{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2026,8,19]],"date-time":"2026-08-19T18:09:35Z","timestamp":1787162975206,"version":"build-2736575974"},"reference-count":25,"publisher":"MDPI AG","issue":"7","license":[{"start":{"date-parts":[[2022,6,21]],"date-time":"2022-06-21T00:00:00Z","timestamp":1655769600000},"content-version":"vor","delay-in-days":0,"URL":"https:\/\/creativecommons.org\/licenses\/by\/4.0\/"}],"content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":["J. Imaging"],"abstract":"<jats:p>Background: This study investigates the effects of vertical off-centring, localiser direction, tube voltage, and phantom positioning (supine and prone) on computed tomography (CT) numbers and radiation dose. Methods: An anthropomorphic phantom was scanned using a Discovery CT750 HD\u2014128 slice (GE Healthcare) scanner at different tube voltages (80, 120, and 140 kVp). Images employing 0\u00b0 and 180\u00b0 localisers were acquired in supine and prone positions for each vertical off-centring (\u00b1100, \u00b160, and \u00b130 mm from the iso-centre). CT numbers and displayed volume CT dose index (CTDIvol) were recorded. The relationship between dose variation and CT number was investigated. Results: The maximum changes in CT number between the two phantom positions as a function of vertical-off-centring were for the upper thorax 34 HU (0\u00b0 localiser, 120 kVp), mid thorax 43 HU (180\u00b0 localiser, 80 kVp), and for the abdominal section 31 HU (0\u00b0 localiser, 80 kVp) in the prone position. A strong positive correlation was reported between the variation in dose and CT number (r = 0.969, p &lt; 0.001); 95% CI (0.93, 0.99). Conclusions: Patient positioning demands an approach with a high degree of accuracy, especially in cases where clinical decisions depend on CT number accuracy for tissue lesion characterisation.<\/jats:p>","DOI":"10.3390\/jimaging8070175","type":"journal-article","created":{"date-parts":[[2022,6,21]],"date-time":"2022-06-21T21:55:19Z","timestamp":1655848519000},"page":"175","update-policy":"https:\/\/doi.org\/10.3390\/mdpi_crossmark_policy","source":"Crossref","is-referenced-by-count":4,"title":["The Impacts of Vertical Off-Centring, Localiser Direction, Phantom Positioning and Tube Voltage on CT Number Accuracy: An Experimental Study"],"prefix":"10.3390","volume":"8","author":[{"given":"Yazan","family":"Al-Hayek","sequence":"first","affiliation":[{"name":"School of Dentistry and Medical Sciences, Faculty of Science and Health, Charles Sturt University, Wagga Wagga, NSW 2650, Australia"},{"name":"Department of Medical Imaging, Faculty of Applied Health Sciences, The Hashemite University, Zarqa 13133, Jordan"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"ORCID":"https:\/\/orcid.org\/0000-0003-0285-9962","authenticated-orcid":false,"given":"Kelly","family":"Spuur","sequence":"additional","affiliation":[{"name":"School of Dentistry and Medical Sciences, Faculty of Science and Health, Charles Sturt University, Wagga Wagga, NSW 2650, Australia"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Rob","family":"Davidson","sequence":"additional","affiliation":[{"name":"School of Health Sciences, Faculty of Health, University of Canberra, Canberra, ACT 2601, Australia"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Christopher","family":"Hayre","sequence":"additional","affiliation":[{"name":"School of Dentistry and Medical Sciences, Faculty of Science and Health, Charles Sturt University, Wagga Wagga, NSW 2650, Australia"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"ORCID":"https:\/\/orcid.org\/0000-0003-0918-2624","authenticated-orcid":false,"given":"Xiaoming","family":"Zheng","sequence":"additional","affiliation":[{"name":"School of Dentistry and Medical Sciences, Faculty of Science and Health, Charles Sturt University, Wagga Wagga, NSW 2650, Australia"}],"role":[{"vocabulary":"crossref","role":"author"}]}],"member":"1968","published-online":{"date-parts":[[2022,6,21]]},"reference":[{"key":"ref_1","doi-asserted-by":"crossref","first-page":"966","DOI":"10.1088\/1361-6498\/ab1826","article-title":"NCRP Report no.180-management of exposure to ionizing radiation: NCRP radiation protection guidance for the United States","volume":"39","author":"Cool","year":"2019","journal-title":"J. Radiol. Prot."