{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2026,4,29]],"date-time":"2026-04-29T12:12:20Z","timestamp":1777464740923,"version":"3.51.4"},"reference-count":25,"publisher":"SAGE Publications","issue":"5","content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":["BME"],"published-print":{"date-parts":[[2020,11,13]]},"abstract":"<jats:p>BACKGROUND: The stability of the implant-abutment interface is an important factor that influences load distribution on the marginal bone. OBJECTIVE: In this study, three dental implants with the same connection were subjected to different dynamic loading cycles. The fracture strengths and the horizontal compatibility of implants were assessed. METHODS: Eighty four implant specimens were embedded in a polyacetal cylinder as simulated bone loss of 3 mm from the implant platform. Three of the implants were used to determine the endurance limit. The other specimens were subdivided into four subgroups (n = 6): three for dynamic + static loading, and one for static loading (control group). The tests were performed by applying a compression load. The dynamic loading experiments included three different cycles with endurance upper limit loads at a frequency of 10 Hz. RESULTS: The differences between the fracture strength values of the implant brands were found to be statistically significant. However, there were no meaningful differences between the fracture strength values of implants of the same brand. The specimens of the DTI implant system had the lowest strength (647.9 \u00b1 41.5\u00a0N) and the SEM analysis indicated that the Implantium implant system had the shortest horizontal gaps. CONCLUSIONS: There was a negative correlation between the fracture strengths and size of the microgaps. The importance of these in vitro results needs to be validated by clinical trials because the loads in the mouth can be applied from various angles.<\/jats:p>","DOI":"10.3233\/bme-206008","type":"journal-article","created":{"date-parts":[[2020,10,30]],"date-time":"2020-10-30T12:28:46Z","timestamp":1604060926000},"page":"319-328","source":"Crossref","is-referenced-by-count":0,"title":["Dynamic and static load performance of dental biomaterial systems with conical implant-abutment connections"],"prefix":"10.1177","volume":"31","author":[{"given":"Mehmet Can","family":"Basgil","sequence":"first","affiliation":[{"name":"Department of Prosthodontics, Istanbul Okmeydani Dental Health Hospital, Istanbul, Turkey"}],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Cem","family":"Kurtoglu","sequence":"additional","affiliation":[{"name":", , Cukurova University, , Turkey"}],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Koray","family":"Soygun","sequence":"additional","affiliation":[{"name":", , Cukurova University, , Turkey"}],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Yunus","family":"Uslan","sequence":"additional","affiliation":[{"name":", TOBB University of Economics and Technology, , Turkey"}],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Teyfik","family":"Demi\u0307r","sequence":"additional","affiliation":[{"name":", TOBB University of Economics and Technology, , Turkey"}],"role":[{"role":"author","vocabulary":"crossref"}]}],"member":"179","reference":[{"issue":"5","key":"10.3233\/BME-206008_ref1","first-page":"528","article-title":"Cement-retained versus screw-retained implant restorations: A critical review","volume":"18","author":"Michalakis","year":"2003","journal-title":"Int J Oral Max Impl"},{"issue":"4","key":"10.3233\/BME-206008_ref2","first-page":"540","article-title":"Microbial leakage and marginal fit of the implant-abutment interface","volume":"12","author":"Jansen","year":"1997","journal-title":"Int J Oral Max Impl"},{"issue":"1","key":"10.3233\/BME-206008_ref3","doi-asserted-by":"crossref","first-page":"19","DOI":"10.1034\/j.1600-0501.2002.130101.x","article-title":"Infectious risks for oral implants: A review of the literature","volume":"13","author":"Quirynen","year":"2002","journal-title":"Clin Oral Implan