{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2026,11,13]],"date-time":"2026-11-13T06:28:43Z","timestamp":1794551323474,"version":"4.9.1"},"reference-count":68,"publisher":"MDPI AG","issue":"1","license":[{"start":{"date-parts":[[2021,12,29]],"date-time":"2021-12-29T00:00:00Z","timestamp":1640736000000},"content-version":"vor","delay-in-days":0,"URL":"https:\/\/creativecommons.org\/licenses\/by\/4.0\/"}],"funder":[{"DOI":"10.13039\/100000038","name":"United States Food and Drug Administration","doi-asserted-by":"publisher","award":["16ECDRH407"],"award-info":[{"award-number":["16ECDRH407"]}],"id":[{"id":"10.13039\/100000038","id-type":"DOI","asserted-by":"publisher"}]}],"content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":["Sensors"],"abstract":"<jats:p>Infrared thermographs (IRTs) implemented according to standardized best practices have shown strong potential for detecting elevated body temperatures (EBT), which may be useful in clinical settings and during infectious disease epidemics. However, optimal IRT calibration methods have not been established and the clinical performance of these devices relative to the more common non-contact infrared thermometers (NCITs) remains unclear. In addition to confirming the findings of our preliminary analysis of clinical study results, the primary intent of this study was to compare methods for IRT calibration and identify best practices for assessing the performance of IRTs intended to detect EBT. A key secondary aim was to compare IRT clinical accuracy to that of NCITs. We performed a clinical thermographic imaging study of more than 1000 subjects, acquiring temperature data from several facial locations that, along with reference oral temperatures, were used to calibrate two IRT systems based on seven different regression methods. Oral temperatures imputed from facial data were used to evaluate IRT clinical accuracy based on metrics such as clinical bias (\u0394cb), repeatability, root-mean-square difference, and sensitivity\/specificity. We proposed several calibration approaches designed to account for the non-uniform data density across the temperature range and a constant offset approach tended to show better ability to detect EBT. As in our prior study, inner canthi or full-face maximum temperatures provided the highest clinical accuracy. With an optimal calibration approach, these methods achieved a \u0394cb between \u00b10.03 \u00b0C with standard deviation (\u03c3\u0394cb) less than 0.3 \u00b0C, and sensitivity\/specificity between 84% and 94%. Results of forehead-center measurements with NCITs or IRTs indicated reduced performance. An analysis of the complete clinical data set confirms the essential findings of our preliminary evaluation, with minor differences. Our findings provide novel insights into methods and metrics for the clinical accuracy assessment of IRTs. Furthermore, our results indicate that calibration approaches providing the highest clinical accuracy in the 37\u201338.5 \u00b0C range may be most effective for measuring EBT. While device performance depends on many factors, IRTs can provide superior performance to NCITs.<\/jats:p>","DOI":"10.3390\/s22010215","type":"journal-article","created":{"date-parts":[[2021,12,29]],"date-time":"2021-12-29T08:12:15Z","timestamp":1640765535000},"page":"215","update-policy":"https:\/\/doi.org\/10.3390\/mdpi_crossmark_policy","source":"Crossref","is-referenced-by-count":44,"title":["Infrared Thermography for Measuring Elevated Body Temperature: Clinical Accuracy, Calibration, and Evaluation"],"prefix":"10.3390","volume":"22","author":[{"ORCID":"https:\/\/orcid.org\/0000-0002-0557-7058","authenticated-orcid":false,"given":"Quanzeng","family":"Wang","sequence":"first","affiliation":[{"name":"Center for Devices and Radiological Health, Food and Drug Administration, Silver Spring, MD 20993, USA"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Yangling","family":"Zhou","sequence":"additional","affiliation":[{"name":"Center for Devices and Radiological Health, Food and Drug Administration, Silver Spring, MD 