},{"key":"ref_2","doi-asserted-by":"crossref","first-page":"352","DOI":"10.1097\/RLI.0000000000000133","article-title":"Automated tube voltage selection in thoracoabdominal computed tomography at high pitch using a third-generation dual-source scanner: Image quality and radiation dose performance","volume":"50","author":"Lurz","year":"2015","journal-title":"Investig. Radiol."},{"key":"ref_3","doi-asserted-by":"crossref","first-page":"3618","DOI":"10.1007\/s00330-017-4745-8","article-title":"Efficacy of a dynamic collimator for overranging dose reduction in a second- and third-generation dual source CT scanner","volume":"27","author":"Booij","year":"2017","journal-title":"Eur. Radiol."},{"key":"ref_4","doi-asserted-by":"crossref","first-page":"261","DOI":"10.1007\/s00247-002-0678-7","article-title":"Dose reduction opportunities for CT scanners","volume":"32","author":"Toth","year":"2002","journal-title":"Pediatr. Radiol."},{"key":"ref_5","doi-asserted-by":"crossref","first-page":"96","DOI":"10.4274\/mirt.galenos.2019.83723","article-title":"Effects of the Use of Automatic Tube Current Modulation on Patient Dose and Image Quality in Computed Tomography","volume":"28","author":"Yurt","year":"2019","journal-title":"Mol. Imaging Radionucl. Ther."},{"key":"ref_6","unstructured":"Kalender, W. (2011). Computed Tomography: Fundamentals, System Technology, Image Quality, Applications, Publicis Publishing. [3rd ed.]."},{"key":"ref_7","doi-asserted-by":"crossref","first-page":"191","DOI":"10.1016\/j.ejmp.2011.06.002","article-title":"Impact of miscentering on patient dose and image noise in X-ray CT imaging: Phantom and clinical studies","volume":"28","author":"Habibzadeh","year":"2012","journal-title":"Phys. Med."},{"key":"ref_8","doi-asserted-by":"crossref","first-page":"1439","DOI":"10.1088\/0031-9155\/49\/8\/005","article-title":"Monte Carlo simulations in CT for the study of the surface air kerma and energy imparted to phantoms of varying size and position","volume":"49","author":"Dance","year":"2004","journal-title":"Phys. Med. Biol."},{"key":"ref_9","doi-asserted-by":"crossref","first-page":"34","DOI":"10.1067\/j.cpradiol.2018.12.010","article-title":"Effect of CT Localizer Radiographs on Radiation Dose Associated with Automatic Tube Current Modulation: A Multivendor Study","volume":"49","author":"Paolicchi","year":"2020","journal-title":"Curr. Probl. Diagn. Radiol."},{"key":"ref_10","doi-asserted-by":"crossref","first-page":"199","DOI":"10.1016\/0730-4862(83)90099-9","article-title":"Dependence of substance CT number on scanning technique and position within scanner","volume":"7","author":"Hunter","year":"1983","journal-title":"Comput. Radiol."},{"key":"ref_11","doi-asserted-by":"crossref","first-page":"547","DOI":"10.2214\/AJR.06.0370","article-title":"Automatic patient centering for MDCT: Effect on radiation dose","volume":"188","author":"Li","year":"2007","journal-title":"AJR Am. J. Roentgenol."},{"key":"ref_12","doi-asserted-by":"crossref","first-page":"4417","DOI":"10.1120\/jacmp.v14i6.4417","article-title":"Scanner and kVp dependence of measured CT numbers in the ACR CT phantom","volume":"14","author":"Cropp","year":"2013","journal-title":"J. Appl. Clin. Med. Phys."},{"key":"ref_13","doi-asserted-by":"crossref","first-page":"1064","DOI":"10.2214\/AJR.16.17215","article-title":"Variation in CT Number and Image Noise Uniformity According to Patient Positioning in MDCT","volume":"208","author":"Szczykutowicz","year":"2017","journal-title":"AJR Am. J. Roentgenol."