{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2026,5,29]],"date-time":"2026-05-29T09:34:17Z","timestamp":1780047257231,"version":"3.53.1"},"reference-count":36,"publisher":"SLACK, Inc.","issue":"2","content-domain":{"domain":[],"crossmark-restriction":false},"short-container-title":["J Refract Surg"],"published-print":{"date-parts":[[2024,2]]},"abstract":"<jats:sec>\n            <jats:title>Purpose:<\/jats:title>\n            <jats:p>To determine the prevalence of opacification of a hydrophylic intraocular lens (IOL) with hydrophobic coverage and the percentage of explantations required, determining possible risk factors associated with these opacifications.<\/jats:p>\n          <\/jats:sec>\n          <jats:sec>\n            <jats:title>Methods:<\/jats:title>\n            <jats:p>This ambispective study enrolled 575 eyes of 296 patients (age 36 to 87 years) that underwent cataract surgery between 2010 and 2017 with implantation of different models of Lentis Mplus IOLs: LS-312-MF30, LS-313-MF15, LS-313-MF30, LU-313-MF30, and LU-313-MF30T (Oculentis GmbH). Visual, refractive, and slit-lamp biomicroscopic changes were evaluated in a long-term follow-up. The percentage of cases with IOL opacification and the percentage of cases in which IOL explantation was required due to such opacifications were calculated at each visit. Five of the explanted IOLs from this series could be analyzed by scanning electron microscopy (SEM-EDX).<\/jats:p>\n          <\/jats:sec>\n          <jats:sec>\n            <jats:title>Results:<\/jats:title>\n            <jats:p>\n              IOL opacification developed in 63 eyes (11.0%). With 95% confidence, the prevalence of IOL opacification was between 842 and 1,401 cases per 10,000 eyes. The time elapsed between surgery and the presence of IOL opacification ranged between 0 and 9.3 years (mean: 4.7 \u00b1 2.2 years). No significant differences in terms of IOL opacification rate were found according to gender (\n              <jats:italic>P<\/jats:italic>\n              = .378). No significant differences were found in arterial hypertension, diabetes, hypothyroidism, or hyperthyroidism rates between eyes with or without IOL opacification (\n              <jats:italic>P<\/jats:italic>\n              \u2265 .053). IOL explantation was needed in 9 eyes (1.57%). SEM-EDX analysis confirmed the presence of rough areas on the IOL surface containing different components, such as calcium, phosphorous, copper, or nitrogen.\n            <\/jats:p>\n          <\/jats:sec>\n          <jats:sec>\n            <jats:title>Conclusions:<\/jats:title>\n            <jats:p>The prevalence of opacification with time in Lentis Mplus IOLs is high, with no systemic risk factors associated with this complication, suggesting that it may be attributable to the material and\/or the manufacturing process.<\/jats:p>\n            <jats:p>\n              <jats:bold>\n                [\n                <jats:italic>J Refract Surg<\/jats:italic>\n                . 2024;40(2):e98\u2013e107.]