{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2026,4,6]],"date-time":"2026-04-06T10:01:31Z","timestamp":1775469691678,"version":"3.50.1"},"reference-count":9,"publisher":"SLACK, Inc.","issue":"7","content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":["J Refract Surg"],"published-print":{"date-parts":[[2017,7]]},"abstract":"<jats:sec>\n            <jats:title>PURPOSE:<\/jats:title>\n            <jats:p>To compare the alignment meridian generated by the Zeiss Callisto Eye (Carl Zeiss AG, Dublin, CA) and the Alcon Verion Image Guided System (Alcon Laboratories, Inc., Fort Worth, TX).<\/jats:p>\n          <\/jats:sec>\n          <jats:sec>\n            <jats:title>METHODS:<\/jats:title>\n            <jats:p>In this retrospective comparative evaluation of technology, intraoperative images were captured at different steps in the same surgery, allowing the comparison of the guidance lines generated by the Verion system to the parallel guidance lines generated by the Callisto Eye system. Measurements of each hemi-meridian were quantified using Adobe Photoshop 2015 CC software (Adobe Systems, San Jose, CA). The numbers of degrees separating these alignment meridians were calculated, entered into a database, and analyzed.<\/jats:p>\n          <\/jats:sec>\n          <jats:sec>\n            <jats:title>RESULTS:<\/jats:title>\n            <jats:p>\n              The authors found that of 98 captured images of 16 eyes, the two technologies were identical in 0 eyes (\u03b8\n              <jats:sub>1<\/jats:sub>\n              = \u03b8\n              <jats:sub>2<\/jats:sub>\n              = 0), similar by 3\u00b0 in 52 (53%) captured images (\u03b8\n              <jats:sub>1<\/jats:sub>\n              \u2260 \u03b8\n              <jats:sub>2<\/jats:sub>\n              \u2260 0), and different by at least 3\u00b0 in 46 (47%) captured images (\u03b8\n              <jats:sub>1<\/jats:sub>\n              \u2260 \u03b8\n              <jats:sub>2<\/jats:sub>\n              \u2260 0). The target meridians were superimposed, the target lines were minimally separated, and the target lines were dissimilar. It was noted that some intraoperative variation occurred from measurement to measurement. Within the small group of 16 cases of routine toric lens implantation in this study, the absolute average number of degrees of misalignment between the Verion and Callisto Eye systems was 3.355 for \u03b8\n              <jats:sub>1<\/jats:sub>\n              and 3.838 for \u03b8\n              <jats:sub>2<\/jats:sub>\n              . On average, the intraoperative variation termed \u201cdrift\u201d was noted to be 3.963\u00b0 for \u03b8\n              <jats:sub>1<\/jats:sub>\n              , and 4.557\u00b0 for \u03b8\n              <jats:sub>2<\/jats:sub>\n              .\n            <\/jats:p>\n          <\/jats:sec>\n          <jats:sec>\n            <jats:title>CONCLUSIONS:<\/jats:title>\n            <jats:p>The authors found that small deviations were frequent when comparing two sophisticated technologies. Although deviations greater than 3\u00b0 occurred in less than 47% of captured images from 16 eyes, smaller but significant variations of less than 3\u00b0 occurred in 53% of captured images from 16 eyes. It was rare to identify a large deviation. However, the authors identified \u201cdrift\u201d in the same eye when measurements were taken at different times. The results indicate that the two systems are not currently interchangeable. Superiority of one system over the other was not determined.<\/jats:p>\n            <jats:p>\n              <jats:bold>\n                [\n                <jats:italic>J Refract Surg.