{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2026,8,13]],"date-time":"2026-08-13T13:33:28Z","timestamp":1786628008045,"version":"build-2736575974"},"reference-count":23,"publisher":"MDPI AG","issue":"16","license":[{"start":{"date-parts":[[2026,8,13]],"date-time":"2026-08-13T00:00:00Z","timestamp":1786579200000},"content-version":"vor","delay-in-days":0,"URL":"https:\/\/creativecommons.org\/licenses\/by\/4.0\/"}],"content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":["Diagnostics"],"abstract":"<jats:p>Background\/Objectives: Corneal power has the largest impact on the variability of intraocular lens (IOL) power predictions, and corneal data from different biometers cannot be used interchangeably. We quantified the systematic differences between the Zeiss IOLMaster 700 (IOLM) and the Heidelberg Engineering Anterion and derived strategies for using both devices interchangeably in IOL power calculation. Methods: In this retrospective single-centre study, 837 eyes of 837 cataract patients were measured preoperatively with both biometers. Harmonic mean corneal front and back surface radii were derived from the flat and steep meridians, and corneal power referenced to the front apex plane was expressed as spherocylindrical power vectors (spherical equivalent, SEQ; astigmatic components C0 and C45). Three mapping strategies were compared using Bland\u2013Altman and double-angle plots: linear regression of corneal radii without (MR) and with (MRI) intercept and multivariate linear regression of the power vector components (MMV). Results: Corneal front surface radii agreed well between devices (MR slope 1.000), whereas the IOLM reported systematically flatter posterior radii (MR slope 0.943), giving a systematically higher total corneal power (43.105 D versus 42.760 D). MRI mapping largely removed the systematic offset in the corneal radii but did not fully correct the astigmatic centroids, whereas MMV mapping aligned both the SEQ and the astigmatic centroids at the origin and yielded smaller confidence ellipses. Conclusions: Measurements from the two devices are not directly interchangeable, primarily because of systematic discrepancies in the reported posterior corneal curvature. Where identical IOL calculation concepts and formula constants are used across devices, conversion of corneal data is mandatory, and multivariate power vector mapping provides superior harmonisation compared with radius-based approaches.<\/jats:p>","DOI":"10.3390\/diagnostics16162556","type":"journal-article","created":{"date-parts":[[2026,8,13]],"date-time":"2026-08-13T12:47:11Z","timestamp":1786625231000},"page":"2556","update-policy":"https:\/\/doi.org\/10.3390\/mdpi_crossmark_policy","source":"Crossref","is-referenced-by-count":0,"title":["Systematic Differences in Corneal Curvature and Power Measurements Between the IOLMaster 700 and the Anterion and Mapping Strategies for Cross-Device Use"],"prefix":"10.3390","volume":"16","author":[{"ORCID":"https:\/\/orcid.org\/0000-0001-9175-6177","authenticated-orcid":false,"given":"Achim","family":"Langenbucher","sequence":"first","affiliation":[{"name":"Department of Experimental Ophthalmology, Saarland University, 66424 Homburg, Germany"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"ORCID":"https:\/\/orcid.org\/0000-0003-4145-2559","authenticated-orcid":false,"given":"Jascha Armin","family":"Wendelstein","sequence":"additional","affiliation":[{"name":"Department of Experimental Ophthalmology, Saarland University, 66424 Homburg, Germany"},{"name":"Department of Ophthalmology, Ludwig-Maximilian University, 80336 Munich, Germany"},{"name":"IROC AG, Institut f\u00fcr Refraktive und Ophthalmo-Chirurgie, 8006 Zurich, Switzerland"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"ORCID":"https:\/\/orcid.org\/0000-0003-0779-7449","authenticated-orcid":false,"given":"Alan","family":"Cayless","sequence":"additional","affiliation":[{"name":"School of Physical Sciences, The Open University, Milton Keynes MK7 