\n              <\/jats:bold>\n            <\/jats:p>\n          <\/jats:sec>","DOI":"10.3928\/1081597x-20240115-01","type":"journal-article","created":{"date-parts":[[2024,2,12]],"date-time":"2024-02-12T19:03:03Z","timestamp":1707764583000},"source":"Crossref","is-referenced-by-count":4,"title":["Long-term Prevalence of Opacification of a Hydrophylic Acrylic Rotationally Asymmetric Refractive Multifocal Intraocular Lens"],"prefix":"10.3928","volume":"40","author":[{"given":"Mar\u00eda Teresa","family":"\u00c1lvarez-Garc\u00eda","sequence":"first","affiliation":[],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Esther","family":"Rivera-Ruiz","sequence":"additional","affiliation":[],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Jorge L.","family":"Ali\u00f3","sequence":"additional","affiliation":[],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"David P.","family":"Pi\u00f1ero","sequence":"additional","affiliation":[],"role":[{"vocabulary":"crossref","role":"author"}]}],"member":"2308","reference":[{"key":"e_1_3_2_2_2","doi-asserted-by":"publisher","DOI":"10.3109\/02713683.2014.978476"},{"key":"e_1_3_2_3_2","doi-asserted-by":"publisher","DOI":"10.1111\/ceo.13650"},{"key":"e_1_3_2_4_2","doi-asserted-by":"publisher","DOI":"10.1136\/bjo.36.3.113"},{"key":"e_1_3_2_5_2","doi-asserted-by":"publisher","DOI":"10.1016\/S0886-3350(13)80808-5"},{"issue":"5","key":"e_1_3_2_6_2","first-page":"683","article-title":"[Late opacification of a polymethylmethacrylate (PMMA) intraocular lens: case report]","volume":"68","author":"Maia OO","year":"2005","unstructured":"Maia OO Jr, Nakashima AF, Barbosa EP, Primiano J\u00fanior HP, Nakashima Y. [Late opacification of a polymethylmethacrylate (PMMA) intraocular lens: case report]. Arq Bras Oftalmol. 2005;68(5):683\u2013685. PMID:16322870","journal-title":"Arq Bras Oftalmol"},{"key":"e_1_3_2_7_2","doi-asserted-by":"publisher","DOI":"10.1016\/j.sjopt.2011.12.001"},{"key":"e_1_3_2_8_2","doi-asserted-by":"publisher","DOI":"10.1016\/j.jcrs.2016.06.003"},{"key":"e_1_3_2_9_2","volume-title":"Soft Implant Lenses in Cataract Surgery","author":"Epstein E","year":"1986","unstructured":"Epstein E. History of intraocular lens implant surgery. In: Mazzocico TRRG, Epstein E, eds. Soft Implant Lenses in Cataract Surgery. SLACK Incorporated; 1986."},{"issue":"7","key":"e_1_3_2_10_2","first-page":"913","article-title":"Silicone intraocular lens implant discoloration in humans","volume":"109","author":"No authors listed","year":"1991","unstructured":"No authors listed. Silicone intraocular lens implant discoloration in humans. Arch Ophthalmol. 1991;109(7):913\u2013915. PMID:2064564","journal-title":"Arch Ophthalmol"},{"key":"e_1_3_2_11_2","doi-asserted-by":"publisher","DOI":"10.1001\/archopht.1992.01080150017007"},{"key":"e_1_3_2_12_2","doi-asserted-by":"publisher","DOI":"10.1038\/eye.1999.241"},{"key":"e_1_3_2_13_2","doi-asserted-by":"publisher","DOI":"10.1016\/j.ophtha.2004.06.032"},{"key":"e_1_3_2_14_2","doi-asserted-by":"publisher","DOI":"10.1038\/sj.eye.6701621"},{"key":"e_1_3_2_15_2","doi-asserted-by":"publisher","DOI":"10.1016\/j.jcrs.2006.02.047"},{"issue":"3","key":"e_1_3_2_16_2","first-page":"96","article-title":"Inesperada prevalencia de lentes intraoculares opacificadas","volume":"3","author":"Arg\u00fcello M","year":"2009","unstructured":"Arg\u00fcello M, Gallina P, Venier G, Gonorazky S. Inesperada prevalencia de lentes intraoculares opacificadas. Oftalmol Clin Exp. 2009;3(3):96\u2013101.","journal-title":"Oftalmol Clin