20993, USA"},{"name":"Department of Mechanical Engineering, University of Maryland, Baltimore County, Baltimore, MD 21250, USA"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"ORCID":"https:\/\/orcid.org\/0000-0003-4810-9454","authenticated-orcid":false,"given":"Pejman","family":"Ghassemi","sequence":"additional","affiliation":[{"name":"Center for Devices and Radiological Health, Food and Drug Administration, Silver Spring, MD 20993, USA"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"David","family":"McBride","sequence":"additional","affiliation":[{"name":"University Health Center, University of Maryland, College Park, MD 20742, USA"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Jon P.","family":"Casamento","sequence":"additional","affiliation":[{"name":"Center for Devices and Radiological Health, Food and Drug Administration, Silver Spring, MD 20993, USA"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"T. Joshua","family":"Pfefer","sequence":"additional","affiliation":[{"name":"Center for Devices and Radiological Health, Food and Drug Administration, Silver Spring, MD 20993, USA"}],"role":[{"vocabulary":"crossref","role":"author"}]}],"member":"1968","published-online":{"date-parts":[[2021,12,29]]},"reference":[{"key":"ref_1","doi-asserted-by":"crossref","first-page":"26","DOI":"10.1177\/101053950501700107","article-title":"Infrared thermography to mass-screen suspected SARS patients with fever","volume":"17","author":"Chiu","year":"2005","journal-title":"Asia-Pac. J. Public Health"},{"key":"ref_2","doi-asserted-by":"crossref","unstructured":"Nishiura, H., and Kamiya, K. (2011). Fever screening during the influenza (H1N1-2009) pandemic at Narita International Airport, Japan. BMC Infect. Dis., 11.","DOI":"10.1186\/1471-2334-11-111"},{"key":"ref_3","doi-asserted-by":"crossref","first-page":"425","DOI":"10.1016\/S1473-3099(20)30086-4","article-title":"Radiological findings from 81 patients with COVID-19 pneumonia in Wuhan, China: A descriptive study","volume":"20","author":"Shi","year":"2020","journal-title":"Lancet Infect. Dis."},{"key":"ref_4","doi-asserted-by":"crossref","first-page":"475","DOI":"10.1016\/S2213-2600(20)30079-5","article-title":"Clinical course and outcomes of critically ill patients with SARS-CoV-2 pneumonia in Wuhan, China: A single-centered, retrospective, observational study","volume":"8","author":"Yang","year":"2020","journal-title":"Lancet Respir. Med."},{"key":"ref_5","first-page":"442","article-title":"Ebola virus disease: A review on epidemiology, symptoms, treatment and pathogenesis","volume":"72","author":"Goeijenbier","year":"2014","journal-title":"Neth. J. Med."},{"key":"ref_6","doi-asserted-by":"crossref","first-page":"497","DOI":"10.1016\/S0140-6736(20)30183-5","article-title":"Clinical features of patients infected with 2019 novel coronavirus in Wuhan, China","volume":"395","author":"Huang","year":"2020","journal-title":"Lancet"},{"key":"ref_7","doi-asserted-by":"crossref","first-page":"551","DOI":"10.15585\/mmwr.mm6918e2","article-title":"Public health response to the initiation and spread of pandemic COVID-19 in the United States, February 24\u2013April 21, 2020","volume":"69","author":"Schuchat","year":"2020","journal-title":"Morb. Mortal. Weekly Rep."},{"key":"ref_8","unstructured":"Widmaier, E.P., Raff, H., and Strang, K.T. (2008). Regulation of Organic Metabolism and Energy Balance-Section B: Regulation of Total-Body Energy Balance and Temperature. Vander\u2019s Human Physiology, McGraw-Hill."},{"key":"ref_9","doi-asserted-by":"crossref","first-page":"661","DOI":"10.1016\/j.ijnurstu.2008.11.006","article-title":"Normal body temperature and the effects of age, sex, ambient temperature and body mass index on normal oral temperature: A prospective, comparative study","volume":"46","author":"Lu","year":"2009","journal-title":"Int. J. Nurs. Stud."},{"key":"ref_10","doi-asserted-by":"crossref","first-page":"6593","DOI":"10.1167\/iovs.10-5659","article-title":"The relationship between body and ambient temperature and corneal temperature","volume":"51","author":"Kessel","year":"2010","journal-title":"Investig. Ophthalmol. Vis. Sci."