Res"},{"issue":"6","key":"10.3233\/BME-206008_ref4","doi-asserted-by":"crossref","first-page":"599","DOI":"10.1016\/0022-3913(94)90443-X","article-title":"Tightening characteristics for screwed joints in osseointegrated dental implants","volume":"71","author":"Burguete","year":"1994","journal-title":"J Prosthet Dent"},{"issue":"2","key":"10.3233\/BME-206008_ref5","first-page":"463","article-title":"Influence of implant\/abutment joint designs on abutment screw loosening in a dental implant system","volume":"75","author":"Kitagawa","year":"2005","journal-title":"J Biomed Mater Res B Appl Biomater"},{"issue":"1","key":"10.3233\/BME-206008_ref6","doi-asserted-by":"crossref","first-page":"6","DOI":"10.1563\/1548-1336(2008)34[1:AFEAOT]2.0.CO;2","article-title":"A finite element analysis of two different dental implants: Stress distribution in the prosthesis, abutment, implant, and supporting bone","volume":"34","author":"Quaresma","year":"2008","journal-title":"J Oral Implantol"},{"issue":"5","key":"10.3233\/BME-206008_ref7","doi-asserted-by":"crossref","first-page":"416","DOI":"10.1111\/j.1600-0722.2007.00473.x","article-title":"Factorial analysis of variables influencing mechanical characteristics of a single tooth implant placed in the maxilla using finite element analysis and the statistics-based Taguchi method","volume":"115","author":"Lin","year":"2007","journal-title":"Eur J Oral Sci"},{"issue":"1","key":"10.3233\/BME-206008_ref8","doi-asserted-by":"crossref","first-page":"97","DOI":"10.1016\/j.jbiomech.2004.03.006","article-title":"Mechanics of the taper integrated screwed-in (TIS) abutments used in dental implants","volume":"38","author":"Bozkaya","year":"2005","journal-title":"J Biomech"},{"issue":"5","key":"10.3233\/BME-206008_ref9","doi-asserted-by":"crossref","first-page":"464","DOI":"10.1034\/j.1600-0501.2000.011005458.x","article-title":"An in vitro evaluation of the strength of a 1-piece and 2-piece conical abutment joint in implant design","volume":"11","author":"Norton","year":"2000","journal-title":"Clin Oral Implants Res"},{"issue":"3","key":"10.3233\/BME-206008_ref10","first-page":"286","article-title":"Fracture resistance of the implant-abutment connection in implants with internal hex and internal conical connections under oblique compressive loading: An in vitro study","volume":"22","author":"Copped\u00ea","year":"2009","journal-title":"Int J Prosthodont"},{"issue":"1","key":"10.3233\/BME-206008_ref11","first-page":"26","article-title":"Zirconium implant abutments: Fracture strength and influence of cyclic loading on retaining-screw loosening","volume":"37","author":"Gehrke","year":"2006","journal-title":"Quintessence Int"},{"issue":"2","key":"10.3233\/BME-206008_ref12","doi-asserted-by":"crossref","first-page":"32","DOI":"10.1097\/ID.0b013e318211fce8","article-title":"Comparative strain gauge analysis of external and internal hexagon, Morse taper, and influence of straight and ofset implant configuration","volume":"20","author":"Nishioka","year":"2011","journal-title":"Implant Dent"},{"issue":"1","key":"10.3233\/BME-206008_ref13","doi-asserted-by":"crossref","first-page":"78","DOI":"10.1111\/j.1365-2842.2006.01545.x","article-title":"In vitro differences of stress concentrations for internal and external hex implant-abutment connections: A short communication","volume":"33","author":"Maeda","year":"2006","journal-title":"J Oral Rehabil"},{"issue":"1","key":"10.3233\/BME-206008_ref14","doi-asserted-by":"crossref","first-page":"24","DOI":"10.1016\/j.jpor.2011.02.002","article-title":"Influence of the interface design on the yield force of the implant\u2013abutment complex before and after cyclic mechanical loading","volume":"56","author":"Dittmer","year":"2012","journal-title":"J Prosthodont Res"},{"issue":"1","key":"10.3233\/BME-206008_ref15","doi-asserted-by":"crossref","first-page":"18","DOI":"10.1111\/j.1600-0501.2010.02141.x","article-title":"Bending moments of zirconia and titanium abutments with internal and external implant-abutment connections after aging and chewing