{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2024,6,2]],"date-time":"2024-06-02T11:40:03Z","timestamp":1717328403203},"reference-count":19,"publisher":"SLACK, Inc.","issue":"9","content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":["J Refract Surg"],"published-print":{"date-parts":[[2014,8]]},"abstract":"<jats:sec><jats:title>PURPOSE:<\/jats:title><jats:p>To evaluate LASIK corneal flaps using a multifunctional femtosecond laser suitable for cataract and corneal surgery (LenSx; Alcon Laboratories, Inc., Aliso Viejo, CA) and to compare the planned and postoperatively measured flap thickness using an anterior segment optical coherence tomography device (AS-OCT).<\/jats:p><\/jats:sec><jats:sec><jats:title>METHODS:<\/jats:title><jats:p>Twenty patients (38 eyes) diagnosed as having myopia and myopic astigmatism were enrolled. LASIK was performed using the LenSx femtosecond laser for intracorneal flaps and the Wavelight Allegretto 400 excimer laser (Alcon Laboratories, Inc.) for intrastromal photoablation. Desired flap thickness and diameter were 140.0 \u00b1 0.0<jats:italic>\u00b5<\/jats:italic>m and 8.5 \u00b1 0.0 mm, respectively, whereas mean ablation depth and diameter of the excimer laser treatment were 67.9 \u00b1 24.18<jats:italic>\u00b5<\/jats:italic>m and 6.5 \u00b1 0.08 mm, respectively. Entered data of the LenSx femtosecond laser were used to determine desired flap thickness, whereas AS-OCT (RTVue; Optovue, Inc., Fremont, CA) was used to measure flap thickness postoperatively. The Wilcoxon signed-rank test, dependent paired<jats:italic>t<\/jats:italic>test, and Friedman test were used for comparison of dependent and repeated measures.<\/jats:p><\/jats:sec><jats:sec><jats:title>RESULTS:<\/jats:title><jats:p>There was no statistically significant difference in the planned and postoperatively measured flap thickness (140.0 \u00b1 0.0 vs 140.28 \u00b1 8.0<jats:italic>\u00b5<\/jats:italic>m;<jats:italic>P<\/jats:italic>= .4067). Interfaces of the flaps had even surfaces according to the images and calculations on the AS-OCT device (<jats:italic>P<\/jats:italic>= .058).<\/jats:p><\/jats:sec><jats:sec><jats:title>CONCLUSIONS:<\/jats:title><jats:p>Application of this multifunctional femtosecond laser performing LASIK proved to be a safe and effective method regarding predictability of flap thickness.<\/jats:p><jats:p><jats:bold>[<jats:italic>J Refract Surg.<\/jats:italic>2014;30(9):622\u2013626.]<\/jats:bold><\/jats:p><\/jats:sec>","DOI":"10.3928\/1081597x-20140827-01","type":"journal-article","created":{"date-parts":[[2014,9,10]],"date-time":"2014-09-10T15:00:03Z","timestamp":1410361203000},"page":"622-626","source":"Crossref","is-referenced-by-count":10,"title":["Analysis of Planned and Postoperatively Measured Flap Thickness After LASIK Using the LenSx Multifunctional Femtosecond Laser System"],"prefix":"10.3928","volume":"30","author":[{"given":"Eva","family":"Juhasz","sequence":"first","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Tamas","family":"Filkorn","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Kinga","family":"Kranitz","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Gabor Laszlo","family":"Sandor","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Andrea","family":"Gyenes","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Zoltan Zsolt","family":"Nagy","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]}],"member":"2308","reference":[{"key":"e_1_3_2_2_2","doi-asserted-by":"publisher","DOI":"10.1126\/science.1957169"},{"key":"e_1_3_2_3_2","doi-asserted-by":"publisher","DOI":"10.3928\/1081-597X-20070601-03"},{"key":"e_1_3_2_4_2","doi-asserted-by":"publisher","DOI":"10.3928\/1081597X-20110812-03"},{"key":"e_1_3_2_5_2","first-page":"350","article-title":"Comparison of visual acuity, refractive results and complications of femtosecond laser with mechanical microkeratome in LASIK","volume":"6","author":"Cosar CB","year":"2013","unstructured":"Cosar CB, Gonen T, Moray M, Sener AB. Comparison of visual acuity, refractive results and complications of femtosecond laser with mechanical microkeratome in LASIK. Int J Ophthalmol. 