},{"key":"ref_11","doi-asserted-by":"crossref","first-page":"358","DOI":"10.1055\/s-2008-1025792","article-title":"Exercise and the circadian variation in body temperature measures","volume":"7","author":"Reilly","year":"1986","journal-title":"Int. J. Sports Med."},{"key":"ref_12","doi-asserted-by":"crossref","first-page":"871","DOI":"10.1016\/j.metabol.2009.02.017","article-title":"Is obesity associated with lower body temperatures? Core temperature: A forgotten variable in energy balance","volume":"58","author":"Landsberg","year":"2009","journal-title":"Metabolism"},{"key":"ref_13","doi-asserted-by":"crossref","first-page":"249","DOI":"10.1081\/CBI-100103189","article-title":"Circadian rhythmicity of cortisol and body temperature: Morningness-eveningness effects","volume":"18","author":"Bailey","year":"2001","journal-title":"Chronobiol. Int."},{"key":"ref_14","doi-asserted-by":"crossref","first-page":"3","DOI":"10.1186\/1740-3391-3-3","article-title":"Daily rhythm of cerebral blood flow velocity","volume":"3","author":"Conroy","year":"2005","journal-title":"J. Circadian Rhythm."},{"key":"ref_15","doi-asserted-by":"crossref","first-page":"289","DOI":"10.1159\/000333148","article-title":"Age-dependent changes in temperature regulation\u2013A mini review","volume":"58","author":"Blatteis","year":"2012","journal-title":"Gerontology"},{"key":"ref_16","doi-asserted-by":"crossref","first-page":"405","DOI":"10.1016\/0002-9378(72)90617-5","article-title":"A composite picture of the menstrual cycle","volume":"114","author":"Moghissi","year":"1972","journal-title":"Am. J. Obstet. Gynecol."},{"key":"ref_17","doi-asserted-by":"crossref","first-page":"1311","DOI":"10.1111\/j.1365-2702.2010.03565.x","article-title":"Performance of non-contact infrared thermometer for detecting febrile children in hospital and ambulatory settings","volume":"20","author":"Chiappini","year":"2011","journal-title":"J. Clin. Nurs."},{"key":"ref_18","doi-asserted-by":"crossref","first-page":"471","DOI":"10.1111\/j.1365-2214.2011.01264.x","article-title":"Clinical accuracy of a non-contact infrared skin thermometer in paediatric practice","volume":"38","author":"Teran","year":"2011","journal-title":"Child Care Health Dev."},{"key":"ref_19","doi-asserted-by":"crossref","first-page":"76","DOI":"10.1109\/MEMB.2008.931018","article-title":"Remote-sensing infrared thermography","volume":"28","author":"Ng","year":"2009","journal-title":"IEEE Eng. Med. Biol. Mag."},{"key":"ref_20","doi-asserted-by":"crossref","first-page":"19115","DOI":"10.2807\/ese.14.06.19115-en","article-title":"International travels and fever screening during epidemics: A literature review on the effectiveness and potential use of non-contact infrared thermometers","volume":"14","author":"Bitar","year":"2009","journal-title":"Eurosurveillance"},{"key":"ref_21","doi-asserted-by":"crossref","first-page":"305","DOI":"10.1097\/PEC.0b013e3182854465","article-title":"Mass screening for fever in children: A comparison of 3 infrared thermal detection systems","volume":"29","author":"Selent","year":"2013","journal-title":"Pediatr. Emerg. Care"},{"key":"ref_22","doi-asserted-by":"crossref","first-page":"1471","DOI":"10.1016\/j.puhe.2015.07.023","article-title":"Comparison of Infrared Thermal Detection Systems for mass fever screening in a tropical healthcare setting","volume":"129","author":"Tay","year":"2015","journal-title":"Public Health"},{"key":"ref_23","doi-asserted-by":"crossref","first-page":"1109","DOI":"10.1086\/502351","article-title":"Limitations of forehead infrared body temperature detection for fever screening for severe acute respiratory syndrome","volume":"25","author":"Liu","year":"2004","journal-title":"Infect. Control Hosp. Epidemiol."},{"key":"ref_24","doi-asserted-by":"crossref","first-page":"1710","DOI":"10.3201\/eid1611.100703","article-title":"Comparison of 3 infrared thermal detection systems and self-report for mass fever screening","volume":"16","author":"Nguyen","year":"2010","journal-title":"Emerg. Infect. Dis."