Exp"},{"key":"e_1_3_2_17_2","doi-asserted-by":"publisher","DOI":"10.1016\/S0161-6420(01)00778-3"},{"key":"e_1_3_2_18_2","doi-asserted-by":"publisher","DOI":"10.1016\/S0886-3350(03)00461-9"},{"key":"e_1_3_2_19_2","doi-asserted-by":"publisher","DOI":"10.1016\/j.jcrs.2007.01.015"},{"key":"e_1_3_2_20_2","doi-asserted-by":"publisher","DOI":"10.1016\/j.jcrs.2007.10.029"},{"key":"e_1_3_2_21_2","doi-asserted-by":"publisher","DOI":"10.1016\/j.jcrs.2010.08.055"},{"key":"e_1_3_2_22_2","doi-asserted-by":"publisher","DOI":"10.17116\/oftalma2015131474-78"},{"key":"e_1_3_2_23_2","doi-asserted-by":"publisher","DOI":"10.1016\/j.jcrs.2016.06.032"},{"key":"e_1_3_2_24_2","doi-asserted-by":"publisher","DOI":"10.1177\/1120672119830581"},{"key":"e_1_3_2_25_2","unstructured":"Oculentis GmbH. Voluntary recall LENTIS HydroSmart foldable intraocular lenses in glass vials (CE1275). 2014. http:\/\/www.moph.gov.lb\/userfiles\/medical_care\/Lentis.pdf"},{"key":"e_1_3_2_26_2","doi-asserted-by":"publisher","DOI":"10.1016\/j.ajo.2016.04.018"},{"key":"e_1_3_2_27_2","doi-asserted-by":"publisher","DOI":"10.1016\/j.jcrs.2018.07.026"},{"key":"e_1_3_2_28_2","doi-asserted-by":"publisher","DOI":"10.1097\/MD.0000000000018379"},{"key":"e_1_3_2_29_2","doi-asserted-by":"publisher","DOI":"10.1016\/S0886-3350(01)00841-0"},{"key":"e_1_3_2_30_2","doi-asserted-by":"publisher","DOI":"10.1097\/j.jcrs.0000000000000342"},{"key":"e_1_3_2_31_2","doi-asserted-by":"publisher","DOI":"10.1016\/j.ajoc.2020.100672"},{"key":"e_1_3_2_32_2","volume-title":"Foundation for Research and Tecnology-Hellas, Institute of Chemical Engineering Sciences (Forth\/ICE-HT)","author":"Oculentis Intraocular Lens Investigation","unstructured":"Oculentis Intraocular Lens Investigation. Foundation for Research and Tecnology-Hellas, Institute of Chemical Engineering Sciences (Forth\/ICE-HT), University of Patras, Greece. Accessed September 30, 2019. http:\/\/www.oculentis.com\/Downloads\/StudyPatrasUniversitySummary.pdf"},{"key":"e_1_3_2_33_2","doi-asserted-by":"publisher","DOI":"10.1016\/j.jcrs.2013.03.035"},{"key":"e_1_3_2_34_2","doi-asserted-by":"publisher","DOI":"10.1016\/j.ophtha.2011.08.036"},{"key":"e_1_3_2_35_2","doi-asserted-by":"publisher","DOI":"10.3928\/1081597X-20111213-02"},{"key":"e_1_3_2_36_2","doi-asserted-by":"publisher","DOI":"10.1016\/j.jcrs.2010.08.043"},{"issue":"3","key":"e_1_3_2_37_2","first-page":"501","article-title":"Error induced by the estimation of the corneal power and the effective lens position with a rotationally asymmetric refractive multifocal intraocular lens","volume":"8","author":"Pi\u00f1ero DP","year":"2015","unstructured":"Pi\u00f1ero DP, Camps VJ, Ram\u00f3n ML, Mateo V, P\u00e9rez-Cambrod\u00ed RJ. Error induced by the estimation of the corneal power and the effective lens position with a rotationally asymmetric refractive multifocal intraocular lens. Int J Ophthalmol. 2015;8(3):501\u2013507. PMID:26085998","journal-title":"Int J Ophthalmol"}],"container-title":["Journal of Refractive Surgery"],"original-title":[],"language":"en","deposited":{"date-parts":[[2024,2,12]],"date-time":"2024-02-12T19:03:11Z","timestamp":1707764591000},"score":1,"resource":{"primary":{"URL":"https:\/\/journals.healio.com\/doi\/10.3928\/1081597X-20240115-01"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[2024,2]]},"references-count":36,"journal-issue":{"issue":"2","published-print":{"date-parts":[[2024,2]]}},"alternative-id":["10.3928\/1081597X-20240115-01"],"URL":"https:\/\/doi.org\/10.3928\/1081597x-20240115-01","relation":{},"ISSN":["1081-597X"],"issn-type":[{"value":"1081-597X","type":"print"}],"subject":[],"published":{"date-parts":[[2024,2]]}}}