},{"key":"ref_14","doi-asserted-by":"crossref","first-page":"403","DOI":"10.1148\/radiol.2352040515","article-title":"Measuring coronary calcium on CT images adjusted for attenuation differences","volume":"235","author":"Nelson","year":"2005","journal-title":"Radiology"},{"key":"ref_15","doi-asserted-by":"crossref","first-page":"1","DOI":"10.1186\/s13244-019-0688-8","article-title":"Imaging features of adrenal masses","volume":"10","author":"Albano","year":"2019","journal-title":"Insights Imaging"},{"key":"ref_16","doi-asserted-by":"crossref","first-page":"S101","DOI":"10.1148\/rg.24si045515","article-title":"From the RSNA refresher courses: A practical approach to the cystic renal mass","volume":"24","author":"Hartman","year":"2004","journal-title":"Radiographics"},{"key":"ref_17","doi-asserted-by":"crossref","first-page":"563","DOI":"10.1148\/radiol.2015151169","article-title":"Radiomics: Images Are More than Pictures, They Are Data","volume":"278","author":"Gillies","year":"2016","journal-title":"Radiology"},{"key":"ref_18","doi-asserted-by":"crossref","first-page":"8262","DOI":"10.1038\/s41598-021-87598-w","article-title":"Robustness of radiomic features in CT images with different slice thickness, comparing liver tumour and muscle","volume":"11","author":"Rundo","year":"2021","journal-title":"Sci. Rep."},{"key":"ref_19","doi-asserted-by":"crossref","first-page":"10545","DOI":"10.1038\/s41598-018-28895-9","article-title":"Voxel size and gray level normalization of CT radiomic features in lung cancer","volume":"8","author":"Latifi","year":"2018","journal-title":"Sci. Rep."},{"key":"ref_20","first-page":"5","article-title":"0\u00b0 vs. 180\u00b0 CT localiser: The effect of vertical off-centring, phantom positioning and tube voltage on dose optimisation in multidetector computed tomography","volume":"69","author":"Zheng","year":"2021","journal-title":"J. Med. Radiat. Sci."},{"key":"ref_21","unstructured":"(2022, April 18). ImageJ Software, Available online: https:\/\/imagej.nih.gove\/ij\/."},{"key":"ref_22","doi-asserted-by":"crossref","unstructured":"Hsieh, J. (2015). Computed Tomography: Principles, Design, Artifacts, and Recent Advances, SPIE. [3rd ed.].","DOI":"10.1117\/3.2197756"},{"key":"ref_23","first-page":"1","article-title":"The effects of CT X-ray tube voltage and current variations on the relative electron density (RED) and CT number conversion curves","volume":"13","author":"Afifi","year":"2020","journal-title":"J. Radiat. Res. Appl. Sci."},{"key":"ref_24","doi-asserted-by":"crossref","first-page":"578","DOI":"10.1093\/jrr\/rrs109","article-title":"Determining photon energy absorption parameters for different soil samples","volume":"54","author":"Kucuk","year":"2013","journal-title":"J. Radiat. Res."},{"key":"ref_25","doi-asserted-by":"crossref","first-page":"72","DOI":"10.1186\/s13014-019-1276-2","article-title":"Changes of lung parenchyma density following high dose radiation therapy for thoracic carcinomas\u2014An automated analysis of follow up CT scans","volume":"14","author":"Kirschner","year":"2019","journal-title":"Radiat. Oncol."}],"container-title":["Journal of Imaging"],"original-title":[],"language":"en","link":[{"URL":"https:\/\/www.mdpi.com\/2313-433X\/8\/7\/175\/pdf","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2025,10,10]],"date-time":"2025-10-10T23:36:38Z","timestamp":1760139398000},"score":1,"resource":{"primary":{"URL":"https:\/\/www.mdpi.com\/2313-433X\/8\/7\/175"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[2022,6,21]]},"references-count":25,"journal-issue":{"issue":"7","published-online":{"date-parts":[[2022,7]]}},"alternative-id":["jimaging8070175"],"URL":"https:\/\/doi.org\/10.3390\/jimaging8070175","relation":{},"ISSN":["2313-433X"],"issn-type":[{"value":"2313-433X","type":"electronic"}],"subject":[],"published":{"date-parts":[[2022,6,21]]}}}