{"status":"ok","message-type":"work","message-version":"1.0.0","message":{"indexed":{"date-parts":[[2026,8,18]],"date-time":"2026-08-18T16:34:40Z","timestamp":1787070880687,"version":"build-2736575974"},"reference-count":13,"publisher":"Ovid Technologies (Wolters Kluwer Health)","issue":"Suppl 2","content-domain":{"domain":["lww.com","ovid.com"],"crossmark-restriction":true},"short-container-title":[],"published-print":{"date-parts":[[2023,1]]},"abstract":"<jats:title>ABSTRACT<\/jats:title>\n                  <jats:sec>\n                    <jats:title>Background:<\/jats:title>\n                    <jats:p>The present study aims to evaluate the performance of an Electronic portal imaging device (EPID) for measuring dosimetric parameters and for verification of dose in small photon fields.<\/jats:p>\n                  <\/jats:sec>\n                  <jats:sec>\n                    <jats:title>Material and Methods:<\/jats:title>\n                    <jats:p>\n                      In this study, the beam profiles were obtained using the amorphous silicon (a-Si) EPID for field sizes ranging from 1 \u00d7 1 to 10 \u00d7 10 cm\n                      <jats:sup>2<\/jats:sup>\n                      at energies 6 and 18 mega-voltage (MV). For comparison, the dosimetric parameters, including penumbra widths and field sizes, were measured with the pinpoint, diode, and Semiflex dosimeters. Finally, Rando Phantom was used to compare the two-dimensional (2D) Dose distribution between EPID and Treatment Planning System (TPS).\n                    <\/jats:p>\n                  <\/jats:sec>\n                  <jats:sec>\n                    <jats:title>Results:<\/jats:title>\n                    <jats:p>In both 5 cm and 10 cm depths, there were large differences between the measured doses obtained from TPS, Pinpoint detector, and Farmer detector in 1 \u00d7 1 field size. The differences become negligible as the field sizes increase and from 3 \u00d7 3 field size to 10 \u00d7 10 field size, the maximum observed differences are 2 cGy and 2.4 cGy for 5 cm and 10 cm depths, respectively. The results indicate that the penumbra widths are smaller in the Gantry-Target (GT) direction compared to the Right-Left (RL) direction. The maximum difference (47.6%) was observed for EPID in the 10 \u00d7 10 field size, and the minimum difference (16.6%) was observed for TPS in the 1 \u00d7 1 field size. Finally, 2D dose distributions obtained by EPID and TPS exhibit excellent agreement.<\/jats:p>\n                  <\/jats:sec>\n                  <jats:sec>\n                    <jats:title>Conclusion:<\/jats:title>\n                    <jats:p>EPID is an excellent tool for the measurement of dosimetry parameters such as dose profiles, penumbra widths, field sizes, and pretreatment verification of 2D dose distributions, especially in small fields.<\/jats:p>\n                  <\/jats:sec>","DOI":"10.4103\/jcrt.jcrt_2398_22","type":"journal-article","created":{"date-parts":[[2023,12,13]],"date-time":"2023-12-13T04:05:41Z","timestamp":1702440341000},"page":"S815-S820","update-policy":"https:\/\/doi.org\/10.1097\/lww.0000000000001000","source":"Crossref","is-referenced-by-count":0,"title":["The feasibility of the pretreatment verification of 2D dose distributions in radiation therapy with small fields using the electronic portal imaging device"],"prefix":"10.4103","volume":"19","author":[{"given":"Nina","family":"Tuncel","sequence":"first","affiliation":[{"name":"Radiation Oncology Department, School of Medicine, Akdeniz University, Antalya, Turkey"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Seied Rabi","family":"Mahdavi","sequence":"additional","affiliation":[{"name":"Department of Medical Physics, School of Medicine, Iran University of Medical Sciences, Tehran, Iran"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Mohammad","family":"Haghparast","sequence":"additional","affiliation":[{"name":"Department of Radiology, Faculty of Paramedicine, Hormozgan University of Medical Sciences, Bandar Abbas, Iran"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Ismail","family":"Karakus","sequence":"additional","affiliation":[{"name":"Radiation Oncology Department, School of Medicine, Akdeniz University, Antalya, Turkey"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Abbas","family":"Haghparast","sequence":"additional","affiliation":[{"name":"Department of Medical Physics, Faculty of Medicine, Kermanshah University of Medical Sciences, Kermanshah, Iran"}],"role":[{"vocabulary":"crossref","role":"author"}]},{"given":"Vahideh","family":"Nazari","sequence":"additional","affiliation":[{"name":"Department of Medical Physics, Hamadan University of Medical Sciences, Hamadan, Iran"}],"role":[{"vocabulary":"crossref","role":"author"}]}],"member":"276","published-online":{"date-parts":[[2023,11,17]]},"reference":[{"key":"R1-20241021","doi-asserted-by":"crossref","first-page":"3095","DOI":"10.1118\/1.2040711","article-title":"Dose-response