simulation","volume":"23","author":"Truninger","year":"2012","journal-title":"Clin Oral Implants Res"},{"issue":"5","key":"10.3233\/BME-206008_ref16","first-page":"617","article-title":"A study of the bending resistance of implant-supported reinforced alumina and machined zirconia abutments and copies","volume":"24","author":"Sundh","year":"2008","journal-title":"Dent Mate"},{"issue":"6","key":"10.3233\/BME-206008_ref17","first-page":"671","article-title":"Evaluation of an endosseous oral implant system according to UNI EN ISO 14801 fatigue test protocol","volume":"23","author":"Marchetti","year":"2014","journal-title":"Implant Dent"},{"issue":"1","key":"10.3233\/BME-206008_ref18","doi-asserted-by":"crossref","first-page":"7","DOI":"10.1563\/0-809.1","article-title":"Comparison of the compressive strength of 3 different implant design systems","volume":"33","author":"Pedroza","year":"2007","journal-title":"J Oral Implantol"},{"key":"10.3233\/BME-206008_ref19","doi-asserted-by":"crossref","first-page":"346","DOI":"10.1016\/j.jmbbm.2016.05.011","article-title":"Fracture strength of implant abutments after fatigue testing: A systematic review and a meta-analysis","volume":"62","author":"Coray","year":"2016","journal-title":"J Mech Behav Biomed Mater"},{"issue":"2","key":"10.3233\/BME-206008_ref20","doi-asserted-by":"crossref","first-page":"202","DOI":"10.1111\/j.1600-0501.2012.02443.x","article-title":"Reliability and failure modes of internal conical dental implant connections","volume":"24","author":"Freitas-J\u00fanior","year":"2013","journal-title":"Clin Oral Implants Res"},{"issue":"4","key":"10.3233\/BME-206008_ref21","doi-asserted-by":"crossref","first-page":"416","DOI":"10.1016\/j.dental.2014.01.005","article-title":"Influence of abutment-to-fixture design on reliability and failure mode of all-ceramic crown systems","volume":"30","author":"Delben","year":"2014","journal-title":"Dent Mater"},{"issue":"1","key":"10.3233\/BME-206008_ref22","doi-asserted-by":"crossref","first-page":"118","DOI":"10.1902\/jop.2011.110087","article-title":"Influences of internal tapered abutment designs on bone stresses around a dental implant: Three-dimensional finite element method with statistical evaluation","volume":"83","author":"Chu","year":"2012","journal-title":"J Periodontol"},{"issue":"4","key":"10.3233\/BME-206008_ref23","doi-asserted-by":"crossref","first-page":"266","DOI":"10.1097\/00008505-199807040-00003","article-title":"Loss of crestal bone around dental implants: A retrospective study","volume":"7","author":"Callan","year":"1998","journal-title":"Implant Dent"},{"issue":"9","key":"10.3233\/BME-206008_ref24","doi-asserted-by":"crossref","first-page":"1083","DOI":"10.1111\/clr.12217","article-title":"Microleakage of the prosthetic abutment\/implant interface with internal and external connection: In vitro study","volume":"25","author":"Larrucea Verdugo","year":"2014","journal-title":"Clin Oral Implants Res"},{"issue":"3","key":"10.3233\/BME-206008_ref25","doi-asserted-by":"crossref","first-page":"685","DOI":"10.1097\/SCS.0000000000002563","article-title":"Evaluation of microgap with three-dimensional x-ray microtomography: Internal hexagon versus cone morse","volume":"27","author":"Scarano","year":"2016","journal-title":"J Craniofac Surg"}],"container-title":["Bio-Medical Materials and Engineering"],"original-title":[],"link":[{"URL":"https:\/\/content.iospress.com\/download?id=10.3233\/BME-206008","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2026,4,28]],"date-time":"2026-04-28T13:23:59Z","timestamp":1777382639000},"score":1,"resource":{"primary":{"URL":"https:\/\/journals.sagepub.com\/doi\/full\/10.3233\/BME-206008"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[2020,11,13]]},"references-count":25,"journal-issue":{"issue":"5"},"URL":"https:\/\/doi.org\/10.3233\/bme-206008","relation":{},"ISSN":["0959-2989","1878-3619"],"issn-type":[{"value":"0959-2989","type":"print"},{"value":"1878-3619","type":"electronic"}],"subject":[],"published":{"date-parts":[[2020,11,13]]}}}