},{"key":"ref_25","first-page":"109","article-title":"Utility of infrared thermography for screening febrile subjects","volume":"19","author":"Chan","year":"2013","journal-title":"Hong Kong Med. J."},{"key":"ref_26","doi-asserted-by":"crossref","first-page":"504","DOI":"10.1086\/659404","article-title":"Evaluation of an infrared thermal detection system for fever recognition during the H1N1 influenza pandemic","volume":"32","author":"Hewlett","year":"2011","journal-title":"Infect. Control Hosp. Epidemiol."},{"key":"ref_27","doi-asserted-by":"crossref","unstructured":"Priest, P.C., Duncan, A.R., Jennings, L.C., and Baker, M.G. (2011). Thermal Image Scanning for Influenza Border Screening: Results of an Airport Screening Study. PLoS ONE, 6.","DOI":"10.1371\/journal.pone.0014490"},{"key":"ref_28","doi-asserted-by":"crossref","first-page":"e2014004","DOI":"10.4178\/epih\/e2014004","article-title":"Fever screening and detection of febrile arrivals at an international airport in Korea: Association among self-reported fever, infrared thermal camera scanning, and tympanic temperature","volume":"36","author":"Cho","year":"2014","journal-title":"Epidemiol. Health"},{"key":"ref_29","doi-asserted-by":"crossref","unstructured":"Mouchtouri, V.A., Christoforidou, E.P., Lemos, C.M., Fanos, M., Rexroth, U., Grote, U., Belfroid, E., Swaan, C., and Hadjichristodoulou, C. (2019). Exit and entry screening practices for infectious diseases among travelers at points of entry: Looking for evidence on public health impact. Int. J. Env. Res. Public Health, 16.","DOI":"10.3390\/ijerph16234638"},{"key":"ref_30","unstructured":"IEC & ISO (2017). IEC 80601-2-59: Medical Electrical Equipment-Part 2-59: Particular Requirements for the Basic Safety and Essential Performance of Screening Thermographs for Human Febrile Temperature Screening, International Electrotechnical Commission, International Organization for Standardization."},{"key":"ref_31","unstructured":"ISO (2017). ISO\/TR 13154: Medical Electrical Equipment\u2014Deployment, Implementation and Operational Guidelines for Identifying Febrile Humans Using a Screening Thermograph, International Organization for Standardization."},{"key":"ref_32","doi-asserted-by":"crossref","unstructured":"Ghassemi, P., Pfefer, T.J., Casamento, J.P., Simpson, R., and Wang, Q. (2018). Best practices for standardized performance testing of infrared thermographs intended for fever screening. PLoS ONE, 13.","DOI":"10.1371\/journal.pone.0203302"},{"key":"ref_33","unstructured":"ASTM (2016). ASTM E1965-98: Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature, ASTM Committee E20 on Temperature Measurement."},{"key":"ref_34","unstructured":"ISO (2017). ISO 80601-2-56: Medical Electrical Equipment-Part 2-56: Particular Requirements for Basic Safety and Essential Performance of Clinical Thermometers for Body Temperature Measurement, International Organization for Standardization."},{"key":"ref_35","unstructured":"Shiraki, K., and Yousef, M. (1987). Dilemma of body temperature measurement. Man in Stressful Environments: Thermal and Work Physiology, Charles C. Thomas."},{"key":"ref_36","doi-asserted-by":"crossref","first-page":"879","DOI":"10.2165\/00007256-200232140-00001","article-title":"Core temperature measurement","volume":"32","author":"Moran","year":"2002","journal-title":"Sports Med."},{"key":"ref_37","doi-asserted-by":"crossref","first-page":"769","DOI":"10.1016\/S0022-3476(06)80024-7","article-title":"Comparison of temperature measurements by an aural infrared thermometer with measurements by traditional rectal and axillary techniques","volume":"122","author":"Yetman","year":"1993","journal-title":"J. Pediatr."},{"key":"ref_38","doi-asserted-by":"crossref","first-page":"17","DOI":"10.1111\/j.1553-2712.1995.tb03072.x","article-title":"Cerumen occlusion lowers infrared tympanic membrane temperature measurement","volume":"2","author":"Doezema","year":"1995","journal-title":"Acad. Emerg. Med."},{"key":"ref_39","doi-asserted-by":"crossref","first-page":"331","DOI":"10.1007\/BF00423239","article-title":"Oral temperature as an index of core temperature during heat transients","volume":"50","author":"Mairiaux","year":"1983","journal-title":"Eur. J. Appl. Physiol. Occup. Physiol."