"b8-0070366","series-title":"Nippon Rinsho","first-page":"1670","article-title":"[Surgical treatment for osteoporotic spinal fractures in the thoracolumbar spine.]","volume":"64","author":"Ito","year":"2006"},{"key":"b9-0070366","series-title":"Clin Biomech (Bristol, Avon)","doi-asserted-by":"publisher","first-page":"676","DOI":"10.1016\/j.clinbiomech.2006.02.005","article-title":"Biomechanical performance of the new BeadEx inplant in the treatment of osteoporotic vertebral body compression fractures: restoration and maintenance of height and stability","volume":"21","author":"Kettler","year":"2006"},{"key":"b10-0070366","series-title":"Z Orthop Ihre Grenzgeb","first-page":"464","article-title":"[Increasing bone screw anchoring in the femur head by cement administration via the implant\u2014a biomechanical study.]","volume":"138","author":"Kramer","year":"2000"},{"key":"b11-0070366","series-title":"Clin Biomech (Bristol, Avon)","doi-asserted-by":"publisher","first-page":"S25","DOI":"10.1016\/S0268-0033(03)00081-0","article-title":"A biomechanical study on the cortex-anchorage vertebral screw","volume":"18","author":"Lin","year":"2003"},{"key":"b12-0070366","series-title":"J Orthop Res","doi-asserted-by":"publisher","first-page":"545","DOI":"10.1016\/S0736-0266(00)00052-8","article-title":"An expandable anchor for fixation in osteoporotic bone","volume":"19","author":"McKoy","year":"2001"},{"key":"b13-0070366","series-title":"Spine","doi-asserted-by":"publisher","first-page":"E867","DOI":"10.1097\/01.brs.0000244658.35865.59","article-title":"A comparison of pullout strength for pedicle screws of different designs: a study using tapped and untapped pilot holes","volume":"31","author":"Pfeiffer","year":"2006"},{"issue":"1 Suppl","key":"b14-0070366","series-title":"J Neurosurg","first-page":"50","article-title":"Enhanced primary stability through additional cementable cannulated rescue screw for anterior thoracolumbar plate application","volume":"98","author":"Schultheiss","year":"2003"},{"issue":"2 Suppl","key":"b15-0070366","series-title":"Osteoporos Int","doi-asserted-by":"publisher","first-page":"S107","DOI":"10.1007\/s00198-004-1777-0","article-title":"Mechanical behavior of screws in normal and osteoporotic bone","volume":"16","author":"Seebeck","year":"2005"},{"key":"b16-0070366","series-title":"J Spinal Disord Tech","doi-asserted-by":"publisher","first-page":"413","DOI":"10.1097\/01.bsd.0000173840.59099.06","article-title":"Open vertebral cement augmentation combined with lumbar decompression for the operative management of thoracolumbar stenosis secondary to osteoporotic burst fractures","volume":"18","author":"Singh","year":"2005"},{"key":"b17-0070366","series-title":"Eur Spine J","doi-asserted-by":"publisher","first-page":"528","DOI":"10.1007\/s005860000169","article-title":"Augmentation of (pedicle) screws with calcium apatite cement in patients with severe progressive osteoporotic spinal deformities: an innovative technique","volume":"9","author":"Wuisman","year":"2000"}],"container-title":["Journal of Neurosurgery: Spine"],"original-title":[],"link":[{"URL":"https:\/\/thejns.org\/view\/journals\/j-neurosurg-spine\/7\/3\/article-p366.xml","content-type":"text\/html","content-version":"vor","intended-application":"text-mining"},{"URL":"https:\/\/thejns.org\/downloadpdf\/journals\/j-neurosurg-spine\/7\/3\/article-p366.xml","content-type":"text\/html","content-version":"vor","intended-application":"syndication"},{"URL":"https:\/\/thejns.org\/downloadpdf\/journals\/j-neurosurg-spine\/7\/3\/article-p366.xml","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2023,4,15]],"date-time":"2023-04-15T23:27:22Z","timestamp":1681601242000},"score":1,"resource":{"primary":{"URL":"https:\/\/thejns.org\/view\/journals\/j-neurosurg-spine\/7\/3\/article-p366.xml"}},"subtitle":["Technical note and report of three cases"],"short-title":[],"issued":{"date-parts":[[2007,9]]},"references-count":17,"journal-issue":{"issue":"3"},"URL":"https:\/\/doi.org\/10.3171\/spi-07\/09\/366","relation":{},"ISSN":["1547-5654"],"issn-type":[{"value":"1547-5654","type":"print"}],"subject":[],"published":{"date-parts":[[2007,9]]}}}