<\/jats:italic>\n                2017;33(7):482\u2013487.]\n              <\/jats:bold>\n            <\/jats:p>\n          <\/jats:sec>","DOI":"10.3928\/1081597x-20170504-02","type":"journal-article","created":{"date-parts":[[2017,7,6]],"date-time":"2017-07-06T12:05:54Z","timestamp":1499342754000},"page":"482-487","source":"Crossref","is-referenced-by-count":36,"title":["Comparing the Zeiss Callisto Eye and the Alcon Verion Image Guided System Toric Lens Alignment Technologies"],"prefix":"10.3928","volume":"33","author":[{"given":"Arjan S.","family":"Hura","sequence":"first","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]},{"given":"Robert H.","family":"Osher","sequence":"additional","affiliation":[],"role":[{"role":"author","vocabulary":"crossref"}]}],"member":"2308","reference":[{"key":"e_1_3_2_2_2","volume-title":"Presented at: American Intraocular Implant Society Annual Meeting","author":"Osher RH","unstructured":"Osher RH. Combining phacoemulsification with corneal relaxing incisions for reduction of preexisting astigmatism. Presented at: American Intraocular Implant Society Annual Meeting; 1984; Los Angeles, CA."},{"key":"e_1_3_2_3_2","doi-asserted-by":"publisher","DOI":"10.1016\/S0886-3350(13)80232-5"},{"key":"e_1_3_2_4_2","doi-asserted-by":"publisher","DOI":"10.1016\/j.jcrs.2013.02.020"},{"key":"e_1_3_2_5_2","unstructured":"Beaver-Visitec International Launches New Toric Marker. Beaver-Visitec International Launches New Toric Marker 2013. Available at: http:\/\/beaver-visitec.com\/news-events\/press-release.cfm?customel_datapageid_2208=2815."},{"key":"e_1_3_2_6_2","doi-asserted-by":"publisher","DOI":"10.1016\/j.jcrs.2009.09.021"},{"key":"e_1_3_2_7_2","volume-title":"Presented at: XXIX Congress of the European Society of Cataract and Refractive Surgeons","author":"Nuijts R","unstructured":"Nuijts R. Optimal cataract workflow with Haag-Streit Lenstar and SMI surgery guidance. Presented at: XXIX Congress of the European Society of Cataract and Refractive Surgeons; September 17\u201321, 2011; Vienna."},{"key":"e_1_3_2_8_2","unstructured":"Trinh L Villette V Dupont-Monod S Auclin F Baudouin C. Marker-Based Versus Markerless Toric IOL Alignment: Demonstrated Benefit Of An Automated Markerless System . White Paper. 2015. Available at: https:\/\/applications.zeiss.com\/C1257A290053AE30\/0\/CBDEB9B1E125C397C1257DE30052352C\/$FILE\/WhitePaper_MARKERLESS_TORIC_IOL_dvk_03.pdf"},{"key":"e_1_3_2_9_2","volume-title":"Conference in Sarasota","author":"Eye Photo Systems Launches New Toric Alignment System by Osher Software for Slit Lamp Photography","year":"2015","unstructured":"Eye Photo Systems Launches New Toric Alignment System by Osher Software for Slit Lamp Photography. 2015 Cataract Surgery: Telling It Like It Is! Conference in Sarasota, Florida; January 15, 2015."},{"key":"e_1_3_2_10_2","doi-asserted-by":"publisher","DOI":"10.1016\/j.jcrs.2012.08.019"}],"container-title":["Journal of Refractive Surgery"],"original-title":[],"language":"en","link":[{"URL":"https:\/\/www.healio.com\/doiresolver?doi=10.3928\/1081597X-20170504-02","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2023,5,3]],"date-time":"2023-05-03T09:27:14Z","timestamp":1683106034000},"score":1,"resource":{"primary":{"URL":"https:\/\/journals.healio.com\/doi\/10.3928\/1081597X-20170504-02"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[2017,7]]},"references-count":9,"journal-issue":{"issue":"7","published-print":{"date-parts":[[2017,7]]}},"alternative-id":["10.3928\/1081597X-20170504-02"],"URL":"https:\/\/doi.org\/10.3928\/1081597x-20170504-02","relation":{},"ISSN":["1081-597X"],"issn-type":[{"value":"1081-597X","type":"print"}],"subject":[],"published":{"date-parts":[[2017,7]]}}}