2013;6:350\u2013355.","journal-title":"Int J Ophthalmol"},{"key":"e_1_3_2_6_2","doi-asserted-by":"crossref","first-page":"2440","DOI":"10.3760\/cma.j.issn.0366-6999.20130918","article-title":"Comparison of the flaps made by femtosecond laser and automated keratomes for sub-bowman keratomileusis","volume":"126","author":"Zhai CB","year":"2013","unstructured":"Zhai CB, Tian L, Zhou YH, Zhang QW, Zhang J. Comparison of the flaps made by femtosecond laser and automated keratomes for sub-bowman keratomileusis. Chin Med J (Engl). 2013;126:2440\u20132444.","journal-title":"Chin Med J (Engl)"},{"key":"e_1_3_2_7_2","first-page":"347","article-title":"Ultra-short pulse (femtosecond) laser surgery: initial use in LASIK flap creation","volume":"14","author":"Ratkay-Traub I","year":"2001","unstructured":"Ratkay-Traub I, Juhasz T, Horvath C, et al. Ultra-short pulse (femtosecond) laser surgery: initial use in LASIK flap creation. Ophthalmol Clin North Am. 2001;14:347\u2013355.","journal-title":"Ophthalmol Clin North Am"},{"key":"e_1_3_2_8_2","doi-asserted-by":"publisher","DOI":"10.1097\/ICU.0b013e3283477c9c"},{"key":"e_1_3_2_9_2","doi-asserted-by":"publisher","DOI":"10.3928\/1081597X-20091117-04"},{"key":"e_1_3_2_10_2","doi-asserted-by":"publisher","DOI":"10.1126\/scitranslmed.3001305"},{"key":"e_1_3_2_11_2","doi-asserted-by":"publisher","DOI":"10.1167\/iovs.06-0192"},{"key":"e_1_3_2_12_2","doi-asserted-by":"publisher","DOI":"10.3928\/1081-597X-20070601-03"},{"key":"e_1_3_2_13_2","volume-title":"Presented at the ASCRS Annual Meeting","author":"Knorz M","unstructured":"Knorz M. Analysis of 100 consecutive capsulotomies created with femtosecond laser for cataract surgery and novel patient interface technology. Presented at the ASCRS Annual Meeting; April 19\u201323, 2013; San Francisco, CA."},{"key":"e_1_3_2_14_2","doi-asserted-by":"publisher","DOI":"10.1136\/bjophthalmol-2011-300073"},{"key":"e_1_3_2_15_2","doi-asserted-by":"publisher","DOI":"10.3928\/1081597X-20130617-04"},{"key":"e_1_3_2_16_2","doi-asserted-by":"publisher","DOI":"10.1016\/S0161-6420(01)00966-6"},{"key":"e_1_3_2_17_2","doi-asserted-by":"publisher","DOI":"10.1016\/S0002-9394(03)00222-8"},{"key":"e_1_3_2_18_2","first-page":"69","article-title":"Corneal flap morphological analysis using anterior segment optical coherence tomography in laser in situ keratomileusis with femtosecond lasers versus mechanical microkeratome","volume":"5","author":"Zhang XX","year":"2012","unstructured":"Zhang XX, Zhong XW, Wu JS, et al. Corneal flap morphological analysis using anterior segment optical coherence tomography in laser in situ keratomileusis with femtosecond lasers versus mechanical microkeratome. Int J Ophthalmol. 2012;5:69\u201373.","journal-title":"Int J Ophthalmol"},{"key":"e_1_3_2_19_2","doi-asserted-by":"publisher","DOI":"10.1016\/j.ophtha.2011.12.025"},{"key":"e_1_3_2_20_2","doi-asserted-by":"publisher","DOI":"10.1055\/s-0032-1328384"}],"container-title":["Journal of Refractive Surgery"],"original-title":[],"language":"en","link":[{"URL":"https:\/\/www.healio.com\/doiresolver?doi=10.3928\/1081597X-20140827-01","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2024,6,2]],"date-time":"2024-06-02T11:09:41Z","timestamp":1717326581000},"score":1,"resource":{"primary":{"URL":"https:\/\/journals.healio.com\/doi\/10.3928\/1081597X-20140827-01"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[2014,8]]},"references-count":19,"journal-issue":{"issue":"9","published-print":{"date-parts":[[2014,8]]}},"alternative-id":["10.3928\/1081597X-20140827-01"],"URL":"https:\/\/doi.org\/10.3928\/1081597x-20140827-01","relation":{},"ISSN":["1081-597X"],"issn-type":[{"value":"1081-597X","type":"print"}],"subject":[],"published":{"date-parts":[[2014,8]]}}}