6AA, UK"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Peter","family":"Hoffmann","sequence":"additional","affiliation":[{"name":"Augen- und Laserklinik Castrop-Rauxel, 44575 Castrop-Rauxel, Germany"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"N\u00f3ra","family":"Szentm\u00e1ry","sequence":"additional","affiliation":[{"name":"Department of Experimental Ophthalmology, Saarland University, 66424 Homburg, Germany"},{"name":"Department of Ophthalmology, Semmelweis University, M\u00e1ria u. 39, 1085 Budapest, Hungary"}],"role":[{"vocabulary":"crossref","role":"author"}]}],"member":"1968","published-online":{"date-parts":[[2026,8,13]]},"reference":[{"key":"ref_1","doi-asserted-by":"crossref","first-page":"7","DOI":"10.1097\/j.jcrs.0000000000001300","article-title":"Performance of IOL calculation formulas that use measured posterior corneal power in eyes following myopic laser vision correction","volume":"50","author":"Kenny","year":"2024","journal-title":"J. Cataract Refract. Surg."},{"key":"ref_2","doi-asserted-by":"crossref","first-page":"360","DOI":"10.1097\/j.jcrs.0000000000001363","article-title":"Evaluation of corneal power from an AS-OCT thick lens model and ray tracing: Reliability of the keratometer index","volume":"50","author":"Langenbucher","year":"2024","journal-title":"J. Cataract Refract. Surg."},{"key":"ref_3","unstructured":"Langenbucher, A., Szentm\u00e1ry, N., Cayless, A., Hoffmann, P., and Wendelstein, J. (2025). Corneal power modelling with OCT data\u2014Thin and thick lens paraxial models versus raytracing. Acta Ophthalmol., in press."},{"key":"ref_4","doi-asserted-by":"crossref","first-page":"185","DOI":"10.1111\/opo.12909","article-title":"Prediction of corneal back surface power\u2014Deep learning algorithm versus multivariate regression","volume":"42","author":"Langenbucher","year":"2022","journal-title":"Ophthalmic Physiol. Opt."},{"key":"ref_5","doi-asserted-by":"crossref","first-page":"2080","DOI":"10.1016\/j.jcrs.2012.08.036","article-title":"Contribution of posterior corneal astigmatism to total corneal astigmatism","volume":"38","author":"Koch","year":"2012","journal-title":"J. Cataract Refract. Surg."},{"key":"ref_6","doi-asserted-by":"crossref","first-page":"1645","DOI":"10.1016\/j.jcrs.2014.01.046","article-title":"Influence of posterior corneal astigmatism on total corneal astigmatism in eyes with moderate to high astigmatism","volume":"40","author":"Savini","year":"2014","journal-title":"J. Cataract Refract. Surg."},{"key":"ref_7","doi-asserted-by":"crossref","first-page":"291","DOI":"10.1167\/iovs.14-15659","article-title":"Anterior surface-based keratometry compared with Scheimpflug tomography-based total corneal astigmatism","volume":"56","author":"Tonn","year":"2015","journal-title":"Invest. Ophthalmol. Vis. Sci."},{"key":"ref_8","doi-asserted-by":"crossref","unstructured":"Langenbucher, A., Wendelstein, J., Cayless, A., Hoffmann, P., and Szentm\u00e1ry, N. (2024). Vector analysis of corneal astigmatism in cataractous eyes based on IOLMaster 700 biometry. PLoS ONE, 19.","DOI":"10.1371\/journal.pone.0300576"},{"key":"ref_9","first-page":"1112","article-title":"Anterior segment optical coherence tomography in modern ophthalmic diagnostics","volume":"165","year":"2024","journal-title":"Orvosi Hetil."},{"key":"ref_10","doi-asserted-by":"crossref","first-page":"104186","DOI":"10.1016\/j.jfo.2024.104186","article-title":"Comparison of refractive outcomes obtained with two swept-source OCT-based optical biometers after cataract surgery: A study of 152 eyes","volume":"47","author":"Agard","year":"2024","journal-title":"J. Fr. Ophtalmol."},{"key":"ref_11","doi-asserted-by":"crossref","unstructured":"Chai, F., Ma, J.X., Wang, X., Liu, J., Jiang, L., Wu, W., and Li, J. (2024). Total keratometry for toric intraocular lens calculation: Comparison from two swept-source optical coherence tomography biometers. Front. Med., 10.","DOI":"10.3389\/fmed.2023.1276186"},{"key":"ref_12","doi-asserted-by":"crossref","first-page":"48","DOI":"10.3928\/1081597X-20221115-02","article-title":"Comparison of refractive prediction accuracy with three optical devices","volume":"39","author":"Cheng","year":"2023","journal-title":"J. Refract. Surg."