characteristics of an amorphous silicon EPID","volume":"32","author":"Winkler","year":"2005","journal-title":"Med Phys"},{"key":"R2-20241021","doi-asserted-by":"crossref","first-page":"82","DOI":"10.1120\/jacmp.v13i4.3736","article-title":"EPID dosimetry for pretreatment quality assurance with two commercial systems","volume":"13","author":"Bailey","year":"2012","journal-title":"J Appl Clin Med Phys"},{"key":"R3-20241021","article-title":"Evaluation of Perkin elmer amorphous silicon electronic portal imaging device for small photon field dosimetry","author":"Haghparast","year":"2022","journal-title":"J Biomed Phys Eng"},{"key":"R4-20241021","doi-asserted-by":"crossref","first-page":"1297","DOI":"10.1016\/S0360-3016(99)00328-4","article-title":"Transit dosimetry with an electronic portal imaging device (EPID) for 115 prostate cancer patients","volume":"45","author":"Pasma","year":"1999","journal-title":"Int J Radiat Oncol"},{"key":"R6-20241021","doi-asserted-by":"crossref","first-page":"22","DOI":"10.1120\/jacmp.v6i4.2104","article-title":"A method for deconvolution of integrated electronic portal images to obtain incident fluence for dose reconstruction","volume":"6","author":"Renner","year":"2005","journal-title":"J Appl Clin Med Phys"},{"key":"R7-20241021","doi-asserted-by":"crossref","first-page":"21","DOI":"10.4103\/2228-7477.199158","article-title":"Multi-parametric improvements in the CCD camera-based EPID for portal dosimetry","volume":"7","author":"Nazari","year":"2017","journal-title":"J Med Signals Sens"},{"key":"R8-20241021","doi-asserted-by":"crossref","first-page":"273","DOI":"10.1186\/s13014-014-0273-8","article-title":"Feasibility study on the verification of actual beam delivery in a treatment room using EPID transit dosimetry","volume":"9","author":"Baek","year":"2014","journal-title":"Radiat Oncol"},{"key":"R9-20241021","doi-asserted-by":"crossref","first-page":"041707","DOI":"10.1118\/1.4868461","article-title":"A practical and theoretical definition of very small field size for radiotherapy output factor measurements","volume":"41","author":"Charles","year":"2014","journal-title":"Med Phys"},{"key":"R10-20241021","doi-asserted-by":"crossref","first-page":"59","DOI":"10.4103\/jmp.JMP_131_16","article-title":"Investigating the electronic portal imaging device for small radiation field measurements","volume":"42","author":"Agarwal","year":"2017","journal-title":"J Med Phys"},{"key":"R11-20241021","doi-asserted-by":"crossref","first-page":"9","DOI":"10.1002\/acm2.12007","article-title":"Pixel response-based EPID dosimetry for patient specific QA","volume":"18","author":"Han","year":"2017","journal-title":"J Appl Clin Med Phys"},{"key":"R12-20241021","doi-asserted-by":"crossref","first-page":"12","DOI":"10.4103\/0971-6203.177284","article-title":"Analysis of small field percent depth dose and profiles: Comparison of measurements with various detectors and effects of detector orientation with different jaw settings","volume":"41","author":"Godson","year":"2016","journal-title":"J Med Phys"},{"key":"R13-20241021","doi-asserted-by":"crossref","first-page":"238","DOI":"10.1120\/jacmp.v11i4.3268","article-title":"Portal dosimetry for pretreatment verification of IMRT plan: A comparison with 2D ion chamber array","volume":"11","author":"Sharma","year":"2010","journal-title":"J Appl Clin Med Phys"},{"key":"R14-20241021","doi-asserted-by":"crossref","first-page":"289","DOI":"10.1016\/j.radonc.2008.07.008","article-title":"A literature review of electronic portal imaging for radiotherapy dosimetry","volume":"88","author":"van Elmpt","year":"2008","journal-title":"Radiother Oncol"}],"container-title":["Journal of Cancer Research and Therapeutics"],"original-title":[],"language":"en","link":[{"URL":"https:\/\/www.ovid.com\/10.4103\/jcrt.jcrt_2398_22","content-type":"unspecified","content-version":"vor","intended-application":"similarity-checking"}],"deposited":{"date-parts":[[2026,8,18]],"date-time":"2026-08-18T15:38:08Z","timestamp":1787067488000},"score":1,"resource":{"primary":{"URL":"https:\/\/www.ovid.com\/10.4103\/jcrt.jcrt_2398_22"}},"subtitle":[],"short-title":[],"issued":{"date-parts":[[2023,1]]},"references-count":13,"journal-issue":{"issue":"Suppl 2","published-print":{"date-parts":[[2023]]}},"URL":"https:\/\/doi.org\/10.4103\/jcrt.jcrt_2398_22","relation":{},"ISSN":["0973-1482","1998-4138"],"issn-type":[{"value":"0973-1482","type":"print"},{"value":"1998-4138","type":"electronic"}],"subject":[],"published":{"date-parts":[[2023,1]]},"assertion":[{"value":"2022-11-13","name":"received","label":"Received","group":{"name":"publication_history","label":"Publication History"}},{"value":"2023-02-06","name":"accepted","label":"Accepted","group":{"name":"publication_history","label":"Publication History"}}]}}