},{"key":"ref_40","doi-asserted-by":"crossref","first-page":"ofz032","DOI":"10.1093\/ofid\/ofz032","article-title":"Normal body temperature: A systematic review","volume":"6","author":"Geneva","year":"2019","journal-title":"Open Forum Infect. Dis."},{"key":"ref_41","first-page":"097002","article-title":"Clinical evaluation of fever-screening thermography: Impact of consensus guidelines and facial measurement location","volume":"25","author":"Zhou","year":"2020","journal-title":"J. Biomed. Opt."},{"key":"ref_42","unstructured":"Acharya, U.R., Ng, Y.K.E., and Suri, J.S. (2008). Clinical Implications for Thermography in the Eye World: A short History of Clinical Ocular Thermography. Image Modeling of the Human Eye, Artech House."},{"key":"ref_43","doi-asserted-by":"crossref","first-page":"686","DOI":"10.1088\/0031-9155\/18\/5\/307","article-title":"Spectral emissivity of skin and pericardium","volume":"18","author":"Steketee","year":"1973","journal-title":"Phys. Med. Biol."},{"key":"ref_44","unstructured":"Tk\u00e1\u010dov\u00e1, M., \u017div\u010d\u00e1k, J., and Foffov\u00e1, P. (2011, January 27\u201330). A Reference for Human Eye Surface Temperature Measurements in Diagnostic Process of Ophthalmologic Diseases. Proceedings of the Measurement 2011, Smolenice, Slovakia."},{"key":"ref_45","first-page":"22079","article-title":"Clinical Evaluation of Non-Contact Infrared Thermometers","volume":"11","author":"Sullivan","year":"2021","journal-title":"Res. Sq."},{"key":"ref_46","doi-asserted-by":"crossref","unstructured":"Chenna, Y.N.D., Ghassemi, P., Pfefer, T.J., Casamento, J., and Wang, Q. (2018). Free-form deformation approach for registration of visible and infrared facial images in fever screening. Sensors, 18.","DOI":"10.3390\/s18010125"},{"key":"ref_47","first-page":"869","article-title":"Infrared ear thermometry","volume":"54","author":"Ng","year":"2004","journal-title":"Br. J. Gen. Pract."},{"key":"ref_48","first-page":"67","article-title":"Fever screening and infrared thermal imaging: Concerns and guidelines","volume":"19","author":"Mercer","year":"2009","journal-title":"Thermol. Int."},{"key":"ref_49","unstructured":"Walker, H.K., Hall, W.D.H., and Hurst, J.W. (1990). Temperature. Clinical Methods: The History, Physical, and Laboratory Examinations, Butterworth Publishers, a Division of Reed Publishing."},{"key":"ref_50","unstructured":"Clinical and Laboratory Standards Institute (2018). EP09c: Measurement Procedure Comparison and Bias Estimation Using Patient Samples, Clinical and Laboratory Standards Institute."},{"key":"ref_51","unstructured":"ISO (2017). ISO 80601-2-61: Medical Electrical Equipment\u2014Part 2-61: Particular Requirements for Basic Safety and Essential Performance of Pulse Oximeter Equipment, International Organization for Standardization."},{"key":"ref_52","doi-asserted-by":"crossref","first-page":"277","DOI":"10.1007\/s13312-011-0055-4","article-title":"Receiver operating characteristic (ROC) curve for medical researchers","volume":"48","author":"Kumar","year":"2011","journal-title":"Indian Pediatr."},{"key":"ref_53","doi-asserted-by":"crossref","first-page":"29","DOI":"10.1148\/radiology.143.1.7063747","article-title":"The meaning and use of the area under a receiver operating characteristic (ROC) curve","volume":"143","author":"Hanley","year":"1982","journal-title":"Radiology"},{"key":"ref_54","doi-asserted-by":"crossref","first-page":"1085","DOI":"10.1016\/S0140-6736(95)91748-9","article-title":"Comparing methods of measurement: Why plotting difference against standard method is misleading","volume":"346","author":"Bland","year":"1995","journal-title":"Lancet"},{"key":"ref_55","unstructured":"Centers for Disease Control and Prevention (2014). Non-Contact Temperature Measurement Devices: Considerations for Use in Port of Entry Screening Activities, Centers for Disease Control and Prevention."