},{"key":"ref_13","doi-asserted-by":"crossref","first-page":"2625","DOI":"10.1038\/s41433-024-03094-1","article-title":"Comparison of corneal topography maps obtained using the IOLMaster 700\u00ae and the Anterion\u00ae in candidates for toric IOL implantation","volume":"38","author":"Chudzinski","year":"2024","journal-title":"Eye"},{"key":"ref_14","doi-asserted-by":"crossref","first-page":"87","DOI":"10.1097\/j.jcrs.0000000000000373","article-title":"Comparison of 2 swept-source optical coherence tomography-based biometry devices","volume":"47","author":"Hirnschall","year":"2021","journal-title":"J. Cataract Refract. Surg."},{"key":"ref_15","doi-asserted-by":"crossref","first-page":"e1330","DOI":"10.3928\/1081597X-20251027-02","article-title":"Comparison of posterior corneal curvature measurements by four different tomographers","volume":"41","author":"Grendele","year":"2025","journal-title":"J. Refract. Surg."},{"key":"ref_16","first-page":"287","article-title":"Comparison of accuracy and axial length acquisition success rate by three types of swept-source OCT-based biometers","volume":"51","author":"Kato","year":"2025","journal-title":"J. Cataract Refract. Surg."},{"key":"ref_17","doi-asserted-by":"crossref","first-page":"999","DOI":"10.1007\/s00417-022-05870-9","article-title":"Comparison of 2 modern swept-source optical biometers\u2014IOLMaster 700 and Anterion","volume":"261","author":"Langenbucher","year":"2023","journal-title":"Graefes Arch. Clin. Exp. Ophthalmol."},{"key":"ref_18","doi-asserted-by":"crossref","first-page":"1594","DOI":"10.4103\/ijo.IJO_2433_21","article-title":"Comparison of ocular biometry and refractive outcome between Anterion and IOLMaster 700","volume":"70","author":"Panda","year":"2022","journal-title":"Indian J. Ophthalmol."},{"key":"ref_19","doi-asserted-by":"crossref","first-page":"23","DOI":"10.4103\/IJO.IJO_3122_24","article-title":"Comparison meta-analysis of two swept-source optical coherence biometers","volume":"74","author":"Romero","year":"2026","journal-title":"Indian J. Ophthalmol."},{"key":"ref_20","doi-asserted-by":"crossref","first-page":"1117","DOI":"10.1097\/j.jcrs.0000000000001515","article-title":"Comparison of keratometry and total corneal power, as measured by an SS-OCT-based optical biometer, for intraocular lens power calculation in Asian eyes","volume":"50","author":"Savini","year":"2024","journal-title":"J. Cataract Refract. Surg."},{"key":"ref_21","doi-asserted-by":"crossref","first-page":"1","DOI":"10.1186\/s40662-022-00320-3","article-title":"Comprehensive comparisons of ocular biometry: A network-based big data analysis","volume":"10","author":"Yu","year":"2023","journal-title":"Eye Vis."},{"key":"ref_22","doi-asserted-by":"crossref","first-page":"347","DOI":"10.1016\/S0161-6420(79)35504-X","article-title":"Applied optical principles: Keratometry","volume":"86","author":"Sampson","year":"1979","journal-title":"Ophthalmology"},{"key":"ref_23","doi-asserted-by":"crossref","first-page":"858","DOI":"10.1097\/j.jcrs.0000000000000890","article-title":"Total keratometric power vs total corneal power","volume":"48","author":"Holladay","year":"2022","journal-title":"J. Cataract Refract. Surg."}],"container-title":["Diagnostics"],"original-title":[],"language":"en","link":[{"URL":"https:\/\/www.mdpi.com\/2075-4418\/16\/16\/2556\/pdf","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2026,8,13]],"date-time":"2026-08-13T12:51:27Z","timestamp":1786625487000},"score":1,"resource":{"primary":{"URL":"https:\/\/www.mdpi.com\/2075-4418\/16\/16\/2556"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[2026,8,13]]},"references-count":23,"journal-issue":{"issue":"16","published-online":{"date-parts":[[2026,8]]}},"alternative-id":["diagnostics16162556"],"URL":"https:\/\/doi.org\/10.3390\/diagnostics16162556","relation":{},"ISSN":["2075-4418"],"issn-type":[{"value":"2075-4418","type":"electronic"}],"subject":[],"published":{"date-parts":[[2026,8,13]]}}}