},{"key":"ref_56","doi-asserted-by":"crossref","first-page":"93","DOI":"10.1016\/j.medengphy.2021.05.007","article-title":"Performance characterization of non-contact infrared thermometers (NCITs) for forehead temperature measurement","volume":"93","author":"Sullivan","year":"2021","journal-title":"Med. Eng. Phys."},{"key":"ref_57","doi-asserted-by":"crossref","first-page":"65","DOI":"10.1080\/03091902.2017.1409818","article-title":"Comparison of non-contact infrared skin thermometers","volume":"42","author":"Fletcher","year":"2018","journal-title":"J. Med. Eng. Technol."},{"key":"ref_58","unstructured":"(2011). ASTM. ASTM E1112-00: Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature, ASTM Committee F04 on Medical and Surgical Materials and Devices."},{"key":"ref_59","doi-asserted-by":"crossref","first-page":"141","DOI":"10.11613\/BM.2015.015","article-title":"Understanding bland altman analysis","volume":"25","author":"Giavarina","year":"2015","journal-title":"Biochem. Med. Biochem. Med."},{"key":"ref_60","doi-asserted-by":"crossref","first-page":"778","DOI":"10.1002\/sim.3086","article-title":"Why Bland\u2013Altman plots should use X, not (Y + X)\/2 when X is a reference method","volume":"27","author":"Krouwer","year":"2008","journal-title":"Stat. Med."},{"key":"ref_61","doi-asserted-by":"crossref","first-page":"12305","DOI":"10.3390\/s140712305","article-title":"Infrared thermography for temperature measurement and non-destructive testing","volume":"14","author":"Usamentiaga","year":"2014","journal-title":"Sensors"},{"key":"ref_62","unstructured":"FDA (2021, October 29). Enforcement Policy for Telethermographic Systems During the Coronavirus Disease 2019 (COVID-19) Public Health Emergency, Available online: https:\/\/www.fda.gov\/media\/137079\/download."},{"key":"ref_63","first-page":"31","article-title":"Detection of body temperature with infrared thermography: Accuracy in detection of fever","volume":"18","author":"Cheung","year":"2012","journal-title":"Hong Kong Med. J."},{"key":"ref_64","doi-asserted-by":"crossref","first-page":"212","DOI":"10.1080\/03091900600711530","article-title":"Traceability and calibration in temperature measurement: A clinical necessity","volume":"30","author":"Simpson","year":"2006","journal-title":"J. Med. Eng. Technol."},{"key":"ref_65","doi-asserted-by":"crossref","first-page":"329","DOI":"10.1093\/sleep\/2.3.329","article-title":"Timing of REM sleep is coupled to the circadian rhythm of body temperature in man","volume":"2","author":"Charles","year":"1980","journal-title":"Sleep"},{"key":"ref_66","doi-asserted-by":"crossref","unstructured":"Harding, C., Pompei, F., Bordonaro, S.F., McGillicuddy, D.C., Burmistrov, D., and Sanchez, L.D. (2020). Fevers Are Rarest in the Morning: Could We Be Missing Infectious Disease Cases by Screening for Fever Then?. medRxiv.","DOI":"10.1101\/2020.05.23.20093484"},{"key":"ref_67","doi-asserted-by":"crossref","first-page":"69","DOI":"10.4137\/CMAMD.S5916","article-title":"Patellar skin surface temperature by thermography reflects knee osteoarthritis severity","volume":"3","author":"Denoble","year":"2010","journal-title":"Clin. Med. Insights. Arthritis Musculoskelet. Disord."},{"key":"ref_68","first-page":"e13474","article-title":"Diagnostic accuracy of infrared thermal imaging for detecting covid-19 infection in minimally symptomatic patients","volume":"51","year":"2020","journal-title":"Eur. J. Clin. Investig."}],"container-title":["Sensors"],"original-title":[],"language":"en","link":[{"URL":"https:\/\/www.mdpi.com\/1424-8220\/22\/1\/215\/pdf","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2025,10,11]],"date-time":"2025-10-11T07:55:19Z","timestamp":1760169319000},"score":1,"resource":{"primary":{"URL":"https:\/\/www.mdpi.com\/1424-8220\/22\/1\/215"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[2021,12,29]]},"references-count":68,"journal-issue":{"issue":"1","published-online":{"date-parts":[[2022,1]]}},"alternative-id":["s22010215"],"URL":"https:\/\/doi.org\/10.3390\/s22010215","relation":{},"ISSN":["1424-8220"],"issn-type":[{"value":"1424-8220","type":"electronic"}],"subject":[],"published":{"date-parts":[[2021,12,29]]}}}